Towards optimization in digital chest radiography using Monte Carlo modelling

Ullman, Gustaf (författare)
Linköpings universitet,Medicinsk radiofysik,Hälsouniversitetet
Sandborg, Michael (författare)
Linköpings universitet,Medicinsk radiofysik,Hälsouniversitetet
Dance, David R. (författare)
Joint Department of Physics, The Royal Marsden NHS Foundation Trust and Institute of Cancer Research, London, UK
visa fler...
Hunt, Roger A. (författare)
Joint Department of Physics, The Royal Marsden NHS Foundation Trust and Institute of Cancer Research, London, UK
Alm Carlsson, Gudrun (författare)
Linköpings universitet,Medicinsk radiofysik,Hälsouniversitetet
visa färre...
 (creator_code:org_t)
2006-05-09
2006
Engelska.
Ingår i: Physics in medicine and biology. - : IOP Publishing. - 0031-9155 .- 1361-6560. ; 51:11, s. 2729-2743
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • A Monte Carlo based computer model of the x-ray imaging system was used to investigate how various image quality parameters of interest in chest PA radiography and the effective dose E vary with tube voltage (90–150 kV), additional copper filtration (0–0.5 mm), anti-scatter method (grid ratios 8–16 and air gap lengths 20–40 cm) and patient thickness (20–28 cm) in a computed radiography (CR) system. Calculated quantities were normalized to a fixed value of air kerma (5.0 µGy) at the automatic exposure control chambers. Soft-tissue nodules were positioned at different locations in the anatomy and calcifications in the apical region. The signal-to-noise ratio, SNR, of the nodules and the nodule contrast relative to the contrast of bone (C/CB) as well as relative to the dynamic range in the image (Crel) were used as image quality measures. In all anatomical regions, except in the densest regions in the thickest patients, the air gap technique provides higher SNR and contrast ratios than the grid technique and at a lower effective dose E. Choice of tube voltage depends on whether quantum noise (SNR) or the contrast ratios are most relevant for the diagnostic task. SNR increases with decreasing tube voltage while C/CB increases with increasing tube voltage.

Nyckelord

MEDICINE
MEDICIN

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy