Use of an enhanced gradient system for diffusion MR imaging with motion-artifact reduction

Brockstedt, Sara, (författare)
Lunds universitet, Lund University, Medicinska fakulteten, Faculty of Medicine, Institutionen för kliniska vetenskaper, Lund, Department of Clinical Sciences, Lund, Sektion V, Section V, Medicinsk strålningsfysik, Lund, Medical Radiation Physics, Lund
Thomsen, C (författare)
Wirestam, Ronnie, (författare)
Lunds universitet, Lund University, Medicinska fakulteten, Faculty of Medicine, Institutionen för kliniska vetenskaper, Lund, Department of Clinical Sciences, Lund, Sektion V, Section V, Medicinsk strålningsfysik, Lund, Medical Radiation Physics, Lund
visa fler...
De Poorter, J (författare)
De Wagter, C (författare)
Salford, Leif, (författare)
Lunds universitet, Lund University, Medicinska fakulteten, Faculty of Medicine, Institutionen för kliniska vetenskaper, Lund, Department of Clinical Sciences, Lund, Sektion IV, Section IV, Neurokirurgi, Neurosurgery
Holtås, Stig, (författare)
Lunds universitet, Lund University, Medicinska fakulteten, Faculty of Medicine, Institutionen för kliniska vetenskaper, Lund, Department of Clinical Sciences, Lund, Sektion V, Section V, Diagnostisk radiologi, Lund, Diagnostic Radiology, (Lund)
Ståhlberg, Freddy, (författare)
Lunds universitet, Lund University, Medicinska fakulteten, Faculty of Medicine, Institutionen för kliniska vetenskaper, Lund, Department of Clinical Sciences, Lund, Sektion V, Section V, Diagnostisk radiologi, Lund, Diagnostic Radiology, (Lund), Medicinsk strålningsfysik, Lund, Medical Radiation Physics, Lund
visa färre...
1995
Engelska.
Ingår i: Acta Radiologica. - SAGE Publications. - 1600-0455. ; 36:6, s. 662-670
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • PURPOSE: A spin-echo diffusion-sensitized pulse sequence using high gradients (23 mT/m) is introduced. MATERIAL AND METHODS: In order to minimize motion artefacts, velocity-compensating gradients, ECG-triggering and post-processing with phase correction and raw data averaging using navigator echoes was performed. The in vitro ratio of diffusion coefficients for water and acetone was determined and the water self-diffusion coefficient at different temperatures was evaluated. The pulse sequence was tested in 7 healthy volunteers and in 2 tumour patients with astrocytomas of grades I-II and III-IV. Both single-slice and multi-slice techniques were used. RESULTS: The incorporation of phase correction clearly improved the quality of both diffusion-encoded images and the calculated diffusion maps. Mean values of the diffusion coefficients in vivo were for CSF 2.66 x 10(-9) m2/s and for white and grey matter 0.69 x 10(-9) m2/s and 0.87 x 10(-9) m2/s, respectively. CONCLUSION: Velocity-compensating gradients in combination with a high gradient strength were shown to be useful for in vivo diffusion MR imaging.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Publikations- och innehållstyp

art (ämneskategori)
ref (ämneskategori)
art (ämneskategori)
ref (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy