SwePub
Sök i LIBRIS databas

  Utökad sökning

(id:"swepub:oai:DiVA.org:umu-111468")
 

Sökning: (id:"swepub:oai:DiVA.org:umu-111468") > Fast 4D flow MRI in...

Fast 4D flow MRI intracranial segmentation and quantification in tortuous arteries

Schrauben, Eric (författare)
Umeå universitet,Institutionen för strålningsvetenskaper,Wisconsin, USA
Wåhlin, Anders (författare)
Umeå universitet,Radiofysik,Umeå centrum för funktionell hjärnavbildning (UFBI),Department of Radiology, University of Wisconsin – Madison, Madison, Wisconsin, USA
Ambarki, Khalid (författare)
Umeå universitet,Institutionen för strålningsvetenskaper,Centrum för medicinsk teknik och fysik (CMTF)
visa fler...
Spaak, Erik (författare)
Umeå universitet,Institutionen för strålningsvetenskaper
Malm, Jan (författare)
Umeå universitet,Klinisk neurovetenskap
Wieben, Oliver (författare)
University of Wisconsin – Madison, Madison, Wisconsin, USA
Eklund, Anders (författare)
Umeå universitet,Centrum för medicinsk teknik och fysik (CMTF),Radiofysik
visa färre...
 (creator_code:org_t)
2015-04-02
2015
Engelska.
Ingår i: Journal of Magnetic Resonance Imaging. - : John Wiley & Sons. - 1053-1807 .- 1522-2586. ; 42:5, s. 1458-1464
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • PurposeTo describe, validate, and implement a centerline processing scheme (CPS) for semiautomated segmentation and quantification in carotid siphons of healthy subjects. 4D flow MRI enables blood flow measurement in all major cerebral arteries with one scan. Clinical translational hurdles are time demanding postprocessing and user-dependence induced variability during analysis. Materials and MethodsA CPS for 4D flow data was developed to automatically separate cerebral artery trees. Flow parameters were quantified at planes along the centerline oriented perpendicular to the vessel path. At 3T, validation against 2D phase-contrast (PC) magnetic resonance imaging (MRI) and 4D flow manual processing was performed on an intracranial flow phantom for constant flow, while pulsatile flow validation was performed in the internal carotid artery (ICA) of 10 healthy volunteers. The CPS and 4D manual processing times were measured and compared. Flow and area measurements were also demonstrated along the length of the ICA siphon. ResultsPhantom measurements for area and flow were highly correlated between the CPS and 2D measurements (area: R=0.95, flow: R=0.94), while in vivo waveforms were highly correlated (R=0.93). Processing time was reduced by a factor of 4.6 compared with manual processing. Whole ICA measurements revealed a significantly decreased area in the most distal segment of the carotid siphon (P=0.0017), with flow unchanged (P=0.84). ConclusionThis study exhibits fast semiautomated analysis of intracranial 4D flow MRI. Internal consistency was shown through flow conservation along the tortuous ICA siphon, which is typically difficult to assess. J. Magn. Reson. Imaging 2015;42:1458-1464.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

4D flow MRI
vascular
segmentation
automated processing

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy