SwePub
Tyck till om SwePub Sök här!
Sök i SwePub databas

  Utökad sökning

Träfflista för sökning "WFRF:(Knöös Tommy) "

Sökning: WFRF:(Knöös Tommy)

  • Resultat 1-10 av 67
Sortera/gruppera träfflistan
   
NumreringReferensOmslagsbildHitta
1.
  • Aarup, Lasse Rye, et al. (författare)
  • The effect of different lung densities on the accuracy of various radiotherapy dose calculation methods: Implications for tumour coverage
  • 2009
  • Ingår i: Radiotherapy and Oncology. - : Elsevier BV. - 1879-0887 .- 0167-8140. ; 91:3, s. 405-414
  • Tidskriftsartikel (refereegranskat)abstract
    • Purpose: To evaluate against Monte-Carlo the performance of various dose calculations algorithms regarding lung turnout coverage in stereotactic body radiotherapy (SBRT) conditions. Materials and methods: Dose distributions in virtual lung phantoms have been calculated using four commercial Treatment Planning System (TPS) algorithms and one Monte Carlo (MC) system (EGSnrc). We compared the performance of the algorithms in calculating the target dose for different degrees of lung inflation. The phantoms had a cubic 'body' and 'lung' and a central 2-cm diameter spherical 'tumour' (the body and turnout have unit density). The lung tissue was assigned five densities (rho(lung)): 0.01, 0.1, 0.2, 0.4 and 1 g/cm(3). Four-field treatment plans were calculated with 6- and 18 MV narrow beams for each value of rho(lung). We considered the Pencil Beam Convolution (PBCEl) and the Analytical Anisotropic Algorithm (AAA(ECl)) from Varian Eclipse and the Pencil Beam Convolution (PBCOMP) and the Collapsed Cone Convolution (CCCOMP) algorithms from Oncentra MasterPlan. Results: When changing rho(lung) from 0.4 to 0.1 g/cm(3), the MC median target dose decreased from 89.2% to 74.9% for 6 MV and from 83.3% to 61.6% for 18 MV (of dose maximum in the homogenous case at both energies), while for both PB algorithms the median target dose was virtually independent of lung density. Conclusions: Both PB algorithms overestimated the target dose, the overestimation increasing as rho(lung) decreased. Concerning target dose, the AAA(ECl) and CCCOMP algorithms appear to be adequate alternatives to MC. (C) 2009 Elsevier Ireland Ltd. All rights reserved. Radiotherapy and oncology 91 (2009) 405-414
  •  
2.
  • Ahnesjö, Anders, et al. (författare)
  • Application of the convolution method for calculation of output factors for therapy photon beams
  • 1992
  • Ingår i: Medical Physics. - : Wiley. - 0094-2405. ; 19:2, s. 295-301
  • Tidskriftsartikel (refereegranskat)abstract
    • The output factor for a therapy photon beam is defined as the dose per monitor unit relative to the dose per monitor unit in a reference field. Convolution models for photon dose calculations yield the dose in units normalized to the incident energy fluence with phantom scatter intrinsically modeled. Output factors calculated with the convolution method as the dose per unit energy fluence relative to the calculated dose per unit energy fluence in a reference field could deviate as much as 5% if corrections are not made for perturbations due to treatment head scatter. Significant perturbations are particles backscattered from the collimators to the monitor and photons forward scattered from the filter and collimators in the treatment head. The forward scatter adds an "unmonitored" contribution to the total energy fluence of the beam. A model is developed that describes the field size dependence of these perturbations for conversion of output factors, calculated with the convolution method, to machine output factors as an integrated part in treatment planning. The necessary machine characteristics are derived from measurements of the output in air for a limited set of field sizes. The method has been tested using five different multileaf collimated irregular fields at 6 MV and for a large set of rectangular fields at 5, 6, and 18 MV and found to predict output factors with an accuracy better than 1%.
  •  
3.
  •  
4.
  • Björk, Peter, et al. (författare)
  • Comparative dosimetry of diode and diamond detectors in electron beams for intraoperative radiation therapy
  • 2000
  • Ingår i: Medical Physics. - : Wiley. - 0094-2405. ; 27:11, s. 2580-2588
  • Tidskriftsartikel (refereegranskat)abstract
    • The aim of the present study is to examine the validity of using silicon semiconductor detectors in degraded electron beams with a broad energy spectrum and a wide angular distribution. A comparison is made with diamond detector measurements, which is the dosimeter considered to give the best results provided that dose rate effects are corrected for. Two-dimensional relative absorbed dose distributions in electron beams (6-20 MeV) for intraoperative radiation therapy (IORT) are measured in a water phantom. To quantify deviations between the detectors, a dose comparison tool that simultaneously examines the dose difference and distance to agreement (DTA) is used to evaluate the results in low- and high-dose gradient regions, respectively. Uncertainties of the experimental measurement setup (+/- 1% and +/- 0.5 mm) are taken into account by calculating a composite distribution that fails this dose-difference and DTA acceptance limit. Thus, the resulting area of disagreement should be related to differences in detector performance. The dose distributions obtained with the diode are generally in very good agreement with diamond detector measurements. The buildup region and the dose falloff region show good agreement with increasing electron energy, while the region outside the radiation field close to the water surface shows an increased difference with energy. The small discrepancies in the composite distributions are due to several factors: (a) variation of the silicon-to-water collision stopping-power ratio with electron energy, (b) a more pronounced directional dependence for diodes than for diamonds, and (c) variation of the electron fluence perturbation correction factor with depth. For all investigated treatment cones and energies, the deviation is within dose-difference and DTA acceptance criteria of +/- 3% and +/- 1 mm, respectively. Therefore, p-type silicon diodes are well suited, in the sense that they give results in close agreement with diamond detectors, for practical measurements of relative absorbed dose distributions in degraded electron beams used for IORT.
  •  
5.
  • Björk, Peter, et al. (författare)
  • Design and dosimetry characteristics of a soft-docking system for intraoperative radiation therapy
  • 2000
  • Ingår i: International Journal of Radiation Oncology, Biology, Physics. - 0360-3016. ; 47:2, s. 527-533
  • Tidskriftsartikel (refereegranskat)abstract
    • PURPOSE: The design concept and the dosimetric characteristics of an applicator system for intraoperative radiation therapy (IORT) with special emphasis on alignment methods, the effect of a plastic scatterer in the beam, radiation leakage, and misalignment dosimetry, are presented in this paper. MATERIALS AND METHODS: A soft-docking system for a linear accelerator, which enables collimation of electron beams (4-22 MeV) for IORT has been developed. The system includes twenty-one circular polymethylmethacrylate (PMMA) treatment cones of different lengths, diameters and end angles. All in-water measurements are made using p-type silicon diode detectors. RESULTS: The effect of introducing a PMMA scatterer in the therapeutic beam includes increased surface dose values (above 83% for all nominal electron energies and for all cones) and improved dose homogeneity within the therapeutic range. Electrons scattered from the inside wall of the cone result in dose profile horns at depth of dose maximum always lower than 109%. The radiation leakage outside the cone is less than 13%. Large changes in the dose profiles occur if the intraoperative cone is misaligned more than 0.5. CONCLUSION: The alignment procedure of the soft-docking system is easy to handle and the applicator design provides adequate collimation of electron beams for IORT.
  •  
6.
  • Björk, Peter, et al. (författare)
  • Dosimetry characteristics of degraded electron beams investigated by Monte Carlo calculations in a setup for intraoperative radiation therapy.
  • 2002
  • Ingår i: Physics in Medicine and Biology. - : IOP Publishing. - 1361-6560 .- 0031-9155. ; 47:2, s. 239-256
  • Tidskriftsartikel (refereegranskat)abstract
    • Degraded electron beams, as used for intraoperative radiation therapy (IORT) or similar complicated dosimetric situations, have different characteristics compared to conventional electron therapy beams. If international dosimetry protocols are applied in a direct manner to such degraded beams, uncertainties will be introduced in the absorbed dose determination. The Monte Carlo method has been used to verify experimentally determined relative absorbed dose distributions and output factors in an IORT geometry. Monte Carlo generated dose distributions are mostly within +/-2% or +/-2 mm of measured data. The simulated output variation between the IORT cones (relative output factors) are mostly within 2% of measured values. By comparing IORT and conventional electron beam characteristics (e.g. energy spectra, angular distributions and the contributions of different system components to these quantities) limitations and uncertainties of commonly used dosimetric techniques in IORT electron fields are quantified. The intraoperative treatment field contains a larger amount of scattered electrons, which leads to a broader energy spectrum as well as a wider angular distribution of electrons at the phantom surface. The dose from the scattered electrons can contribute up to 40% of the total dose at a depth of dose maximum, compared to approximately 10% for standard beams. A study of the energy spectra at the reference depth reveals that an uncertainty of the order of 1% can be introduced if ionization chamber based dosimetry is used to determine output factors for the investigated IORT system. We recommend that relative absorbed dose distributions and output factors in IORT electron beams and for similar complicated dosimetric situations should be determined with detectors having a small energy and angular dependence (e.g. diamond detectors or p-Si diodes).
  •  
7.
  •  
8.
  •  
9.
  •  
10.
  • Blad, Börje, et al. (författare)
  • The influence of air humidity on an unsealed ionization chamber in a linear accelerator
  • 1996
  • Ingår i: Physics in Medicine and Biology. - : IOP Publishing. - 1361-6560 .- 0031-9155. ; 41:11, s. 2541-2548
  • Tidskriftsartikel (refereegranskat)abstract
    • The safe and accurate delivery of the prescribed absorbed dose is the central function of the dose monitoring and beam stabilization system in a medical linear accelerator. The absorbed dose delivered to the patient during radiotherapy is often monitored by a transmission ionization chamber. Therefore it is of utmost importance that the chamber behaves correctly. We have noticed that the sensitivity of an unsealed chamber in a Philips SL linear accelerator changes significantly, especially during and after the summer season. The reason for this is probably a corrosion effect of the conductive plates in the chamber due to the increased relative humidity during hot periods. We have found that the responses of the different ion chamber plates change with variations in air humidity and that they do not return to their original values when the air humidity is returned to ambient conditions.
  •  
Skapa referenser, mejla, bekava och länka
  • Resultat 1-10 av 67

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy