- Malaquias, A, et al.
(författare)
-
Active beam spectroscopy diagnostics for ITER : Present status (invited)
- 2004
-
Ingår i: Review of Scientific Instruments. - : AIP Publishing. - 0034-6748 .- 1089-7623. ; 75:10, s. 3393-3398
-
Tidskriftsartikel (refereegranskat)abstract
- An overview is given of the present design of the active beam spectroscopy diagnostics for the ITER. Present spatial resolution and signal-to-noise indicate that, in principle, all proposed measurements are possible covering the entire plasma minor radius. Calculations show that the mirror based periscopes and the impurity coating of the first mirror affect the signal strength and the polarization characteristics of the measured spectra having an impact in the measurement accuracy. On-line calibration techniques and methods to access the first mirror status are addressed.
|
|