SwePub
Sök i SwePub databas

  Utökad sökning

Träfflista för sökning "hsv:(MEDICIN OCH HÄLSOVETENSKAP) hsv:(Klinisk medicin) hsv:(Radiologi och bildbehandling) "

Sökning: hsv:(MEDICIN OCH HÄLSOVETENSKAP) hsv:(Klinisk medicin) hsv:(Radiologi och bildbehandling)

  • Resultat 5911-5920 av 7848
Sortera/gruppera träfflistan
   
NumreringReferensOmslagsbildHitta
5911.
  • Olander, Birger (författare)
  • Centre of Rotation Determination Using Projection Data in X-ray Micro Computed Tomography
  • 1994
  • Rapport (övrigt vetenskapligt/konstnärligt)abstract
    • There are several methods available to determine the Centre Of Rotation, COR, and align detectors and X-ray focus to COR in X-ray computed tomography. Some methods use narrow rods/needles or specially made alignment objects or phantoms. In X-ray Micro Computed Tomography (or Computerized Micro Tomography), μCT (CMT), methods using sample projection data for COR measurements are preferred since the replacement of alignment objects with samples often displace translation stages and make the alignment obsolete. To achieve an optimal image quality, precise positioning of COR to the detector and X-ray focus is essential. In μCT this can be accomplished in an alignment procedure using sample projection data prior to scanning. This alignment procedure will add examination time and increase the dose to the sample. Therefore the alignment procedure should incorporate as few projections as possible and be insensitive to noise. Some scanning equipment cannot be modified to implement such alignment procedure but actual COR can be determined from projection data and used in reconstruction. This report introduces a new Translated Opposite Projection, TOP, technique using a pair of opposite parallel projections (180° apart). Two TOP methods are developed: TOPlin using linear interpolation in the spatial domain and TOPfft in the frequency domain. The two TOP methods are compared to two Centre Of Gravity, COG, methods. The two COG methods are: COGsin, an enhancement of a method presented by Hogan et al [Hogan93] and COGopp, a simplification of this method possible with a fixed COR. In this report all projections in one scan are assumed to have a fixed, COR, as in third (or higher) generation tomography or first (and second generation) tomography if the translation stage errors is negligible. This also means that the rotation stage errors must be negligible. The COGsin is the only method presented here capable of determine a COR for each projection angle, thus allowing for a COR moving as a function of projection angle. The TOP methods normally give better precision with non-ideal projection data compared to the COG methods. Tests using both simulated and scanned projection data indicate that the TOP methods give higher precision in presence of stochastic errors (noise) and system errors like calibration errors. A μCT scan often takes a long time and detector calibration and X-ray intensity profiles might vary with time giving non-stationary system errors during a single scan. If the system errors can be approximated with simple polynomial functions, a new Baseline Restoration, BR, technique can be used together with the TOP methods to reduce COR errors.
  •  
5912.
  • Olivecrona, Grim, et al. (författare)
  • Assessment of the sacroiliac joint with computed tomography motion analysis : a diagnostic study of 12 patients
  • 2024
  • Ingår i: Acta Orthopaedica. - : Medical Journals Sweden. - 1745-3674 .- 1745-3682. ; 95, s. 20-24
  • Tidskriftsartikel (refereegranskat)abstract
    • Background and purpose - Pain in the sacroiliac joint may be caused by abnormal joint motion. Diagnosis is mainly based on clinical tests. The aims of this study were to examine whether low-dose computed tomography with provocation of the hip could detect sacroiliac joint motion , to study whether provocation of the hip results in greater sacroiliac joint motion in the ipsilateral than in the contralateral sacroiliac joint.Patients and methods - 12 patients with sacroiliac joint pain were examined with low-dose computed tomography scans of the sacroiliac joint, one with the hips in the neutral position, and one each in provocation with the left or the right hip in a figure-of-4 position. Accuracy was tested by comparing internal rotation of the sacrum with internal rotation in the sacroiliac joint. Motion in the sacroiliac joint was assessed by comparing the position of each of the ilia with the reference, the sacrum. Data is shown as mean with 95% confidence interval (CI).Results - We observed greater motion in the sacroiliac joint than internally in the sacrum, i.e., 0.57(degrees) (CI 0.43-0.71) vs. 0.20(degrees) (CI 0.11-0.28). The motion of the geometric center of the moving object for the sacroiliac joint was larger on the provoked side; mean difference 0.17 mm (CI 0.01-0.33), P = 0.04. Corresponding figures for rotation were mean difference 0.19(degrees) (CI 0.10-0.28), P < 0.001. Compared with the sacrum, the largest motion was seen at the anterior superior iliac spine; mean difference 0.38 mm (CI 0.10-0.66), P = 0.001.Conclusion - Provocation in the figure-of-4 position of the hip results in sacroiliac joint motion measurable with computed tomography motion analysis. Provocation of the hip induces larger motion on the ipsilateral than on the contralateral sacroiliac joint.
  •  
5913.
  • Olivecrona, Lotta, et al. (författare)
  • Standard orientation of the pelvis : Validation on a model and ten patients
  • 2005
  • Ingår i: Acta Radiologica. - London, UK : Informa Healthcare. - 0284-1851 .- 1600-0455. ; 46:1, s. 74-82
  • Tidskriftsartikel (refereegranskat)abstract
    • Purpose: To validate an image post-processing method for re-orienting the pelvis in CT volumes to a standardized orientation in a model and in 10 patients. Material and Methods: Twenty-four CT volumes of a pelvic model and 10 pairs of postoperative total hip arthroplasty (THA) patient CT scans were rotated to a defined pelvic standard orientation and the rotation was recorded. For precision, a test-retest procedure was used. For accuracy, three exactly represented coordinate points were used. For clinical application, the standard orientation was used for calculating the direction of acetabular cup migration from a previous model study. Results: Precision of pelvic standard orientation, calculated as maximal directional error, was better than 1degrees in the model study and better than 1.5degrees in the patient study. Accuracy, expressed as angle between ideal and measured coordinate axes, was 0.1degrees for x, y, z axes. No measurable systematic errors were found. When applied to acetabular cup migration in the model, standardization of pelvic orientation had no significant effect on the measurements. Conclusion: Reorienting the pelvis during image post-processing was shown to be accurate. It enables measurements relative to the pelvis and minimizes the dependency of patient positioning.
  •  
5914.
  • Olofsson, Jörgen, 1970- (författare)
  • Developing and evaluating dose calculation models for verification of advanced radiotherapy
  • 2006
  • Doktorsavhandling (övrigt vetenskapligt/konstnärligt)abstract
    • A prerequisite for modern radiotherapy is the ability to accurately determine the absorbed dose (D) that is given to the patient. The subject of this thesis has been to develop and evaluate efficient dose calculation models for high-energy photon beams delivered by linear accelerators. Even though the considered calculation models are general, the work has been focused on quality assurance (QA) tools used to independently verify the dose for individual treatment plans. The purpose of this verification is to guarantee patient safety and to improve the treatment outcome. Furthermore, a vital part of this work has been to explore the prospect of estimating the dose calculation uncertainties associated with individual treatment setups. A discussion on how such uncertainty estimations can facilitate improved clinical QA procedures by providing appropriate action levels has also been included within the scope of this thesis. In order to enable efficient modelling of the physical phenomena that are involved in dose output calculations it is convenient to divide them into two main categories; the first one dealing with the radiation exiting the accelerator’s treatment head and a second one associated with the subsequent energy deposition processes. A multi-source model describing the distribution of energy fluence emitted from the treatment head per delivered monitor unit (MU) is presented and evaluated through comparisons with measurements in multiple photon beams and collimator settings. The calculations show close agreement with the extensive set of experimental data, generally within +/-1% of corresponding measurements. The energy (dose) deposition in the irradiated object has been modelled through a photon pencil kernel solely based on a beam quality index (TPR20,10). This model was evaluated in a similar manner as the multi-source model at three different treatment depths. A separate study was focused on the specific difficulties associated with dose calculations in points located at a distance from the central beam axis. Despite the minimal input data required to characterize individual photon beams, the accuracy proved to be very good when comparing the calculated results with experimental data. The evaluated calculation models were finally used to analyse how well the lateral dose distributions from typical megavoltage photon beams are optimized with respect to the resulting beam flatness characteristics. The results did not reveal any obvious reasons why different manufacturers should provide different lateral dose distributions. Furthermore, the performed lateral optimizations indicate that there is room for improved flatness performance for the investigated linear accelerators.
  •  
5915.
  • Olofsson, Jonas K., et al. (författare)
  • A Designated Odor-Language Integration System in the Human Brain
  • 2014
  • Ingår i: Journal of Neuroscience. - 0270-6474 .- 1529-2401. ; 34:45, s. 14864-14873
  • Tidskriftsartikel (refereegranskat)abstract
    • Odors are surprisingly difficult to name, but the mechanism underlying this phenomenon is poorly understood. In experiments using event-related potentials (ERPs) and functional magnetic resonance imaging (fMRI), we investigated the physiological basis of odor naming with a paradigm where olfactory and visual object cues were followed by target words that either matched or mismatched the cue. We hypothesized that word processing would not only be affected by its semantic congruency with the preceding cue, but would also depend on the cue modality (olfactory or visual). Performance was slower and less precise when linking a word to its corresponding odor than to its picture. The ERP index of semantic incongruity (N400), reflected in the comparison of nonmatching versus matching target words, was more constrained to posterior electrode sites and lasted longer on odor-cue (vs picture-cue) trials. In parallel, fMRI cross-adaptation in the right orbitofrontal cortex (OFC) and the left anterior temporal lobe (ATL) was observed in response to words when preceded by matching olfactory cues, but not by matching visual cues. Time-series plots demonstrated increased fMRI activity in OFC and ATL at the onset of the odor cue itself, followed by response habituation after processing of a matching (vs nonmatching) target word, suggesting that predictive perceptual representations in these regions are already established before delivery and deliberation of the target word. Together, our findings underscore the modality-specific anatomy and physiology of object identification in the human brain.
  •  
5916.
  • Olofsson, Lennart, 1952- (författare)
  • Energy and intensity modulated radiation therapy with electrons
  • 2005
  • Doktorsavhandling (övrigt vetenskapligt/konstnärligt)abstract
    • In recent years intensity modulated radiation therapy with photons (xIMRT) has gained attention due to its ability to reduce the dose in the tissues close to the tumour volume. However, this technique also results in a large low dose volume. Electron IMRT (eIMRT) has the potential to reduce the integral dose to the patient due to the dose fall off in the electron depth dose curves. This dose fall off makes it possible to modulate the dose distribution in the direction of the beam by selecting appropriate electron energies. The use of a computer based energy selection method was examined in combination with the IMRT technique to optimise the electron dose distribution. It is clearly illustrated that the energy optimisation procedure reduces the dose to lung and heart in a breast cancer treatment. To shape the multiple electron subfields (beamlets) that are used in eIMRT, an electron multi leaf collimator (eMLC) is needed. However, photons produced in a conventional electron treatment head could penetrate such an added eMLC, thus producing an undesirable dose contribution. The leakage levels normally achieved are acceptable for standard single electron field treatments but could become unacceptably high in eIMRT treatments where a lot of small subfields are combined. To limit this photon contribution, the photon MLC (xMLC) was used to shield off large parts of the photon leakage. The effect of this xMLC shielding on the reduction of photon leakage, the electron beam penumbras, and electron output (dose level), was studied using Monte Carlo methods for different electron treatment head designs. The use of helium as a mean to reduce the electron scatter in the treatment head, and thus the perturbating effect of the xMLC on electron beam penumbra and output, was also investigated. This thesis shows that the effect of the xMLC shielding on the electron beam penumbra and output can be made negligible while still obtaining a significantly reduced x-ray leakage dose contribution. The result is a large gain in radiation protection of the patient and a better dynamic range for the eIMRT dose optimisation. For this optimisation a computer based electron energy selection method was developed and tested on two clinical cases.
  •  
5917.
  •  
5918.
  • Olsrud, Johan, et al. (författare)
  • A two-compartment gel phantom for optimization and quality assurance in clinical BOLD fMRI
  • 2008
  • Ingår i: Magnetic Resonance Imaging. - : Elsevier BV. - 1873-5894 .- 0730-725X. ; 26:2, s. 279-286
  • Tidskriftsartikel (refereegranskat)abstract
    • Clinical applications of blood oxygen level-dependent (BOLD) functional magnetic resonance imaging (fMRI) depend heavily on robust paradigms, imaging methods and analysis procedures. In this work, as a means to optimize and perform quality assurance of the entire imaging and analysis chain, a phantom that provides a well known and reproducible signal change similar to a block type fMRI experiment is presented. It consists of two gel compartments with slightly different T(2) that dynamically enter and leave the imaged volume. The homogeneous gel in combination with a cylindrical geometry results in a well-defined T(2) difference causing a signal difference between the two compartments in T(2)-weighted MR images. From time series data obtained with the phantom, maps of percent signal change (PSC) and t-values are calculated. As an example of image parameter optimisation, the phantom is demonstrated to be useful for accurate determination of the influence of echo time (TE) on BOLD fMRI results, taking the t-value as a measure of sensitivity. In addition, the phantom is proposed as a tool for quality assurance (QA) since reproducible time series and t-maps are obtained in a series of independent repeat experiments. The phantom is relatively simple to build and can therefore be used by any clinical fMRI center.
  •  
5919.
  •  
5920.
  • Olsrud, Johan, et al. (författare)
  • MRI thermometry in phantoms by use of the proton resonance frequency shift method: application to interstitial laser thermotherapy
  • 1998
  • Ingår i: Physics in Medicine and Biology. - 1361-6560. ; 43:9, s. 2597-2613
  • Tidskriftsartikel (refereegranskat)abstract
    • In this work the temperature dependence of the proton resonance frequency was assessed in agarose gel with a high melting temperature (95 degrees C) and in porcine liver in vitro at temperatures relevant to thermotherapy (25-80 degrees C). Furthermore, an optically tissue-like agarose gel phantom was developed and evaluated for use in MRI. The phantom was used to visualize temperature distributions from a diffusing laser fibre by means of the proton resonance frequency shift method. An approximately linear relationship (0.0085 ppm degrees C(-1)) between proton resonance frequency shift and temperature change was found for agarose gel, whereas deviations from a linear relationship were observed for porcine liver. The optically tissue-like agarose gel allowed reliable MRI temperature monitoring, and the MR relaxation times (T1 and T2) and the optical properties were found to be independently alterable. Temperature distributions around a diffusing laser fibre, during irradiation and subsequent cooling, were assessed with high spatial resolution (voxel size = 4.3 mm3) and with random uncertainties ranging from 0.3 degrees C to 1.4 degrees C (1 SD) with a 40 s scan time.
  •  
Skapa referenser, mejla, bekava och länka
  • Resultat 5911-5920 av 7848
Typ av publikation
tidskriftsartikel (5725)
konferensbidrag (1105)
doktorsavhandling (394)
forskningsöversikt (210)
bokkapitel (134)
annan publikation (100)
visa fler...
rapport (92)
patent (35)
bok (15)
licentiatavhandling (14)
recension (10)
samlingsverk (redaktörskap) (9)
proceedings (redaktörskap) (5)
konstnärligt arbete (1)
visa färre...
Typ av innehåll
refereegranskat (6234)
övrigt vetenskapligt/konstnärligt (1559)
populärvet., debatt m.m. (51)
Författare/redaktör
Forssell-Aronsson, E ... (290)
Båth, Magnus, 1974 (282)
Mattsson, Sören (177)
Tolmachev, Vladimir (173)
Lubberink, Mark (170)
Ljungberg, Michael (142)
visa fler...
Tingberg, Anders (129)
Noz, Marilyn E. (126)
Maguire Jr., Gerald ... (122)
Månsson, Lars Gunnar ... (120)
Ståhlberg, Freddy (116)
Ekberg, Olle (115)
Valentin, Lil (113)
Larsson, Elna-Marie (110)
Isaksson, Mats, 1961 (104)
Strand, Sven-Erik (103)
Sörensen, Jens (102)
Helou, Khalil, 1966 (100)
Spetz, Johan (100)
Holtås, Stig (98)
Wirestam, Ronnie (95)
Bernhardt, Peter, 19 ... (93)
Nilsson, Markus (92)
Zackrisson, Sophia (89)
Langen, Britta (88)
Knutsson, Linda (81)
Ljungberg, Maria (79)
Johnsson, Åse (Allan ... (79)
Nilsson, Per (75)
van Westen, Danielle (75)
Orlova, Anna, 1960- (73)
Svalkvist, Angelica (72)
Sundgren, Pia (71)
Nyman, Ulf (71)
Orlova, Anna (70)
Nyholm, Tufve (68)
Trägårdh, Elin (67)
Leander, Peter (66)
Söderberg, Marcus (64)
Ahlström, Håkan, 195 ... (63)
Björkman-Burtscher, ... (63)
Rääf, Christopher (62)
Antoni, Gunnar (61)
Arheden, Håkan (61)
Andersson, Ingvar (61)
Palm, Stig, 1964 (60)
Mitran, Bogdan (58)
Ceberg, Crister (58)
Szczepankiewicz, Fil ... (58)
Haller, Sven (58)
visa färre...
Lärosäte
Lunds universitet (3389)
Göteborgs universitet (1586)
Uppsala universitet (1415)
Karolinska Institutet (799)
Linköpings universitet (692)
Kungliga Tekniska Högskolan (610)
visa fler...
Umeå universitet (520)
Stockholms universitet (244)
Chalmers tekniska högskola (213)
Örebro universitet (156)
Malmö universitet (48)
Sveriges Lantbruksuniversitet (34)
Jönköping University (26)
Linnéuniversitetet (17)
Mittuniversitetet (15)
Luleå tekniska universitet (12)
RISE (11)
Blekinge Tekniska Högskola (8)
Högskolan i Skövde (7)
Högskolan i Halmstad (6)
Högskolan Kristianstad (4)
Handelshögskolan i Stockholm (4)
Gymnastik- och idrottshögskolan (4)
Högskolan i Gävle (3)
Mälardalens universitet (3)
Högskolan Väst (2)
Karlstads universitet (2)
Marie Cederschiöld högskola (2)
Högskolan i Borås (1)
Högskolan Dalarna (1)
Sophiahemmet Högskola (1)
visa färre...
Språk
Engelska (7648)
Svenska (177)
Tyska (12)
Japanska (3)
Danska (2)
Franska (1)
visa fler...
Ryska (1)
Spanska (1)
Finska (1)
Grekiska, klassisk (1)
Kinesiska (1)
visa färre...
Forskningsämne (UKÄ/SCB)
Medicin och hälsovetenskap (7847)
Naturvetenskap (629)
Teknik (471)
Samhällsvetenskap (69)
Lantbruksvetenskap (30)
Humaniora (11)

År

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy