SwePub
Tyck till om SwePub Sök här!
Sök i SwePub databas

  Utökad sökning

Träfflista för sökning "L773:0277 786X OR L773:1996 756X ;hsvcat:3"

Sökning: L773:0277 786X OR L773:1996 756X > Medicin och hälsovetenskap

  • Resultat 1-10 av 58
Sortera/gruppera träfflistan
   
NumreringReferensOmslagsbildHitta
1.
  • Andersson-Engels, Stefan, et al. (författare)
  • Integrated system for interstitial photodynamic therapy
  • 2002
  • Ingår i: Advanced Optical Devices, Technologies, and Medical Applications. - : SPIE. - 0277-786X .- 1996-756X. ; 5123, s. 293-302
  • Konferensbidrag (refereegranskat)abstract
    • To develop PDT beyond treatment of thin superficial tumours, to also be an efficient treatment alternative for deeply located and/or thick tumours, a system based on interstitial illumination using multiple fibres has been developed. Conditions that could benefit from such a treatment modality are for instance malignant brain tumours and tumours in the oral cavity. In interstitial PDT one needs to use multiple fibres for light delivery in order to allow treatments of tumours larger than a few millimetres in diameter. Our system consists of a laser light source, a beam-splitting system dividing the light into three or six output fibres and a custom-made dosimetry programme. The concept is then to use these fibres not only for delivering the treatment light but also to measure parameters of interest for the treatment outcome. The fluence rate of the light emitted by each fibre is measured at the positions of the other fibre tips. From these results the light dose at all positions could be recalculated. Changes in optical properties as well as bleaching and concentration of the photosensitizer during the treatment could be monitored and compensated for in the dosimetry. Tumours have been treated both in experimental studies and in patients with thick superficial Basal Cell Carcinomas. Almost all treated skin lesions responded with complete response.
  •  
2.
  • Jensen, LK, et al. (författare)
  • Optical coherence tomography in clinical examination of non-pigmented skin malignancies
  • 2003
  • Ingår i: OPTICAL COHERENCE TOMOGRAPHY AND COHERENCE TECHNIQUES. - : SPIE. - 0277-786X .- 1996-756X. - 0819450103 ; 5140, s. 160-160
  • Konferensbidrag (refereegranskat)abstract
    • Optical coherence tomography (OCT) images of basal cell carcinomas (BCCs) have been acquired using a compact handheld probe with an integrated video camera allowing the OCT images to be correlated to a skin surface image. In general the healthy tissue of the skin has an obvious stratified structure, whereas the cancerous tissue shows a more homogenous structure. Thus it was demonstrated that it is possible to distinguish BCCs from healthy tissue by means of OCT. Furthermore different histological types of BCC were identified. Comparison of OCT images taken prior to and immediately after photodynamic therapy clearly shows the tissue response to the treatment, and indicates local oedema in the treated area.
  •  
3.
  • Andersson-Engels, Stefan, et al. (författare)
  • Integrated system for interstitial photodynamic therapy
  • 2003
  • Ingår i: Proceedings of SPIE - The International Society for Optical Engineering. - : SPIE. - 0277-786X .- 1996-756X. ; 5142, s. 42-49
  • Konferensbidrag (refereegranskat)abstract
    • A novel photodynamic therapy system based on interstitial illumination using multiple fibres is under development. The aim with this system is to enable treatment of large tumour volumes and also to utilise real-time measurements to allow on-line dosimetry. Important dosimetric parameters to measure are light fluence rate, sensitizer fluorescence intensity and local blood oxygenation. A construction which allows all functions to be readily performed with a single system is presented. We believe that interstitial PDT utilising this technique may be attractive in many clinical situations.
  •  
4.
  • Aurumskjöld, Marie-Louise, et al. (författare)
  • An investigation of the iterative reconstruction method iDose(4) on a Philips CT Brilliance 64 using a Catphan 600 phantom
  • 2012
  • Ingår i: Medical Imaging 2012: Physics of Medical Imaging. - : SPIE. - 0277-786X .- 1996-756X. ; 8313
  • Konferensbidrag (refereegranskat)abstract
    • The number of CT examinations giving a relatively high patient exposure is increasing. It is therefore important to optimize the imaging conditions at these investigations. Many steps have been taken to reduce the radiation doses in CT examinations. Currently much work is related to iterative image reconstruction methods as alternative to the filtered back projection method. The aim of this work was to evaluate quality parameters in images from a CT (Philips Brilliance 64) equipped with the iterative reconstruction method iDose(4) using a Catphan 600 phantom with and without body simulating ring. CT scans using abdomen protocol were taken with various tube currents and tube voltage and keeping collimation and pitch unchanged for all scans. All collected data were reconstructed with different levels of iDose(4) (Level 2, 4, 6) and traditional filtered back projection. Image quality parameters were evaluated using AutoQA Lite (TM) (Version 2.3 2007 Iris QA, LLC). Results from the study shows that the iterative reconstruction method decreases the noise with 15-45% compared with filtered back projection depending on which level of iDose(4) is used. The percentage reduction in noise level is the same with and without body simulating ring. Low contrast was improved with iDose(4) and spatial resolution is only marginally affected by the method of reconstruction. However by reducing the image noise, the detectability can be improved. Our conclusion is that there is great potential to reduce the noise and thereby improve the image quality by using iterative reconstruction methods. This can also be used to lower radiation dose and maintain image quality or improve image quality.
  •  
5.
  • Aurumskjöld, Marie-Louise, et al. (författare)
  • Model based iterative reconstruction IMR gives possibility to evaluate thinner slice thicknesses than conventional iterative reconstruction iDose(4) - a phantom study
  • 2015
  • Ingår i: Medical Imaging 2015: Physics of Medical Imaging. - : SPIE. - 1996-756X .- 0277-786X. ; 9412, s. 94124-94124
  • Konferensbidrag (refereegranskat)abstract
    • Computed tomography (CT) is one of the most important modalities in a radiological department, which produces images with high diagnostic confidence, but in some cases contributes to a high radiation dose to the patient. The radiation dose can be reduced by the use of advanced image reconstruction algorithms. This study was done on a Philips Brilliance iCT with iterative reconstruction iDose(4) and model-based iterative reconstruction IMR. The purpose was to investigate the effect on the image quality with thin slice images reconstructed with IMR, compared to standard slice thickness reconstructed with iDose(4). Objective measurements of noise and contrast-to-noise ratio were performed using an image quality phantom, an anthropomorphic phantom and clinical cases. Subjective evaluations of low-contrast resolution were performed by observers using an image quality phantom. IMR gives strong noise reduction and enhanced low-contrast and thereby enable selection of thinner slice thickness. Objective evaluation of image noise shows that thin slices reconstructed with IMR provides lower noise than thicker slice images reconstructed with iDose(4). With IMR the slice thickness is of less importance for the noise. With thinner slices the partial volume artefacts becomes less pronounced. In conclusion, we have shown that IMR enables reduction of the slice thickness and at the same time maintain or even reduce the noise level compared to iDose(4) reconstruction with standard slice thickness. This will subsequently result in an improvement of image quality for images reconstructed with IMR.
  •  
6.
  • Axelsson, Rebecca, et al. (författare)
  • Simultaneous digital breast tomosynthesis and mechanical imaging in women recalled from screening - A preliminary analysis
  • 2022
  • Ingår i: 16th International Workshop on Breast Imaging, IWBI 2022. - : SPIE. - 1996-756X .- 0277-786X. - 9781510655843 ; 12286
  • Konferensbidrag (refereegranskat)abstract
    • We have developed a method for simultaneous tomosynthesis and mechanical imaging, called DBTMI. Mechanical imaging measures the stress distribution over the compressed breast surface. Malignant tissue is usually stiffer than benign, which results in higher stress on the compressed breast and enables to distinguish malignant from benign findings. By combining tomosynthesis and mechanical imaging, we could improve cancer detection accuracy by reducing the number of false positive findings. In this study we have analysed clinical DBTMI data, collected from 52 women from an ongoing pilot study at the Skåne University Hospital, Malmö, Sweden. We measured the range of the average stress over the breast surface, the range of average stress over the location of suspected lesions, and the normalized stress over the lesion location. Preliminary results show that the range of stress over the breast surface was 1.23-5.84 kPa, the range over the lesion location 2.10-10.10 kPa, and the normalized stress 1.12-2.44 over the lesion location. Overall, the local stress over malignant lesions was higher than the average stress over the entire breast surface. This is the first step investigating criteria to distinguish between malignant and benign findings based upon clinical DBTMI data.
  •  
7.
  • Bakic, Predrag R., et al. (författare)
  • Evaluation of a flat fielding method for simultaneous DBT and MI acquisition
  • 2020
  • Ingår i: 15th International Workshop on Breast Imaging, IWBI 2020. - : SPIE. - 1996-756X .- 0277-786X. - 9781510638310 ; 11513
  • Konferensbidrag (refereegranskat)abstract
    • We are developing a prototype system for simultaneous digital breast tomosynthesis (DBT) and mechanical imaging (MI). MI maps the local pressure distribution during clinical exams, to distinguish breast abnormalities from the normal tissue. Both DBT alone, and MI when combined with digital mammography, have demonstrated the ability to reduce false positives; however, the benefit of combining DBT with MI has not been investigated. A practical limitation in simultaneous DBT and MI is the presence of the MI sensor in DBT images. Metallic elements of the sensor generate noticeable artifacts, which may interfere with clinical analysis. Previously, we shown that the sensor artifacts can be reduced by flat fielding, which combines projections of the sensor acquired with and without the breast. In this paper we evaluate the flat fielding by assessing artifact reduction and visibility of breast abnormalities. Images of a physical anthropomorphic breast phantom were acquired using a clinical wide-angle DBT system. Visual evaluation was performed by experienced medical physicists. Image quality descriptors were calculated in images with and without flat fielding. To evaluate the visibility of abnormalities we estimated the full width at half maximum (FWHM) for calcifications modeled in the phantom. Our preliminary results suggest a substantial reduction of artifacts by flat fielding (on average 83%). Few noticeable artifacts remain near the breast edge, in the reconstructed image with the sensor in focus. We observed a 17% reduction in the FWHM. Future work would include a detailed assessment, and method optimization using virtual trials as a design aid.
  •  
8.
  • Bejnö, Anna, et al. (författare)
  • Artificial intelligence together with mechanical imaging in mammography
  • 2020
  • Ingår i: 15th International Workshop on Breast Imaging, IWBI 2020. - : SPIE. - 0277-786X .- 1996-756X. - 9781510638310 ; 11513
  • Konferensbidrag (refereegranskat)abstract
    • Artificial intelligence (AI) applications are increasingly seeing use in breast imaging, particularly to assist in or automate the reading of mammograms. Another novel technique is mechanical imaging (MI) which estimates the relative stiffness of suspicious breast abnormalities by measuring the distribution of pressure on the compressed breast. This study investigates the feasibility of combining AI and MI information in breast imaging to provide further diagnostic information. Forty-six women recalled from screening were included in the analysis. Mammograms with findings scored on a suspiciousness scale by an AI tool, and corresponding pressure distributions were collected for each woman. The cases were divided into three groups by diagnosis; biopsy-proven cancer, biopsy-proven benign and non-biopsied, very likely benign. For all three groups, the relative increase of pressure at the location of the finding marked most suspicious by the AI software was recorded. A significant correlation between the relative pressure increase at the AI finding and the AI score was established in the group with cancer (p=0.043), but neither group of healthy women showed such a correlation. This study suggests that AI and MI indicate independent markers for breast cancer. The combination of these two methods has the potential to increase the accuracy of mammography screening, but further research is needed.
  •  
9.
  • Boll, Måns, et al. (författare)
  • Evaluation of 3D printed contrast detail phantoms for mammography quality assurance
  • 2022
  • Ingår i: 16th International Workshop on Breast Imaging, IWBI 2022. - : SPIE. - 0277-786X .- 1996-756X. - 9781510655843 ; 12286
  • Konferensbidrag (refereegranskat)abstract
    • Objects created by 3D printers are increasingly used in various medical applications. Today, affordable 3D printers, using Fused Deposition Modeling are widely available. In this project, a commercially available 3D printer was used to replicate a conventional radiographic contrast detail phantom. Printing materials were selected by comparing their x-ray attenuation properties. Two replicas were printed using polylactic acid, with different filling patterns. The printed phantoms were imaged by a clinical mammography system, using automatic exposure control. Phantom images were visually and quantitively compared to images of the corresponding conventional contrast detail phantom. Visual scoring of the contrast detail elements was performed by a medical physics student. Contrast-to-noise ratio (CNR) was calculated for each phantom element. The diameter and thickness of the smallest visible phantom object were 0.44 mm and 0.09 mm, respectively, for both filling patterns. For the conventional phantom, the diameter and thickness of the smallest visible object were 0.31 mm and 0.09 mm. Visual inspection of printed phantoms revealed some linear artefacts. These artefacts were however not visible on mammographic projections. Quantitively, average CNR of printed phantom objects followed the same trend with an increase of average CNR with increasing disk height. However, there is a limitation of detail objects with disk diameters below 1.25 mm, caused by the available nozzle size. Based upon the encouraging results, future work will explore the use of different materials and smaller nozzle diameters.
  •  
10.
  • Chen, Han, et al. (författare)
  • On imaging with or without grid in digital mammography
  • 2014
  • Ingår i: Proceedings of SPIE, the International Society for Optical Engineering. - : SPIE - International Society for Optical Engineering. - 0277-786X .- 1996-756X. ; 9033, s. 903346-
  • Tidskriftsartikel (refereegranskat)abstract
    • The grids used in digital mammography to reduce scattered radiation from the breast are not perfect and lead to partial absorption of primary radiation at the same time as not all of the scattered radiation is absorbed. It has therefore lately been suggested to remove the grids and correct for effects of scattered radiation by post-processing the images. In this paper, we investigated the dose reduction that might be achieved if the gird were to be removed. Dose reduction is determined as a function of PMMA thickness by comparing the contrast-to-noise ratios (CNRs) of images acquired with and without grid at a constant exposure. We used a theoretical model validated with Monte Carlo simulations and phantom studies. To evaluate the CNR, we applied aluminum filters of two different sizes, 4x8 cm2 and 1x1 cm 2. When the large Al filter was used, the resulting CNR value for the grid-less images was overestimated as a result of a difference in amount of scattered radiation in the background region and of the region covered by the filter, a difference that could be eliminated by selecting a region of interest close to the edge of the filter. The optimal CNR when the PMMA thickness was above about 4 cm was obtained with a grid, whereas removing the grid leaded to a dose saving in thinner PMMAs. The results suggest not removing grids in breast cancer screening.
  •  
Skapa referenser, mejla, bekava och länka
  • Resultat 1-10 av 58
Typ av publikation
konferensbidrag (43)
tidskriftsartikel (15)
Typ av innehåll
refereegranskat (50)
övrigt vetenskapligt/konstnärligt (8)
Författare/redaktör
Zackrisson, Sophia (12)
Båth, Magnus, 1974 (9)
Andersson-Engels, St ... (8)
Svanberg, Katarina (8)
Svanberg, Sune (6)
Bendsöe, Niels (5)
visa fler...
Söderberg, Marcus (4)
Hemdal, Bengt (3)
Stenram, Unne (2)
Zackrisson, S (2)
Johansson, Thomas (2)
Fellman, Vineta (2)
Mattsson, Sören (2)
Andersson, Martin (1)
Chakraborty, D. (1)
Galli, Silvia (1)
Adiels, Caroline B., ... (1)
Goksör, Mattias, 197 ... (1)
Johansson, T (1)
Rosen, A (1)
Svensson, Tomas (1)
Thompson, M (1)
Rystedt, Hans, 1951 (1)
Tranberg, Karl-Göran (1)
Rosen, Arne, 1939 (1)
Abul-Kasim, Kasim (1)
Ohlin, Acke (1)
Gunnarsson, Mikael (1)
Muller, Bert (1)
Wang, Ge (1)
Andersson, Mats (1)
Åkeson, Jonas (1)
Arvidsson, Jonathan (1)
Heuser, Philipp (1)
Maguire Jr., Gerald ... (1)
Dinér, Peter (1)
Iima, Mami (1)
Tylén, Ulf, 1938 (1)
Grøtli, Morten, 1966 (1)
Noz, Marilyn E. (1)
Wårdell, Karin, 1959 ... (1)
Wennerberg, Ann (1)
Zeleznik, Michael P. (1)
Chodorowski, Artur, ... (1)
Alm Carlsson, G (1)
Danielsson, Mats (1)
Ivarsson, Jonas, 197 ... (1)
Mattsson, S. (1)
Zaidi, Habib (1)
Medin, Joakim (1)
visa färre...
Lärosäte
Lunds universitet (40)
Göteborgs universitet (12)
Uppsala universitet (5)
Kungliga Tekniska Högskolan (4)
Chalmers tekniska högskola (4)
Luleå tekniska universitet (2)
visa fler...
Örebro universitet (2)
Linnéuniversitetet (2)
Högskolan Väst (1)
Linköpings universitet (1)
Malmö universitet (1)
visa färre...
Språk
Engelska (58)
Forskningsämne (UKÄ/SCB)
Naturvetenskap (8)
Teknik (5)
Samhällsvetenskap (1)

År

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy