SwePub
Tyck till om SwePub Sök här!
Sök i SwePub databas

  Utökad sökning

Träfflista för sökning "swepub ;pers:(Larsson Anders);mspu:(article)"

Sökning: swepub > Larsson Anders > Tidskriftsartikel

  • Resultat 651-660 av 1282
Sortera/gruppera träfflistan
   
NumreringReferensOmslagsbildHitta
651.
  • Karagiannidis, Christian, et al. (författare)
  • Impact of membrane lung surface area and blood flow on extracorporeal CO2 removal during severe respiratory acidosis
  • 2017
  • Ingår i: Intensive Care Medicine Experimental. - : Springer Science and Business Media LLC. - 2197-425X. ; 5
  • Tidskriftsartikel (refereegranskat)abstract
    • BACKGROUND: Veno-venous extracorporeal CO2 removal (vv-ECCO2R) is increasingly being used in the setting of acute respiratory failure. Blood flow rates through the device range from 200 ml/min to more than 1500 ml/min, and the membrane surface areas range from 0.35 to 1.3 m2. The present study in an animal model with similar CO2 production as an adult patient was aimed at determining the optimal membrane lung surface area and technical requirements for successful vv-ECCO2R.METHODS: Four different membrane lungs, with varying lung surface areas of 0.4, 0.8, 1.0, and 1.3m2 were used to perform vv-ECCO2R in seven anesthetized, mechanically ventilated, pigs with experimentally induced severe respiratory acidosis (pH 7.0-7.1) using a 20Fr double-lumen catheter with a sweep gas flow rate of 8 L/min. During each experiment, the blood flow was increased stepwise from 250 to 1000 ml/min.RESULTS: Amelioration of severe respiratory acidosis was only feasible when blood flow rates from 750 to 1000 ml/min were used with a membrane lung surface area of at least 0.8 m2. Maximal CO2 elimination was 150.8 ml/min, with pH increasing from 7.01 to 7.30 (blood flow 1000 ml/min; membrane lung 1.3 m2). The membrane lung with a surface of 0.4 m2 allowed a maximum CO2 elimination rate of 71.7 mL/min, which did not result in the normalization of pH, even with a blood flow rate of 1000 ml/min. Also of note, an increase of the surface area above 1.0 m2 did not result in substantially higher CO2 elimination rates. The pressure drop across the oxygenator was considerably lower (<10 mmHg) in the largest membrane lung, whereas the smallest revealed a pressure drop of more than 50 mmHg with 1000 ml blood flow/min.CONCLUSIONS: In this porcine model, vv-ECCO2R was most effective when using blood flow rates ranging between 750 and 1000 ml/min, with a membrane lung surface of at least 0.8 m2. In contrast, low blood flow rates (250-500 ml/min) were not sufficient to completely correct severe respiratory acidosis, irrespective of the surface area of the membrane lung being used. The converse was also true, low surface membrane lungs (0.4 m2) were not capable of completely correcting severe respiratory acidosis across the range of blood flows used in this study.
  •  
652.
  •  
653.
  •  
654.
  • Karagiannidis, Christian, et al. (författare)
  • Veno-venous extracorporeal CO2 removal for the treatment of severe respiratory acidosis : pathophysiological and technical considerations
  • 2014
  • Ingår i: Critical Care. - : Springer Science and Business Media LLC. - 1364-8535 .- 1466-609X. ; 18:3, s. R124-
  • Tidskriftsartikel (refereegranskat)abstract
    • INTRODUCTION:While non-invasive ventilation aimed at avoiding intubation has become the modality of choice to treat mild to moderate acute respiratory acidosis, many severely acidotic patients (pH <7.20) still need intubation. Extracorporeal veno-venous CO2 removal (ECCO2R) could prove to be an alternative. The present animal study tested in a systematic fashion technical requirements for successful ECCO2R in terms of cannula size, blood and sweep gas flow.METHODS:ECCO2R with a 0.98 m2 surface oxygenator was performed in six acidotic (pH <7.20) pigs using either a 14.5 French (Fr) or a 19Fr catheter, with sweep gas flow rates of 8 and 16 L/minute, respectively. During each experiment the blood flow was incrementally increased to a maximum of 400 mL/minute (14.5Fr catheter) and 1000 mL/minute (19Fr catheter).RESULTS:Amelioration of severe respiratory acidosis was only feasible when blood flow rates of 750 to 1000 mL/minute (19Fr catheter) were used. Maximal CO2-elimination was 146.1 ± 22.6 mL/minute, while pH increased from 7.13 ± 0.08 to 7.41 ± 0.07 (blood flow of 1000 mL/minute; sweep gas flow 16 L/minute). Accordingly, a sweep gas flow of 8 L/minute resulted in a maximal CO2-elimination rate of 138.0 ± 16.9 mL/minute. The 14.5Fr catheter allowed a maximum CO2 elimination rate of 77.9 mL/minute, which did not result in the normalization of pH.CONCLUSIONS:Veno-venous ECCO2R may serve as a treatment option for severe respiratory acidosis. In this porcine model, ECCO2R was most effective when using blood flow rates ranging between 750 and 1000 mL/minute, while an increase in sweep gas flow from 8 to 16 L/minute had less impact on ECCO2R in this setting.
  •  
655.
  • Karawajczyk, Malgorzata, et al. (författare)
  • High expression of neutrophil and monocyte CD64 with simultaneous lack of upregulation of adhesion receptors CD11b, CD162, CD15, CD65 on neutrophils in severe COVID-19
  • 2021
  • Ingår i: Therapeutic advances in infectious disease. - : Sage Publications. - 2049-9361 .- 2049-937X. ; 8
  • Tidskriftsartikel (refereegranskat)abstract
    • Background and Aims: The pronounced neutrophilia observed in patients with coronavirus disease 2019 (COVID-19) infections suggests a role for these leukocytes in the pathology of the disease. Monocyte and neutrophil expression of CD64 and CD11b have been reported as early biomarkers to detect infections. The aim of this study was to study the expression of receptors for IgG (CD64) and adhesion molecules (CD11b, CD15s, CD65, CD162, CD66b) on neutrophils and monocytes in patients with severe COVID-19 after admission to an intensive care unit (ICU).Methods: The expression of receptors was analyzed using flow cytometry. EDTA blood from 23 patients with confirmed COVID-19 infection was sampled within 48 h of admission to the ICU. Leukocytes were labeled with antibodies to CD11b, CD15s, CD65s, CD162, CD64, and CD66b. Expression of receptors was reported as mean fluorescence intensity (MFI) or the percentage of cells expressing receptors.Results: Results are presented as comparison of COVID-19 patients with the healthy group and the receptor expression as MFI. Neutrophil receptors CD64 (2.5 versus 0.5) and CD66b (44.5 versus 34) were increased and CD15 decreased (21.6 versus 28.3) when CD65 (6.6 versus 4.4), CD162 (21.3 versus 21.1) and CD11b (10.5 versus 12) were in the same range. Monocytes receptors CD64 (30.5 versus 16.6), CD11b (18.7 versus 9.8), and CD162 (38.6 versus 36.5) were increased and CD15 decreased (10.3 versus 17.9); CD65 were in the same range (2.3 versus 1.96).Conclusion: Monocytes and neutrophils are activated during severe COVID-19 infection as shown by strong upregulation of CD64. High monocyte and neutrophil CD64 can be an indicator of a severe form of COVID19. The adhesion molecules (CD11b, CD162, CD65, and CD15) are not upregulated on otherwise activated neutrophils, which might lead to relative impairment of tissue migration. Low adhesion profile of neutrophils suggests immune dysfunction of neutrophils. Monocytes maintain upregulation of some adhesion molecules (CD11b, CD162) suggesting the persistence of an increased ability to migrate into tissues, even during a severe stage of COVID-19. Future research should focus on CD64 and CD11b kinetics in the context of prognosis.
  •  
656.
  •  
657.
  • Karawajczyk, Malgorzata, et al. (författare)
  • Reduced cystatin C-estimated GFR and increased creatinine-estimated GFR in comparison with iohexol-estimated GFR in a hyperthyroid patient : A case report
  • 2008
  • Ingår i: Journal of Medical Case Reports. - : Springer Science and Business Media LLC. - 1752-1947. ; 2:66
  • Tidskriftsartikel (övrigt vetenskapligt/konstnärligt)abstract
    • ABSTRACT: INTRODUCTION: Estimation of the glomerular filtration rate (GFR) is essential for the evaluation of patients with kidney disease, and for treating patients with drugs that are eliminated from the circulation by the kidneys. Cystatin C has been shown to be superior to creatinine for estimating GFR in several studies. However, studies showing that thyroid function has an impact on cystatin C have not addressed the question of whether the changes in cystatin C levels are due to changes in GFR or in cystatin C synthesis. CASE PRESENTATION: We report an account of a hyperthyroid patient with a discrepancy between the GFR estimates from cystatin C and creatinine. The cystatin C concentration (1.36 mg/L) was higher and gave an estimated GFR which was lower (51 mL/min/1.73 m2), while the creatinine concentration was lower (36 mumol/L) and gave a corresponding creatinine-estimated GFR that was higher (145 mL/min/1.73 m2) than the iohexol-estimated GFR (121 mL/min/1.73 m2) during the hyperthyroid period. After thyroidectomy, the creatinine concentration was 36 mumol/L and creatinine-estimated GFR was calculated as 73 mL/min/1.73 m2, while the cystatin C concentration and cystatin C-calculated GFR was 0.78 mg/L and 114 mL/min/1.73 m2, respectively. CONCLUSION: In contrast to creatinine, cystatin C levels rose in the hyperthyroid state as compared to the euthyroid state. The cystatin C-estimated GFR was reduced compared to the iohexol-estimated GFR. This patient case shows that the hyperthyroid-associated changes in cystatin C levels are not due to changes in GFR. Thyroid function should thus be considered when both cystatin C and creatinine are used as markers of kidney function.
  •  
658.
  • Karawajczyk, Malgorzata, et al. (författare)
  • The HemoCue WBC DIFF system could be used for leucocyte and neutrophil counts but not for full differential counts
  • 2017
  • Ingår i: Acta Paediatrica. - : Wiley. - 0803-5253 .- 1651-2227. ; 106:6, s. 974-978
  • Tidskriftsartikel (refereegranskat)abstract
    • AIM: The aim of this study was to evaluate the HemoCue WBC DIFF system for point of care testing of fingerstick samples from paediatric patients.METHODS: We analysed 158 white blood cell counts on both the point of care HemoCue WBC DIFF instrument and the Cell Dyn Sapphire cell counter used by our central laboratory and compared the results. The measurements were performed using fingerstick samples drawn by nurses working in paediatric emergency and paediatric oncology units.RESULTS: There was good agreement between the two instruments for white blood cell and neutrophil counts. The correlation was weaker for lymphocytes and the correlations were poor for monocytes and eosinophils. The HemoCue WBC DIFF flagged 56 of the 148 capillary drawn samples as abnormal, but none of the 10 venously collected samples. Only two of the flagged samples differed significantly between the instruments, with regard to the cell counts.CONCLUSION: The correlations between the white blood cell counts and neutrophil counts in this real life study were good enough to diagnose children in emergency department and oncology unit settings. However, the high number of pathological flags from fingerstick samples, which made reruns necessary, limited the usefulness of the instrument.
  •  
659.
  • Karawajczyk, Malgorzata, et al. (författare)
  • The Near-Patient Testing Instrument HemoScreen can be used for Rapid Evaluation of Peripheral Blood Cell Counts in Patients with COVID-19
  • 2023
  • Ingår i: Family Medicine and Primary Care: Open Access. - : Gavin Publishers. - 2688-7460. ; 7:1, s. 1-7
  • Tidskriftsartikel (refereegranskat)abstract
    • Objectives: Analysis of peripheral blood cell counts are important in the workup of patients with suspected Corona virus disease 2019 (COVID-19). Blood cell counts are usually analyzed on large cell counters that require transport of the samples. Point-of-care (POC) testing reduces the test-turnaround times. The aim of the present study was to evaluate if the HemoScreen™instrument could be used for point of care testing of blood samples from COVID-19 patients. Methods: We compared CBC results from 49 patients with COVID-19 infections analyzed with HemoScreen™ and Sysmex XN-9000. Results: There were strong correlations between the results obtained with Sysmex XN-9000 and HemoScreen™ according to Deming correlations for White Blood Cells (WBC): WBCHemoScreen™ = 0.944* WBCSysmex - 0.093; r2 = 0.988, Neutrophils: NeutrophilsHemoScreen™= 1.029* NeutrophilsSysmex + 0.090; r2 = 0.957, Lymphocytes: LymphocytesHemoScreen™ = 0.944* LymphocytesSysmex - 0.107; r2 = 0.964, Platelets (PLT): PLTHemoScreen™ = 1.056* PLTSysmex + 9.468; r2 = 0.988, and Red blood cells (RBC) RBCHemoScreen™ = 0.955* RBCSysmex - 0.003; r2 = 0.991. Conclusions: The HemoScreen™ instrument provided fast and accurate test results when analyzing blood samples from COVID-19 patients. This instrument can be used in settings where the transport time to the central laboratory from remote locations such as field hospitals can be long.
  •  
660.
  •  
Skapa referenser, mejla, bekava och länka
  • Resultat 651-660 av 1282
Typ av publikation
Typ av innehåll
refereegranskat (1165)
övrigt vetenskapligt/konstnärligt (108)
populärvet., debatt m.m. (9)
Författare/redaktör
Larsson, Anders, 195 ... (165)
Lipcsey, Miklós (96)
Gustavsson, Johan, 1 ... (82)
Lind, Lars (80)
Ärnlöv, Johan, 1970- (62)
visa fler...
Ärnlöv, Johan (62)
Westbergh, Petter, 1 ... (47)
Wang, Shu Min, 1963 (46)
Basu, Samar (44)
Sadeghi, Mahdad, 196 ... (42)
Hedenstierna, Göran (40)
Hedenstierna, Göran, ... (38)
Haglund, Åsa, 1976 (35)
Carlsson, Axel C. (35)
Eriksson, Mats (34)
Sjölin, Jan (33)
Perchiazzi, Gaetano (31)
Frithiof, Robert (30)
Ronquist, Gunnar (30)
Sundström, Johan (30)
Dandona, L (29)
Dandona, R (29)
Yonemoto, N (28)
Carlsson, Lena (27)
Ingelsson, Erik (27)
Gupta, R. (26)
Larsson, Per Anders (26)
Alvis-Guzman, N (25)
Kosen, S (25)
Monasta, L (25)
Venketasubramanian, ... (25)
Vos, T (25)
Fischer, F (24)
Malekzadeh, R (24)
Meretoja, A (24)
Naghavi, M (24)
Abd-Allah, F (23)
Catala-Lopez, F (23)
Mohammed, S (23)
Nangia, V (23)
Bengtsson, Jörgen, 1 ... (23)
Hultström, Michael, ... (23)
Bedi, N (22)
Bikbov, B (22)
Farzadfar, F (22)
Majeed, A (22)
Mendoza, W (22)
Westerman, R (22)
Risérus, Ulf (22)
visa färre...
Lärosäte
Uppsala universitet (996)
Karolinska Institutet (261)
Chalmers tekniska högskola (180)
Högskolan Dalarna (109)
Lunds universitet (105)
Göteborgs universitet (64)
visa fler...
Linköpings universitet (48)
Örebro universitet (33)
Umeå universitet (31)
Sveriges Lantbruksuniversitet (21)
Kungliga Tekniska Högskolan (20)
Luleå tekniska universitet (20)
Stockholms universitet (14)
Mittuniversitetet (12)
RISE (11)
Södertörns högskola (6)
Högskolan i Gävle (5)
Malmö universitet (2)
Linnéuniversitetet (2)
Karlstads universitet (2)
Högskolan i Halmstad (1)
Högskolan i Skövde (1)
Riksantikvarieämbetet (1)
visa färre...
Språk
Engelska (1227)
Svenska (31)
Odefinierat språk (23)
Norska (1)
Forskningsämne (UKÄ/SCB)
Medicin och hälsovetenskap (673)
Teknik (184)
Naturvetenskap (127)
Samhällsvetenskap (33)
Lantbruksvetenskap (13)
Humaniora (3)

År

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy