SwePub
Tyck till om SwePub Sök här!
Sök i LIBRIS databas

  Utökad sökning

WFRF:(Smedby Örjan)
 

Sökning: WFRF:(Smedby Örjan) > Karolinska Institutet > Axelsson Rimma > Comparison of acqui...

Comparison of acquisition protocols for ventilation/perfusion SPECT - a Monte Carlo study

Holstensson, Maria (författare)
Karolinska Institutet
Smedby, Örjan, 1956- (författare)
KTH,Medicinsk avbildning,medicinsk bildbehandling och visualisering
Poludniowski, Gavin (författare)
Karolinska Institutet
visa fler...
Sanches Crespo, Alejandro (författare)
Karolinska Institutet
Savitcheva, Irina (författare)
Öberg, Michael (författare)
Grybäck, Per (författare)
Karolinska Institutet
Gabrielson, Stefan (författare)
Karolinska Institutet
Sandqvist, Patricia (författare)
Karolinska Institutet
Bartholdson, Erika (författare)
Axelsson, Rimma (författare)
Karolinska Institutet
visa färre...
 (creator_code:org_t)
2019-12-05
2019
Engelska.
Ingår i: Physics in Medicine and Biology. - : Institute of Physics (IOP). - 0031-9155 .- 1361-6560. ; 64:23
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • One of the most commonly used imaging techniques for diagnosing pulmonary embolism (PE) is ventilation/perfusion (V/P) scintigraphy. The aim of this study was to evaluate the performance of the currently used imaging protocols for V/P single photon emission computed tomography (V/P SPECT) at two nuclear medicine department sites and to investigate the effect of altering important protocol parameters. 
 
 The Monte Carlo technique was used to simulate 4D digital phantoms with perfusion defects. Six imaging protocols were included in the study and a total of 72 digital patients were simulated. Six dually trained radiologists/nuclear medicine physicians reviewed the images and reported all perfusion mismatch findings. The radiologists also visually graded the image quality. 
 
 No statistically significant differences in diagnostic performance were found between the studied protocols, but visual grading analysis pointed out one protocol as significantly superior to four of the other protocols. Considering the study results, we have decided to harmonize our clinical protocols for imaging patients with suspected PE. The administered Technegas and macro aggregated albumin activities have been altered, a low energy all purpose collimator is used instead of a low energy high resolution collimator and the acquisition times have been lowered.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

Monte Carlo
SPECT
embolism
lung scintigraphy
perfusion
pulmonary
ventilation
Medical Technology
Medicinsk teknologi

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy