SwePub
Sök i SwePub databas

  Utökad sökning

Träfflista för sökning "WFRF:(Markenroth Bloch Karin) ;pers:(Heiberg Einar)"

Sökning: WFRF:(Markenroth Bloch Karin) > Heiberg Einar

  • Resultat 1-7 av 7
Sortera/gruppera träfflistan
   
NumreringReferensOmslagsbildHitta
1.
  • Bock, Jelena, et al. (författare)
  • Validation and reproducibility of cardiovascular 4D-flow MRI from two vendors using 2 × 2 parallel imaging acceleration in pulsatile flow phantom and in vivo with and without respiratory gating
  • 2018
  • Ingår i: Acta Radiologica. - : SAGE Publications. - 0284-1851 .- 1600-0455.
  • Tidskriftsartikel (refereegranskat)abstract
    • Background: 4D-flow magnetic resonance imaging (MRI) is increasingly used. Purpose: To validate 4D-flow sequences in phantom and in vivo, comparing volume flow and kinetic energy (KE) head-to-head, with and without respiratory gating. Material and Methods: Achieva dStream (Philips Healthcare) and MAGNETOM Aera (Siemens Healthcare) 1.5-T scanners were used. Phantom validation measured pulsatile, three-dimensional flow with 4D-flow MRI and laser particle imaging velocimetry (PIV) as reference standard. Ten healthy participants underwent three cardiac MRI examinations each, consisting of cine-imaging, 2D-flow (aorta, pulmonary artery), and 2 × 2 accelerated 4D-flow with (Resp+) and without (Resp−) respiratory gating. Examinations were acquired consecutively on both scanners and one examination repeated within two weeks. Volume flow in the great vessels was compared between 2D- and 4D-flow. KE were calculated for all time phases and voxels in the left ventricle. Results: Phantom results showed high accuracy and precision for both scanners. In vivo, higher accuracy and precision (P < 0.001) was found for volume flow for the Aera prototype with Resp+ (–3.7 ± 10.4 mL, r = 0.89) compared to the Achieva product sequence (–17.8 ± 18.6 mL, r = 0.56). 4D-flow Resp− on Aera had somewhat larger bias (–9.3 ± 9.6 mL, r = 0.90) compared to Resp+ (P = 0.005). KE measurements showed larger differences between scanners on the same day compared to the same scanner at different days. Conclusion: Sequence-specific in vivo validation of 4D-flow is needed before clinical use. 4D-flow with the Aera prototype sequence with a clinically acceptable acquisition time (<10 min) showed acceptable bias in healthy controls to be considered for clinical use. Intra-individual KE comparisons should use the same sequence.
  •  
2.
  • Carlsson, Marcus, et al. (författare)
  • Quantification and visualization of cardiovascular 4D velocity mapping accelerated with parallel imaging or k-t BLAST: head to head comparison and validation at 1.5 T and 3 T
  • 2011
  • Ingår i: Journal of Cardiovascular Magnetic Resonance. - 1097-6647. ; 13:55
  • Tidskriftsartikel (refereegranskat)abstract
    • Background: Three-dimensional time-resolved (4D) phase-contrast (PC) CMR can visualize and quantify cardiovascular flow but is hampered by long acquisition times. Acceleration with SENSE or k-t BLAST are two possibilities but results on validation are lacking, especially at 3 T. The aim of this study was therefore to validate quantitative in vivo cardiac 4D-acquisitions accelerated with parallel imaging and k-t BLAST at 1.5 T and 3 T with 2D-flow as the reference and to investigate if field strengths and type of acceleration have major effects on intracardiac flow visualization. Methods: The local ethical committee approved the study. 13 healthy volunteers were scanned at both 1.5 T and 3 T in random order with 2D-flow of the aorta and main pulmonary artery and two 4D-flow sequences of the heart accelerated with SENSE and k-t BLAST respectively. 2D-image planes were reconstructed at the aortic and pulmonary outflow. Flow curves were calculated and peak flows and stroke volumes (SV) compared to the results from 2D-flow acquisitions. Intra-cardiac flow was visualized using particle tracing and image quality based on the flow patterns of the particles was graded using a four-point scale. Results: Good accuracy of SV quantification was found using 3 T 4D-SENSE (r(2) = 0.86, -0.7 +/- 7.6%) and although a larger bias was found on 1.5 T (r(2) = 0.71, -3.6 +/- 14.8%), the difference was not significant (p = 0.46). Accuracy of 4D k-t BLAST for SV was lower (p < 0.01) on 1.5 T (r(2) = 0.65, -15.6 +/- 13.7%) compared to 3 T (r(2) = 0.64, -4.6 +/- 10.0%). Peak flow was lower with 4D-SENSE at both 3 T and 1.5 T compared to 2D-flow (p < 0.01) and even lower with 4D k-t BLAST at both scanners (p < 0.01). Intracardiac flow visualization did not differ between 1.5 T and 3 T (p = 0.09) or between 4D-SENSE or 4D k-t BLAST (p = 0.85). Conclusions: The present study showed that quantitative 4D flow accelerated with SENSE has good accuracy at 3 T and compares favourably to 1.5 T. 4D flow accelerated with k-t BLAST underestimate flow velocities and thereby yield too high bias for intra-cardiac quantitative in vivo use at the present time. For intra-cardiac 4D-flow visualization, however, 1.5 T and 3 T as well as SENSE or k-t BLAST can be used with similar quality.
  •  
3.
  •  
4.
  • Kanski, Mikael, et al. (författare)
  • Whole-heart four-dimensional flow can be acquired with preserved quality without respiratory gating, facilitating clinical use : A head-to-head comparison
  • 2015
  • Ingår i: BMC Medical Imaging. - : Springer Science and Business Media LLC. - 1471-2342. ; 15:1
  • Tidskriftsartikel (refereegranskat)abstract
    • Background: Respiratory gating is often used in 4D-flow acquisition to reduce motion artifacts. However, gating increases scan time. The aim of this study was to investigate if respiratory gating can be excluded from 4D flow acquisitions without affecting quantitative intracardiac parameters. Methods: Eight volunteers underwent CMR at 1.5 T with a 5-channel coil (5ch). Imaging included 2D flow measurements and whole-heart 4D flow with and without respiratory gating (Resp(+), Resp(-)). Stroke volume (SV), particle-trace volumes, kinetic energy, and vortex-ring volume were obtained from 4D flow-data. These parameters were compared between 5ch Resp(+) and 5ch Resp(-). In addition, 20 patients with heart failure were scanned using a 32-channel coil (32ch), and particle-trace volumes were compared to planimetric SV. Paired comparisons were performed using Wilcoxon's test and correlation analysis using Pearson r. Agreement was assessed as bias ± SD. Results: Stroke volume from 4D flow was lower compared to 2D flow both with and without respiratory gating (5ch Resp(+) 88 ± 18 vs 97 ± 24.0, p = 0.001; 5ch Resp(-) 86 ± 16 vs 97.1 ± 22.7, p < 0.01). There was a good correlation between Resp(+) and Resp(-) for particle-trace derived volumes (R2 = 0.82, 0.2 ± 9.4 ml), mean kinetic energy (R2 = 0.86, 0.07 ± 0.21 mJ), peak kinetic energy (R2 = 0.88, 0.14 ± 0.77 mJ), and vortex-ring volume (R2 = 0.70, -2.5 ± 9.4 ml). Furthermore, good correlation was found between particle-trace volume and planimetric SV in patients for 32ch Resp(-) (R2 = 0.62, -4.2 ± 17.6 ml) and in healthy volunteers for 5ch Resp(+) (R2 = 0.89, -11 ± 7 ml), and 5ch Resp(-) (R2 = 0.93, -7.5 ± 5.4 ml), Average scan duration for Resp(-) was shorter compared to Resp(+) (27 ± 9 min vs 61 ± 19 min, p < 0.05). Conclusions: Whole-heart 4D flow can be acquired with preserved quantitative results without respiratory gating, facilitating clinical use.
  •  
5.
  • Nilsson, Anders, et al. (författare)
  • Accuracy of four-dimensional phase-contrast velocity mapping for blood flow visualizations: a phantom study.
  • 2013
  • Ingår i: Acta Radiologica. - : SAGE Publications. - 1600-0455 .- 0284-1851. ; 54:6, s. 663-671
  • Tidskriftsartikel (refereegranskat)abstract
    • BackgroundTime-resolved three-dimensional, three-directional phase-contrast magnetic resonance velocity mapping (4D PC-MRI) is a powerful technique to depict dynamic blood flow patterns in the human body. However, the impact of phase background effects on flow visualizations has not been thoroughly studied previously, and it has not yet been experimentally demonstrated to what degree phase offsets affect flow visualizations and create errors such as inaccurate particle traces.PurposeTo quantify background phase offsets and their subsequent impact on particle trace visualizations in a 4D PC-MRI sequence. Additionally, we sought to investigate to what degree visualization errors are reduced by background phase correction.Material and MethodsA rotating phantom with a known velocity field was used to quantify background phase of 4D PC-MRI sequences accelerated with SENSE as well as different k-t BLAST speed-up factors. The deviation in end positions between particle traces in the measured velocity fields were compared before and after the application of two different phase correction methods.ResultsPhantom measurements revealed background velocity offsets up to 7 cm/s (7% of velocity encoding sensitivity) in the central slice, increasing with distance from the center. Background offsets remained constant with increasing k-t BLAST speed-up factors. End deviations of up to 5.3 mm (1.8 voxels) in the direction perpendicular to the rotating disc were found between particle traces and the seeding plane of the traces. Phase correction by subtraction of the data from the stationary phantom reduced the average deviation by up to 56%, while correcting the data-set with a first-order polynomial fit to stationary regions decreased average deviation up to 78%.ConclusionPathline visualizations can be significantly affected by background phase errors, highlighting the importance of dedicated and robust phase correction methods. Our results show that pathline deviation can be substantial if adequate phase background errors are not minimized.
  •  
6.
  • Nilsson, Anders, et al. (författare)
  • Variable velocity encoding in a three-dimensional, three-directional phase contrast sequence: Evaluation in phantom and volunteers.
  • 2012
  • Ingår i: Journal of Magnetic Resonance Imaging. - : Wiley. - 1522-2586 .- 1053-1807.
  • Tidskriftsartikel (refereegranskat)abstract
    • PURPOSE: To evaluate accuracy and noise properties of a novel time-resolved, three-dimensional, three-directional phase contrast sequence with variable velocity encoding (denoted 4D-vPC) on a 3 Tesla MR system, and to investigate potential benefits and limitations of variable velocity encoding with respect to depicting blood flow patterns. MATERIALS AND METHODS: A 4D PC-MRI sequence was modified to allow variable velocity encoding (VENC) over the cardiac cycle in all three velocity directions independently. 4D-PC sequences with constant and variable VENC were compared in a rotating phantom with respect to measured velocities and noise levels. Additionally, comparison of flow patterns in the ascending aorta was performed in six healthy volunteers. RESULTS: Phantom measurements showed a linear relationship between velocity noise and velocity encoding. 4D-vPC MRI presented lower noise levels than 4D-PC both in phantom and in volunteer measurements, in agreement with theory. Volunteer comparisons revealed more consistent and detailed flow patterns in early diastole for the variable VENC sequences. CONCLUSION: Variable velocity encoding offers reduced noise levels compared with sequences with constant velocity encoding by optimizing the velocity-to-noise ratio (VNR) to the hemodynamic properties of the imaged area. Increased VNR ratios could be beneficial for blood flow visualizations of pathology in the cardiac cycle. J. Magn. Reson. Imaging 2012. © 2012 Wiley Periodicals, Inc.
  •  
7.
  •  
Skapa referenser, mejla, bekava och länka
  • Resultat 1-7 av 7

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy