SwePub
Sök i SwePub databas

  Utökad sökning

Träfflista för sökning "L773:0094 2405 OR L773:2473 4209 ;pers:(Benmakhlouf Hamza)"

Sökning: L773:0094 2405 OR L773:2473 4209 > Benmakhlouf Hamza

  • Resultat 1-5 av 5
Sortera/gruppera träfflistan
   
NumreringReferensOmslagsbildHitta
1.
  • Benmakhlouf, Hamza, et al. (författare)
  • Output correction factors for nine small field detectors in 6 MV radiation therapy photon beams : A PENELOPE Monte Carlo study
  • 2014
  • Ingår i: Medical physics (Lancaster). - : Wiley. - 0094-2405 .- 2473-4209. ; 41:4, s. 041711-
  • Tidskriftsartikel (refereegranskat)abstract
    • Purpose: To determine detector-specific output correction factors, k(Qclin,Qmsr)(fclin,fmsr) in 6 MV small photon beams for air and liquid ionization chambers, silicon diodes, and diamond detectors from two manufacturers. Methods: Field output factors, defined according to the international formalism published by Alfonso et al. [Med. Phys. 35, 5179-5186 (2008)], relate the dosimetry of small photon beams to that of the machine-specific reference field; they include a correction to measured ratios of detector readings, conventionally used as output factors in broad beams. Output correction factors were calculated with the PENELOPE Monte Carlo (MC) system with a statistical uncertainty (type-A) of 0.15% or lower. The geometries of the detectors were coded using blueprints provided by the manufacturers, and phase-space files for field sizes between 0.5 x 0.5 cm(2) and 10 x 10 cm(2) from a Varian Clinac iX 6 MV linac used as sources. The output correction factors were determined scoring the absorbed dose within a detector and to a small water volume in the absence of the detector, both at a depth of 10 cm, for each small field and for the reference beam of 10 x 10 cm(2). Results: The Monte Carlo calculated output correction factors for the liquid ionization chamber and the diamond detector were within about +/- 1% of unity even for the smallest field sizes. Corrections were found to be significant for small air ionization chambers due to their cavity dimensions, as expected. The correction factors for silicon diodes varied with the detector type (shielded or un-shielded), confirming the findings by other authors; different corrections for the detectors from the two manufacturers were obtained. The differences in the calculated factors for the various detectors were analyzed thoroughly and whenever possible the results were compared to published data, often calculated for different accelerators and using the EGSnrc MC system. The differences were used to estimate a type-B uncertainty for the correction factors. Together with the type-A uncertainty from the Monte Carlo calculations, an estimation of the combined standard uncertainty was made, assigned to the mean correction factors from various estimates. Conclusions: The present work provides a consistent and specific set of data for the output correction factors of a broad set of detectors in a Varian Clinac iX 6 MV accelerator and contributes to improving the understanding of the physics of small photon beams. The correction factors cannot in general be neglected for any detector and, as expected, their magnitude increases with decreasing field size. Due to the reduced number of clinical accelerator types currently available, it is suggested that detector output correction factors be given specifically for linac models and field sizes, rather than for a beam quality specifier that necessarily varies with the accelerator type and field size due to the different electron spot dimensions and photon collimation systems used by each accelerator model. (C) 2014 American Association of Physicists in Medicine.
  •  
2.
  • Ghazal, Mohammed, et al. (författare)
  • 6-MV small field output factors: intra-/intermachine comparison and implementation of TRS-483 using various detectors and several linear accelerators
  • 2019
  • Ingår i: Medical physics (Lancaster). - : WILEY. - 0094-2405 .- 2473-4209. ; 46:11, s. 5350-5359
  • Tidskriftsartikel (refereegranskat)abstract
    • Purpose To investigate the applicability of output correction factors reported in TRS-483 on 6-MV small-field detector-reading ratios using four solid-state detectors. Also, to investigate variations in 6-MV small-field output factors (OF) among nominally matched linear accelerators (linacs). Methods The TRS-483 Code of Practice (CoP) introduced and provided output correction factors to be applied to measured detector-reading ratios to obtain OFs for several small-field detectors. Detector readings for 0.5 cm x 0.5 cm to 8 cm x 8 cm fields were measured and normalized to that of 10 cm x 10 cm field giving the detector-reading ratios. Three silicon diodes, IBA PFD, IBA EFD (IBA, Schwarzenbruck, Germany), PTW T60017, and one microdiamond, PTW T60019 (PTW, Freiburg, Germany), were used. Output correction factors from the CoP were applied to measured detector-reading ratios. Measurements were performed on six Clinac and six TrueBeam linacs (Varian Medical Systems, Palo Alto, USA). An investigation of the relationship between the size of small fields and corresponding detector-reading ratio among the linacs was performed by measuring lateral dose profiles for 0.5 cm x 0.5 cm fields to determine the full width half maximum (FWHM). The relationship between the linacs focal spot size and the small-field detector-reading ratio was investigated by measuring 10 cm x 10 cm lateral dose profiles and determining the penumbra width reflecting the focal spot size. Measurement geometry was as follows: gantry angle = 0 degrees, collimator angle = 0 degrees, source-to surface distance (SSD) = 90 cm, and depth in water = 10 cm. Results For a given linac and 0.5 cm x 0.5 cm field, the deviations in detector-reading ratios among the detectors were 9%-15% for the Clinacs and 4%-5% for the TrueBeams. Use of output correction factors reduced these deviations to 6%-12% and 3%-4%, respectively. For field sizes equal to or larger than 0.8 cm x 0.8 cm, the deviations were corrected to 1% using output correction factors for both Clinacs and TrueBeams. For a given detector and 0.5 cm x 0.5 cm field, the deviations in detector-reading ratios among the linacs were 11%-17% for the Clinacs and 5-6% for the TrueBeams. For 1 cm x 1 cm the deviations were 1%-2% for Clinacs and 1% for TrueBeams. For field sizes larger than 1 cm x 1 cm the deviations were within 1% for both Clinacs and TrueBeams. No relationship between FWHMs and detector-reading ratios for 0.5 cm x 0.5 cm was observed. For Clinacs, larger 10 cm x 10 cm penumbra width yielded lower 0.5 cm x 0.5 cm detector-reading ratio indicating an effect of the focal spot size. For TrueBeams, the spread of penumbra widths was lower compared to Clinacs and no similar relationship was observed. Conclusions Output correction factors from the TRS-483 CoP are not sufficient for accurate determination of OF for 0.5 cm x 0.5 cm fields but are applicable for 0.8 cm x 0.8 cm to 8 cm x 8 cm fields. Nominally matched Clinacs and TrueBeams show large differences in detector-reading ratios for fields smaller than 1 cm x 1 cm.
  •  
3.
  • Häger, Wille, et al. (författare)
  • Experimental investigation of TRS-483 reference dosimetry correction factors for Leksell Gamma Knife (R) Icon (TM) beams
  • 2021
  • Ingår i: Medical physics (Lancaster). - : Wiley. - 0094-2405 .- 2473-4209. ; 48:1, s. 434-444
  • Tidskriftsartikel (refereegranskat)abstract
    • Purpose: Radiosurgery using the Leksell Gamma Knife (R) (LGK) Icon (TM) is an established technique used for treating intracranial lesions. The largest beam field size the LGK Icon can produce is a 16 mm diameter sphere. Despite this, reference dosimetry on the LGK Icon is typically performed using ionization chambers calibrated in 10 x 10 cm(2) fields. Furthermore, plastic phantoms are widely used instead of liquid water phantoms. In an effort to resolve these issues, the International Atomic Energy Agency (IAEA) in collaboration with American Association of Physicists in medicine (AAPM) recently published Technical Report Series No. 483 (TRS-483) as a Code of Practice for small-field dosimetry. TRS-483 includes small-field correction factors, k(Qmsr,Q0)(fmsr,fref), intended to account for the differences between setups when using small-field modalities such as the LGK Icon, and conventional setups. Since the publication of TRS-483, at least three new sets of values of k(Qmsr,Q0)(fmsr,fref) for the LGK Icon have been published. The purpose of this study was to experimentally investigate the published values of k(Qmsr,Q0)(fmsr,fref) for commonly used phantom and ionization chamber (IC) models for the LGK Icon.Methods: Dose-rates from two LGK units were determined using acrylonitrile butadiene styrene (ABS) and Certified Medical Grade Solid Water (R) (SW) phantoms, and PTW 31010 and PTW 31016 ICs. Correction factors were applied, and the resulting dose-rates compared. Relative validity of the correction factors was investigated by taking the ratios of dose-rate correction factor products. Additionally, dose-rates from the individual sectors were determined in order to calculate the beam attenuation caused by the ABS phantom adapter.Results and Conclusions: It was seen that the dose-rate is underestimated by at least 1% when using the ABS phantom, which was attributed to fluence perturbation caused by the IC and phantom adapter. Published correction factors k(Qmsr,Q0)(fmsr,fref) account for these effects to varying degree and should be used. The SW phantom is unlikely to underestimate the dose-rate by more than 1%, and applying k(Qmsr,Q0)(fmsr,fref) could not be shown to be necessary. Out of the two phantom models, the ABS phantom is not recommended for use in LGK reference dosimetry. The use of newly published values of k(Qmsr,Q0)(fmsr,fref) should be considered.
  •  
4.
  • Kaveckyte, Vaiva, et al. (författare)
  • Investigation of a synthetic diamond detector response in kilovoltage photon beams
  • 2020
  • Ingår i: Medical physics (Lancaster). - : Wiley-Blackwell Publishing Inc.. - 0094-2405 .- 2473-4209. ; 47:3, s. 1268-1279
  • Tidskriftsartikel (refereegranskat)abstract
    • Purpose An important characteristic of radiation dosimetry detectors is their energy response which consists of absorbed-dose and intrinsic energy responses. The former can be characterized using Monte Carlo (MC) simulations, whereas the latter (i.e., detector signal per absorbed dose to detector) is extracted from experimental data. Such a characterization is especially relevant when detectors are used in nonrelative measurements at a beam quality that differs from the calibration beam quality. Having in mind the possible application of synthetic diamond detectors (microDiamond PTW 60019, Freiburg, Germany) for nonrelative dosimetry of low-energy brachytherapy (BT) beams, we determined their intrinsic and absorbed-dose energy responses in 25-250 kV beams relative to a Co-60 beam, which is usually the reference beam quality for detector calibration in radiotherapy. Material and Methods Three microDiamond detectors and, for comparison, two silicon diodes (PTW 60017) were calibrated in terms of air-kerma free in air in six x-ray beam qualities (from 25 to 250 kV) and in terms of absorbed dose to water in a Co-60 beam at the national metrology laboratory in Sweden. The PENELOPE/penEasy MC radiation transport code was used to calculate the absorbed-dose energy response of the detectors (modeled based on blueprints) relative to air and water depending on calibration conditions. The MC results were used to extract the relative intrinsic energy response of the detectors from the overall energy response. Measurements using an independent setup with a single ophthalmic BEBIG I25.S16 I-125 BT seed (effective photon energy of 28 keV) were used as a qualitative check of the extracted intrinsic energy response correction factors. Additionally, the impact of the thickness of the active volume as well as the presence of extra-cameral components on the absorbed-dose energy response of a microDiamond detector was studied using MC simulations. Results The relative intrinsic energy response of the microDiamond detectors was higher by a factor of 2 in 25 and 50 kV beams compared to the Co-60 beam. The variation in the relative intrinsic energy response of silicon diodes was within 10% over the investigated photon energy range. The use of relative intrinsic energy response correction factors improved the agreement among the absorbed dose to water values determined using microDiamond detectors and silicon diodes, as well as with the TG-43 formalism-based calculations for the I-125 seed. MC study of microDiamond detector design features provided a possible explanation for inter-detector response variation at low-energy photon beams by differences in the effective thickness of the active volume. Conclusions MicroDiamond detectors had a non-negligible variation in the relative intrinsic energy response (factor of 2) which was comparable to that in the absorbed-dose energy response relative to water at low-energy photon beams. Silicon diodes, in contrast, had an absorbed-dose energy dependence on photon energy that varied by a factor of 6, whereas the intrinsic energy dependence on beam quality was within 10%. It is important to decouple these two responses for a full characterization of detector energy response especially when the user and reference beam qualities differ significantly, and MC alone is not enough.
  •  
5.
  • Kaveckyte, Vaiva, et al. (författare)
  • Suitability of microDiamond detectors for the determination of absorbed dose to water around high-dose-rate Ir-192 brachytherapy sources
  • 2018
  • Ingår i: Medical physics (Lancaster). - : WILEY. - 0094-2405 .- 2473-4209. ; 45:1, s. 429-437
  • Tidskriftsartikel (refereegranskat)abstract
    • Purpose: Experimental dosimetry of high-dose-rate (HDR) Ir-192 brachytherapy (BT) sources is complicated due to high dose and dose-rate gradients, and softening of photon energy spectrum with depth. A single crystal synthetic diamond detector microDiamond (PTW 60019, Freiburg, Germany) has a small active volume, high sensitivity, direct readout, and nearly water-equivalent active volume. The purpose of this study was to evaluate the suitability of microDiamond detectors for the determination of absorbed dose to water around HDR Ir-192 BT sources. Three microDiamond detectors were used, allowing for the comparison of their properties. Methods: In-phantom measurements were performed using microSelectron and VariSource iX HDR Ir-192 BT treatment units. Their treatment planning systems (TPSs), Oncentra (v. 4.3) and BrachyVision (v. 13.6), respectively, were used to create irradiation plans for a cubic PMMA phantom with the microDiamond positioned at one of three source-to-detector distances (SDDs) (1.5, 2.5, and 5.5 cm) at a time. The source was stepped in increments of 0.5 cm over a total length of 6 cm to yield absorbed dose of 2 Gy at the nominal reference-point of the detector. Detectors were calibrated in Co-60 beam in terms of absorbed dose to water, and Monte Carlo (MC) calculated beam quality correction factors were applied to account for absorbed-dose energy dependence. Phantom correction factors were applied to account for differences in dimensions between the measurement phantom and a water phantom used for absorbed dose calculations made with a TPS. The same measurements were made with all three of the detectors. Additionally, dose-rate dependence and stability of the detectors were evaluated in Co-60 beam. Results: The percentage differences between experimentally determined and TPS-calculated absorbed doses to water were from -1.3% to +2.9%. The values agreed to within experimental uncertainties, which were from 1.9% to 4.3% (k = 2) depending on the detector, SDD and treatment delivery unit. No dose-rate or intrinsic energy dependence corrections were applied. All microDiamonds were comparable in terms of preirradiation dose, stability of the readings and energy response, and showed a good agreement. Conclusions: The results indicate that the microDiamond is potentially suitable for the determination of absorbed dose to water around HDR Ir-192 BT sources and may be used for independent verification of TPSs calculations, as well as for QA measurements of HDR Ir-192 BT treatment delivery units at clinical sites. (C) 2017 American Association of Physicists in Medicine
  •  
Skapa referenser, mejla, bekava och länka
  • Resultat 1-5 av 5

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy