Sökning: onr:"swepub:oai:lup.lub.lu.se:844cb989-59eb-4d49-b27e-d44a08872878" >
Use of an enhanced ...
Use of an enhanced gradient system for diffusion MR imaging with motion-artifact reduction
-
- Brockstedt, Sara (författare)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
-
Thomsen, C (författare)
-
- Wirestam, Ronnie (författare)
- Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
-
visa fler...
-
De Poorter, J (författare)
-
De Wagter, C (författare)
-
- Salford, Leif (författare)
- Lund University,Lunds universitet,Neurokirurgi,Sektion IV,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Neurosurgery,Section IV,Department of Clinical Sciences, Lund,Faculty of Medicine
-
- Holtås, Stig (författare)
- Lund University,Lunds universitet,Diagnostisk radiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Diagnostic Radiology, (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
-
- Ståhlberg, Freddy (författare)
- Lund University,Lunds universitet,Diagnostisk radiologi, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medicinsk strålningsfysik, Lund,Institutionen för kliniska vetenskaper, Lund,Diagnostic Radiology, (Lund),Section V,Department of Clinical Sciences, Lund,Faculty of Medicine,Medical Radiation Physics, Lund,Department of Clinical Sciences, Lund
-
visa färre...
-
(creator_code:org_t)
- 1995
- 1995
- Engelska.
-
Ingår i: Acta Radiologica. - 1600-0455. ; 36:6, s. 662-670
- Relaterad länk:
-
https://lup.lub.lu.s...
Abstract
Ämnesord
Stäng
- PURPOSE: A spin-echo diffusion-sensitized pulse sequence using high gradients (23 mT/m) is introduced. MATERIAL AND METHODS: In order to minimize motion artefacts, velocity-compensating gradients, ECG-triggering and post-processing with phase correction and raw data averaging using navigator echoes was performed. The in vitro ratio of diffusion coefficients for water and acetone was determined and the water self-diffusion coefficient at different temperatures was evaluated. The pulse sequence was tested in 7 healthy volunteers and in 2 tumour patients with astrocytomas of grades I-II and III-IV. Both single-slice and multi-slice techniques were used. RESULTS: The incorporation of phase correction clearly improved the quality of both diffusion-encoded images and the calculated diffusion maps. Mean values of the diffusion coefficients in vivo were for CSF 2.66 x 10(-9) m2/s and for white and grey matter 0.69 x 10(-9) m2/s and 0.87 x 10(-9) m2/s, respectively. CONCLUSION: Velocity-compensating gradients in combination with a high gradient strength were shown to be useful for in vivo diffusion MR imaging.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
Publikations- och innehållstyp
- art (ämneskategori)
- ref (ämneskategori)
Hitta via bibliotek
Till lärosätets databas