Potential sources of quantification error when retrospectively assessing metacarpal bone loss from historical radiographs by using digital X-ray radiogrammetry : an experimental study

Kälvesten, Johan (författare)
Linköpings universitet,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Avdelningen för radiologiska vetenskaper,Hälsouniversitetet,Sectra Imtec AB, Linköping, Sweden
Brismar, Torkel B. (författare)
Karolinska Institutet
Persson, Anders (författare)
Östergötlands Läns Landsting,Linköpings universitet,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Avdelningen för radiologiska vetenskaper,Hälsouniversitetet,Röntgenkliniken i Linköping
 (creator_code:org_t)
Elsevier, 2014
2014
Engelska.
Ingår i: Journal of clinical densitometry. - : Elsevier. - 1094-6950 .- 1559-0747. ; 17:1, s. 104-108
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • During the past 15 yr, digital X-ray radiogrammetry (DXR) has been used to measure metacarpal bone mineral density (BMD). BMD is often measured in existing cohorts where X-ray images were not acquired in accordance with the DXR imaging protocol (DIP). The purpose of the present study was to analyze how deviations from DIP in historical radiographs may affect the reproducibility of DXR-BMD measurements. Cadaver hand phantoms were used to conduct repeat measurements of deviations from DIP with respect to voltage, exposure, lateral displacement, supination, combination of lateral displacement and supination or rotation, extension of the wrist, and edge enhancement. Direct digital radiography (Aristos; Siemens Healthcare, Erlangen, Germany) was used for image acquisition and dxr-online (Sectra, Linkoping, Sweden) for DXR-BMD measurements. The influence of the tested deviations from DIP ranged from 0 to 32.5 mg/cm(2) (0-6.8%). On repetition with the same specimen, none of the deviations resulted in a within-specimen reproducibility error greater than 2 mg/cm(2) (0.4%, equivalent to a T-score of 0.042). Among the tested deviations, all except tube voltage had a magnitude greater than the normal measurement noise for the technique and must therefore be considered when planning a study based on historical images.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

Bone mineral density; digital X-ray radiogrammetry; osteoporosis; rheumatoid arthritis
MEDICINE
MEDICIN

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy