SwePub
Tyck till om SwePub Sök här!
Sök i SwePub databas

  Utökad sökning

Träfflista för sökning "WFRF:(Oei Edwin H. G.) srt2:(2023)"

Sökning: WFRF:(Oei Edwin H. G.) > (2023)

  • Resultat 1-3 av 3
Sortera/gruppera träfflistan
   
NumreringReferensOmslagsbildHitta
1.
  • Booij, Ronald, et al. (författare)
  • Assessment of visibility of bone structures in the wrist using normal and half of the radiation dose with photon-counting detector CT
  • 2023
  • Ingår i: European Journal of Radiology. - : ELSEVIER IRELAND LTD. - 0720-048X .- 1872-7727. ; 159
  • Tidskriftsartikel (refereegranskat)abstract
    • Purpose: To quantitatively and qualitatively assess the visibility of bone structures in the wrist on photon-counting detector computed tomography (PCD-CT) images compared to state-of-the-art energy-integrating de-tector CT (EID-CT).Method: Four human cadaveric wrist specimens were scanned with EID-CT and PCD-CT at identical CTDIvolof 12.2 mGy and with 6.1 mGy (half dose PCD-CT). Axial images were reconstructed using the thinnest possible slice thickness, i.e. 0.4 mm on EID-CT and 0.2 mm on PCD-CT, with the largest image matrix size possible using reconstruction kernels optimized for bone (EID-CT: Ur68, PCD-CT: Br92). Quantitative evaluation was performed to determine contrast-noise ratio (CNR) of bone/ fat, cortical and trabecular sharpness. An observer study using visual grading characteristics (VGC) analysis was performed by six observers to assess the visibility of nutrient canals, trabecular architecture, cortical bone and the general image quality.Results: At equal dose, images obtained with PCD-CT had 39 +/- 6 % lower CNR (p = 0.001), 71 +/- 57 % higher trabecular sharpness in the radius (p = 0.02) and 42 +/- 8 % (p < 0.05) sharper cortical edges than those obtained with EID-CT. This was confirmed by VGC analysis showing a superior visibility of nutrient canals, trabeculae and cortical bone area under the curve (AUC) > 0.89) for PCD-CT, even at half dose.Conclusions: Despite a lower CNR and increased noise, the trabecular and cortical sharpness were twofold higher with PCD-CT. Visual grading analysis demonstrated superior visibility of cortical bone, trabeculae, nutrient canals and an overall improved image quality with PCD-CT over EID-CT. At half dose, PCD-CT also yielded superior image quality, both in quantitative measures and as evaluated by radiologists.
  •  
2.
  • Emin, Sevgi, et al. (författare)
  • Imaging-based assessment of fatty acid composition in human bone marrow adipose tissue at 7 T : Method comparison and in vivo feasibility
  • 2023
  • Ingår i: Magnetic Resonance in Medicine. - : Wiley. - 0740-3194 .- 1522-2594. ; 90:1, s. 240-249
  • Tidskriftsartikel (refereegranskat)abstract
    • Purpose: To demonstrate the feasibility and accuracy of chemical shift–encoded imaging of the fatty acid composition (FAC) of human bone marrow adipose tissue at 7 T, and to determine suitable image-acquisition parameters using simulations. Methods: The noise performance of FAC estimation was investigated using simulations with a range of inter-echo time, and accuracy was assessed using a phantom experiment. Furthermore, one knee of 8 knee-healthy subjects (ages 35–54 years) was imaged, and the fractions of saturated fatty acids (SFA) and polyunsaturated fatty acids (PUFA) were mapped. Values were compared between reconstruction methods, and between anatomical regions. Results: Based on simulations, ΔTE = 0.6 ms was chosen. The phantom experiment demonstrated high accuracy of especially SFA using a constrained reconstruction model (slope = 1.1, average bias = −0.2%). The lowest accuracy was seen for PUFA using a free model (slope = 2.0, average bias = 9.0%). For in vivo images, the constrained model resulted in lower intersubject variation compared with the free model (e.g., in the femoral shaft, the SFA percent-point range was within 1.0% [vs. 3.0%]). Furthermore, significant regional FAC differences were detected. For example, using the constrained approach, the femoral SFA in the medial condyle was lower compared with the shaft (median [range]: 27.9% [27.1%, 28.4%] vs. 32.5% [31.8%, 32.8%]). Conclusion: Bone marrow adipose tissue FAC quantification using chemical-shift encoding is feasible at 7 T. Both the noise performance and accuracy of the technique are superior using a constrained signal model.
  •  
3.
  • Woisetschläger, Mischa, et al. (författare)
  • Improved visualization of the bone-implant interface and osseointegration in ex vivo acetabular cup implants using photon-counting detector CT
  • 2023
  • Ingår i: EUROPEAN RADIOLOGY EXPERIMENTAL. - : SPRINGERNATURE. - 2509-9280. ; 7:1
  • Tidskriftsartikel (refereegranskat)abstract
    • BackgroundSuccessful osseointegration of joint replacement implants is required for long-term implant survival. Accurate assessment of osseointegration could enable clinical discrimination of failed implants from other sources of pain avoiding unnecessary surgeries. Photon-counting detector computed tomography (PCD-CT) provides improvements in image resolution compared to conventional energy-integrating detector CT (EID-CT), possibly allowing better visualization of bone-implant-interfaces and osseointegration. The aim of this study was to assess the quality of visualization of bone-implant-interfaces and osseointegration in acetabular cup implants, using PCD-CT compared with EID-CT.MethodsTwo acetabular implants (one cemented, one uncemented) retrieved during revision surgery were scanned using PCD-CT and EID-CT at equal radiation dose. Images were reconstructed using different reconstruction kernels and iterative strengths. Delineation of the bone-implant and bone-cement-interface as an indicator of osseointegration was scored subjectively for image quality by four radiologists on a Likert scale and assessed quantitatively.ResultsDelineation of bone-implant and bone-cement-interfaces was better with PCD-CT compared with EID-CT (p <= 0.030). The highest ratings were given for PCD-CT at sharper kernels for the cemented cup (PCD-CT, median 5, interquartile range 4.25-5.00 versus EID-CT, 3, 2.00-3.75, p < 0.001) and the uncemented cup (5, 4.00-5.00 versus 2, 2-2, respectively, p < 0.001). The bone-implant-interface was 35-42% sharper and the bone-cement-interface was 28-43% sharper with PCD-CT compared with EID-CT, depending on the reconstruction kernel.ConclusionsPCD-CT might enable a more accurate assessment of osseointegration of orthopedic joint replacement implants.
  •  
Skapa referenser, mejla, bekava och länka
  • Resultat 1-3 av 3

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy