Evaluating doppler devices using a moving string test target

Walker, Andrew (författare)
Linköpings universitet,Fysiologisk mätteknik,Tekniska högskolan
Philips, David J. (författare)
Department of Surgery and Center for Bioengineering, University of Washington, Seattle
Powers, Jeffry E. (författare)
Department of Surgery and Center for Bioengineering, University of Washington, Seattle
 (creator_code:org_t)
Wiley InterScience, 1982
1982
Engelska.
Ingår i: Journal of Clinical Ultrasound. - : Wiley InterScience. - 0091-2751 .- 1097-0096. ; 10:1, s. 25-30
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Two moving string test targets have been developed and used to characterize ultrasound Doppler instrumentation. The characteristics investigated were sample volume size and location. The tests were performed on a combined echo/Doppler instrument and on an annular array system. The procedures can be carried out routinely in the clinical laboratory to ensure that the instrument is working properly or as an aid for correct interpretation of acquired data.

Nyckelord

Quality assurance
Doppler
string target
Ultrasound test phantom
Pulsed Doppler
Duplex ultrasound
MEDICINE
MEDICIN

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy