SwePub
Sök i SwePub databas

  Utökad sökning

Träfflista för sökning "L773:0094 2405 OR L773:2473 4209 "

Sökning: L773:0094 2405 OR L773:2473 4209

  • Resultat 21-30 av 346
Sortera/gruppera träfflistan
   
NumreringReferensOmslagsbildHitta
21.
  • Eriksson, O., et al. (författare)
  • Robust automated radiation therapy treatment planning using scenario-specific dose prediction and robust dose mimicking
  • 2022
  • Ingår i: Medical physics (Lancaster). - : Wiley. - 0094-2405 .- 2473-4209. ; 49:6, s. 3564-3573
  • Tidskriftsartikel (refereegranskat)abstract
    • Purpose: We present a framework for robust automated treatment planning using machine learning, comprising scenario-specific dose prediction and robust dose mimicking. Methods: The scenario dose prediction pipeline is divided into the prediction of nominal dose from input image and the prediction of scenario dose from nominal dose, each using a deep learning model with U-net architecture. By using a specially developed dose–volume histogram–based loss function, the predicted scenario doses are ensured sufficient target coverage despite the possibility of the training data being non-robust. Deliverable plans may then be created by solving a robust dose mimicking problem with the predictions as scenario-specific reference doses. Results: Numerical experiments are performed using a data set of 52 intensity-modulated proton therapy plans for prostate patients. We show that the predicted scenario doses resemble their respective ground truth well, in particular while having target coverage comparable to that of the nominal scenario. The deliverable plans produced by the subsequent robust dose mimicking were showed to be robust against the same scenario set considered for prediction. Conclusions: We demonstrate the feasibility and merits of the proposed methodology for incorporating robustness into automated treatment planning algorithms. 
  •  
22.
  • Fan, Peng, et al. (författare)
  • Scatter and crosstalk corrections for (99m)Tc/(123)I dual-radionuclide imaging using a CZT SPECT system with pinhole collimators.
  • 2015
  • Ingår i: Medical Physics. - : Wiley. - 0094-2405 .- 2473-4209. ; 42:12, s. 6895-6911
  • Tidskriftsartikel (refereegranskat)abstract
    • The energy spectrum for a cadmium zinc telluride (CZT) detector has a low energy tail due to incomplete charge collection and intercrystal scattering. Due to these solid-state detector effects, scatter would be overestimated if the conventional triple-energy window (TEW) method is used for scatter and crosstalk corrections in CZT-based imaging systems. The objective of this work is to develop a scatter and crosstalk correction method for (99m)Tc/(123)I dual-radionuclide imaging for a CZT-based dedicated cardiac SPECT system with pinhole collimators (GE Discovery NM 530c/570c).
  •  
23.
  • Georgi, Peter, et al. (författare)
  • Determination of intrinsic energy dependence of point-like inorganic scintillation detector in brachytherapy
  • 2024
  • Ingår i: Medical physics (Lancaster). - : WILEY. - 0094-2405. ; 51:7, s. 5059-5069
  • Tidskriftsartikel (refereegranskat)abstract
    • Background: Inorganic scintillation detectors (ISDs) are promising for in vivo dosimetry in brachytherapy (BT). ISDs have fast response, providing time resolved dose rate information, and high sensitivity, attributed to high atomic numbers. However, the conversion of the detector signal to absorbed dose-to-water is highly dependent on the energy spectrum of the incident radiation. This dependence is comprised of absorbed dose energy dependence, obtainable with Monte Carlo (MC) simulation, and the absorbed dose-to-signal conversion efficiency or intrinsic energy dependence requiring measurements. Studies have indicated negligible intrinsic energy dependence of ZnSe:O-based ISDs in Ir-192 BT. A full characterization has not been performed earlier.Purpose: This study characterizes the intrinsic energy dependence of ZnSe:O-based ISDs for kV X-ray radiation qualities, with energies relevant for BT.Methods: Three point-like ISDs made from fiber-coupled cuboid ZnSe:O-based scintillators were calibrated at the Swedish National Metrology Laboratory for ionizing radiation. The calibration was done in terms of air kerma free-in-air, K-air, in 13 X-ray radiation qualities, Q, from 25 to 300 kVp (CCRI 25-250 kV and ISO 4037 N-series), and in terms of absorbed dose to water, D-w, in a Co-60 beam, Q(0).The mean absorbed dose to the ISDs, relative to K-air and D-w, were obtained with the MC code TOPAS (Geant4) using X-ray spectra obtained with SpekPy software and laboratory filtration data and a generic Co-60 source.The intrinsic energy dependence was determined as a function of effective photon energy, E-eff, (relative to Co-60).The angular dependence of the ISD signal was measured in a 25 kVp (0.20 mm Al HVL) and 135 kVp beam (0.48 mm Cu HVL), by rotating the ISDs 180 degrees around the fiber's longitudinal axis (perpendicular to the beam). A full 360 degrees was not performed due to setup limitations. The impact of detector design was quantified with MC simulation.ResultsAbove 30 keV E-eff the intrinsic energy dependence varied with less than 5 +/- 4% from unity for all detectors (with the uncertainty expressed as the mean of all expanded measurement uncertainties for individual E-eff above 30 keV, k = 2). Below 30 keV, it decreased with up to 17% and inter-detector variations of 13% were observed, likely due to differences in detector geometry not captured by the simulations using nominal geometry. In the 25 kVp radiation quality, the ISD signal varied with 24% over a similar to 45 degrees rotation. For 135 kVp, the corresponding variation was below 3%. Assuming a 0.05 mm thicker layer of reflective paint around the sensitive volume changed the absorbed dose with 6.3% at the lowest E-eff, and with less than 2% at higher energies.ConclusionThe study suggests that the ISDs have an intrinsic energy dependence relative to Co-60 lower than 5 +/- 4% in radiation qualities with E-eff > 30 keV. Therefore, they could in principle be calibrated in a Co-60 beam quality and transferred to such radiation qualities with correction factors determined only by the absorbed dose energy dependence obtained from MC simulations. This encourages exploration of the ISDs' applications in intensity modulated BT with Yb-169 or other novel intermediate energy isotopes.
  •  
24.
  • Ghazal, Mohammed, et al. (författare)
  • 6-MV small field output factors: intra-/intermachine comparison and implementation of TRS-483 using various detectors and several linear accelerators
  • 2019
  • Ingår i: Medical physics (Lancaster). - : WILEY. - 0094-2405 .- 2473-4209. ; 46:11, s. 5350-5359
  • Tidskriftsartikel (refereegranskat)abstract
    • Purpose To investigate the applicability of output correction factors reported in TRS-483 on 6-MV small-field detector-reading ratios using four solid-state detectors. Also, to investigate variations in 6-MV small-field output factors (OF) among nominally matched linear accelerators (linacs). Methods The TRS-483 Code of Practice (CoP) introduced and provided output correction factors to be applied to measured detector-reading ratios to obtain OFs for several small-field detectors. Detector readings for 0.5 cm x 0.5 cm to 8 cm x 8 cm fields were measured and normalized to that of 10 cm x 10 cm field giving the detector-reading ratios. Three silicon diodes, IBA PFD, IBA EFD (IBA, Schwarzenbruck, Germany), PTW T60017, and one microdiamond, PTW T60019 (PTW, Freiburg, Germany), were used. Output correction factors from the CoP were applied to measured detector-reading ratios. Measurements were performed on six Clinac and six TrueBeam linacs (Varian Medical Systems, Palo Alto, USA). An investigation of the relationship between the size of small fields and corresponding detector-reading ratio among the linacs was performed by measuring lateral dose profiles for 0.5 cm x 0.5 cm fields to determine the full width half maximum (FWHM). The relationship between the linacs focal spot size and the small-field detector-reading ratio was investigated by measuring 10 cm x 10 cm lateral dose profiles and determining the penumbra width reflecting the focal spot size. Measurement geometry was as follows: gantry angle = 0 degrees, collimator angle = 0 degrees, source-to surface distance (SSD) = 90 cm, and depth in water = 10 cm. Results For a given linac and 0.5 cm x 0.5 cm field, the deviations in detector-reading ratios among the detectors were 9%-15% for the Clinacs and 4%-5% for the TrueBeams. Use of output correction factors reduced these deviations to 6%-12% and 3%-4%, respectively. For field sizes equal to or larger than 0.8 cm x 0.8 cm, the deviations were corrected to 1% using output correction factors for both Clinacs and TrueBeams. For a given detector and 0.5 cm x 0.5 cm field, the deviations in detector-reading ratios among the linacs were 11%-17% for the Clinacs and 5-6% for the TrueBeams. For 1 cm x 1 cm the deviations were 1%-2% for Clinacs and 1% for TrueBeams. For field sizes larger than 1 cm x 1 cm the deviations were within 1% for both Clinacs and TrueBeams. No relationship between FWHMs and detector-reading ratios for 0.5 cm x 0.5 cm was observed. For Clinacs, larger 10 cm x 10 cm penumbra width yielded lower 0.5 cm x 0.5 cm detector-reading ratio indicating an effect of the focal spot size. For TrueBeams, the spread of penumbra widths was lower compared to Clinacs and no similar relationship was observed. Conclusions Output correction factors from the TRS-483 CoP are not sufficient for accurate determination of OF for 0.5 cm x 0.5 cm fields but are applicable for 0.8 cm x 0.8 cm to 8 cm x 8 cm fields. Nominally matched Clinacs and TrueBeams show large differences in detector-reading ratios for fields smaller than 1 cm x 1 cm.
  •  
25.
  • Giantsoudi, D., et al. (författare)
  • A gEUD-based inverse planning technique for HDR prostate brachytherapy : Feasibility study
  • 2013
  • Ingår i: Medical physics (Lancaster). - : Wiley. - 0094-2405 .- 2473-4209. ; 40:4, s. 041704-
  • Tidskriftsartikel (refereegranskat)abstract
    • Purpose: The purpose of this work was to study the feasibility of a new inverse planning technique based on the generalized equivalent uniform dose for image-guided high dose rate (HDR) prostate cancer brachytherapy in comparison to conventional dose-volume based optimization. Methods: The quality of 12 clinical HDR brachytherapy implants for prostate utilizing HIPO (Hybrid Inverse Planning Optimization) is compared with alternative plans, which were produced through inverse planning using the generalized equivalent uniform dose (gEUD). All the common dose-volume indices for the prostate and the organs at risk were considered together with radiobiological measures. The clinical effectiveness of the different dose distributions was investigated by comparing dose volume histogram and gEUD evaluators. Results: Our results demonstrate the feasibility of gEUD-based inverse planning in HDR brachytherapy implants for prostate. A statistically significant decrease in D-10 or/and final gEUD values for the organs at risk (urethra, bladder, and rectum) was found while improving dose homogeneity or dose conformity of the target volume. Conclusions: Following the promising results of gEUD-based optimization in intensity modulated radiation therapy treatment optimization, as reported in the literature, the implementation of a similar model in HDR brachytherapy treatment plan optimization is suggested by this study. The potential of improved sparing of organs at risk was shown for various gEUD-based optimization parameter protocols, which indicates the ability of this method to adapt to the user's preferences.
  •  
26.
  • Gotz, Malte, et al. (författare)
  • Correction for volume recombination in liquid ionization chambers at high dose-per-pulse
  • 2019
  • Ingår i: Medical physics (Lancaster). - : WILEY. - 0094-2405 .- 2473-4209. ; 46:8, s. 3692-3699
  • Tidskriftsartikel (refereegranskat)abstract
    • Purpose To determine the volume recombination at high dose-per-pulse in liquid ionization chambers (LIC) and to ascertain whether existing calculation methods verified in air-filled chambers may be used to calculate a correction factor. Methods Two LICs, one filled with 2,2,4-trimethylpentane (isooctane) the other with tetramethylsilane (TMS), were irradiated in a pulsed, 20 MeV electron beam. Via reference measurements with a Faraday cup, the saturation correction for volume recombination was determined for dose-per-pulse values ranging from about 5 mGy to 1 Gy for both chambers at a pulse duration of 693 ns. In addition, the isooctane chamber was irradiated with pulses of varying duration, ranging from 5 ps to 10 ms, at a dose-per-pulse of about 76.5 mGy. The dose-per-pulse-dependent measurements were compared to calculations based on Boag's models (with and without a free electron fraction), the two-dose-rate method, and a numerical calculation. The pulse duration dependent measurements were compared only to a numerical calculation that iteratively calculates the charge transport and loss in a 1D model of an ionization chamber. Results In TMS only Boag's model with a free electron fraction and the numerical calculation are in good agreement with the experimental data. However, in isooctane, good agreement is observed between the experimental data, the numerical calculation as well as the two-dose-rate method, and Boag's model including a free electron fraction. Only Boag's model without a free electron fraction shows a good agreement with lesser extend. Furthermore, the pulse duration-dependent data for isooctane are well described by the numerical model. Conclusion With isooctane as an active medium, a LIC could be directly used in a field with high dose-per-pulse utilizing the well-established two-dose-rate method to correct the volume recombination. In addition, pulsed fields with variable pulse duration are easily modeled for this medium using a numerical calculation. Other media, as exemplified by the TMS-filled chamber, might require additional considerations, such as including a fraction of free electrons in the consideration of volume recombination.
  •  
27.
  • Götstedt, Julia, et al. (författare)
  • Development and evaluation of aperture-based complexity metrics using film and EPID measurements of static MLC openings.
  • 2015
  • Ingår i: Medical physics. - : Wiley. - 2473-4209 .- 0094-2405. ; 42:7, s. 3911-21
  • Tidskriftsartikel (refereegranskat)abstract
    • Complexity metrics have been suggested as a complement to measurement-based quality assurance for intensity modulated radiation therapy (IMRT) and volumetric modulated arc therapy (VMAT). However, these metrics have not yet been sufficiently validated. This study develops and evaluates new aperture-based complexity metrics in the context of static multileaf collimator (MLC) openings and compares them to previously published metrics.This study develops the converted aperture metric and the edge area metric. The converted aperture metric is based on small and irregular parts within the MLC opening that are quantified as measured distances between MLC leaves. The edge area metric is based on the relative size of the region around the edges defined by the MLC. Another metric suggested in this study is the circumference/area ratio. Earlier defined aperture-based complexity metrics-the modulation complexity score, the edge metric, the ratio monitor units (MU)/Gy, the aperture area, and the aperture irregularity-are compared to the newly proposed metrics. A set of small and irregular static MLC openings are created which simulate individual IMRT/VMAT control points of various complexities. These are measured with both an amorphous silicon electronic portal imaging device and EBT3 film. The differences between calculated and measured dose distributions are evaluated using a pixel-by-pixel comparison with two global dose difference criteria of 3% and 5%. The extent of the dose differences, expressed in terms of pass rate, is used as a measure of the complexity of the MLC openings and used for the evaluation of the metrics compared in this study. The different complexity scores are calculated for each created static MLC opening. The correlation between the calculated complexity scores and the extent of the dose differences (pass rate) are analyzed in scatter plots and using Pearson's r-values.The complexity scores calculated by the edge area metric, converted aperture metric, circumference/area ratio, edge metric, and MU/Gy ratio show good linear correlation to the complexity of the MLC openings, expressed as the 5% dose difference pass rate, with Pearson's r-values of -0.94, -0.88, -0.84, -0.89, and -0.82, respectively. The overall trends for the 3% and 5% dose difference evaluations are similar.New complexity metrics are developed. The calculated scores correlate to the complexity of the created static MLC openings. The complexity of the MLC opening is dependent on the penumbra region relative to the area of the opening. The aperture-based complexity metrics that combined either the distances between the MLC leaves or the MLC opening circumference with the aperture area show the best correlation with the complexity of the static MLC openings.
  •  
28.
  •  
29.
  •  
30.
  • Holmes, Robin B., et al. (författare)
  • Creation of an anthropomorphic CT head phantom for verification of image segmentation
  • 2020
  • Ingår i: Medical physics (Lancaster). - : Wiley-Blackwell. - 0094-2405 .- 2473-4209. ; 47:6, s. 2380-2391
  • Tidskriftsartikel (refereegranskat)abstract
    • Purpose: Many methods are available to segment structural magnetic resonance (MR) images of the brain into different tissue types. These have generally been developed for research purposes but there is some clinical use in the diagnosis of neurodegenerative diseases such as dementia. The potential exists for computed tomography (CT) segmentation to be used in place of MRI segmentation, but this will require a method to verify the accuracy of CT processing, particularly if algorithms developed for MR are used, as MR has notably greater tissue contrast.Methods: To investigate these issues we have created a three-dimensional (3D) printed brain with realistic Hounsfield unit (HU) values based on tissue maps segmented directly from an individual T1 MRI scan of a normal subject. Several T1 MRI scans of normal subjects from the ADNI database were segmented using SPM12 and used to create stereolithography files of different tissues for 3D printing. The attenuation properties of several material blends were investigated, and three suitable formulations were used to print an object expected to have realistic geometry and attenuation properties. A skull was simulated by coating the object with plaster of Paris impregnated bandages. Using two CT scanners, the realism of the phantom was assessed by the measurement of HU values, SPM12 segmentation and comparison with the source data used to create the phantom.Results: Realistic relative HU values were measured although a subtraction of 60 was required to obtain equivalence with the expected values (gray matter 32.9-35.8 phantom, 29.9-34.2 literature). Segmentation of images acquired at different kVps/mAs showed excellent agreement with the source data (Dice Similarity Coefficient 0.79 for gray matter). The performance of two scanners with two segmentation methods was compared, with the scanners found to have similar performance and with one segmentation method clearly superior to the other.Conclusion: The ability to use 3D printing to create a realistic (in terms of geometry and attenuation properties) head phantom has been demonstrated and used in an initial assessment of CT segmentation accuracy using freely available software developed for MRI.
  •  
Skapa referenser, mejla, bekava och länka
  • Resultat 21-30 av 346
Typ av publikation
tidskriftsartikel (316)
konferensbidrag (25)
forskningsöversikt (4)
doktorsavhandling (1)
Typ av innehåll
refereegranskat (299)
övrigt vetenskapligt/konstnärligt (47)
Författare/redaktör
Karlsson, Mikael (19)
Andreo, P (19)
Carlsson Tedgren, Ås ... (17)
Ahnesjö, Anders (16)
Danielsson, Mats (15)
Poludniowski, G (14)
visa fler...
Maguire Jr., Gerald ... (13)
Noz, Marilyn E. (13)
Papanikolaou, N (13)
Knöös, Tommy (11)
Nyholm, Tufve (11)
Mavroidis, Panayioti ... (10)
Mavroidis, P (10)
Alm Carlsson, Gudrun (10)
Ceberg, Crister (10)
Båth, Magnus, 1974 (9)
Ljungberg, Michael (7)
Ahnesjö, Anders, 195 ... (7)
Toma-Daşu, Iuliana (7)
Bujila, R (7)
Bernhardt, Peter, 19 ... (7)
Brahme, Anders (6)
Bornefalk, Hans (6)
Traneus, Erik (6)
Huq, MS (6)
Palmans, H (6)
Seuntjens, J (6)
Lundqvist, Hans (5)
Carlsson Tedgren, Ås ... (5)
Nilsson, Per (5)
Zeleznik, Michael P. (5)
Andersson, Jonas, 19 ... (5)
Nowik, P. (5)
Bassler, Niels (5)
Gutierrez, A. (4)
Brahme, A (4)
Verhaegen, Frank (4)
Petersson, Kristoffe ... (4)
Strand, Sven-Erik (4)
Weber, Lars (4)
Olsson, Lars E (4)
Sandborg, Michael (4)
Danielsson, Mats, Pr ... (4)
Kindblom, Jon, 1969 (4)
Timberg, Pontus (4)
Dasu, Alexandru (4)
Månsson, Lars Gunnar ... (4)
Mourtada, Firas (4)
Rivard, Mark J. (4)
Thomson, Rowan M. (4)
visa färre...
Lärosäte
Karolinska Institutet (114)
Kungliga Tekniska Högskolan (73)
Lunds universitet (57)
Uppsala universitet (42)
Linköpings universitet (40)
Umeå universitet (39)
visa fler...
Stockholms universitet (34)
Göteborgs universitet (25)
Chalmers tekniska högskola (3)
Örebro universitet (2)
Högskolan i Borås (1)
Sveriges Lantbruksuniversitet (1)
visa färre...
Språk
Engelska (343)
Odefinierat språk (3)
Forskningsämne (UKÄ/SCB)
Medicin och hälsovetenskap (137)
Naturvetenskap (92)
Teknik (43)

År

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy