Risk of nerve injury after use of energy based surgical devices

Carlander, Johan (författare)
Linköpings universitet,Institutionen för klinisk och experimentell medicin,Hälsouniversitetet,Department of Surgery and Center for Clinical Research Uppsala University, Västmanland County Hospital, Västerås, Sweden
Defechereux, T (författare)
Department of Endocrine Surgery, University Hospital CHU Sart-Tilman, Liege, Belgium
Koch, C (författare)
Physikalisch-Technische Bundesanstalt, Braunschweig, Germany
visa fler...
Cheramy, JP. (författare)
Department of animal laboratory, Liege, Belgium
Deprez, M. (författare)
Department of Pathology, Liege University, Belgium
Johansson, K (författare)
Linköpings universitet,Institutionen för klinisk och experimentell medicin,Medicinska fakulteten,Dept Surgery, Västervik Hospital, Sweden / Department of Surgery, Sahlgrenska University Hospital Gothenburg, Sweden
visa färre...
 (creator_code:org_t)
2015
Engelska.
  • Annan publikation (övrigt vetenskapligt/konstnärligt)
Abstract Ämnesord
Stäng  
  • BackgroundThe use of energy based surgical devices (EBD) is well established in surgery. Iatrogenic nerve injury is a common problem in many areas of surgery and may be caused by collateral thermal injury from EBD.MethodsThe sciatic nerve of anaesthetized rats was used in an experimental model. A bipolar scissors (ES) was compared to an ultrasonic device (UD) by cutting the femoris muscle longitudinal during 3 seconds at 1 mm from the sciatic nerve. Temperature and electromyography (EMG) were recorded before, during and after activation of the devices. The nerves were examined blinded with light (LM) and electronic microscope (EM).ResultsThe temperature increase was significantly less and with shorter duration after instrument activationfor the UD compared to bipolar ES. The EMG potential was reduced by 3 % in the UD group and 6% in the ES group respectively (n.s.). LM and EM demonstrated a loss of density in the myelin sheath in a small number of nerves in both groups.ConclusionThis study indicates less thermal spread in tissue following activation of the ultrasonic devices compared to bipolar ES. EMG and morphology assessment with LM and EM indicate a small risk and probably reversible thermal injury after clinical relevant instrument activation at 1 mm from the nerve.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Kirurgi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Surgery (hsv//eng)

Nyckelord

Energy based devices
nerve injury
electron microscope

Publikations- och innehållstyp

vet (ämneskategori)
ovr (ämneskategori)

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy