A Novel Approach for Removing the Hook Effect Artefact from Electrical Bioimpedance Spectroscopy Measurements

Buendia, Ruben, 1982 (författare)
Högskolan i Borås,Institutionen Ingenjörshögskolan
Seoane, Fernando (författare)
Högskolan i Borås,Institutionen Ingenjörshögskolan,Chalmers tekniska högskola,Chalmers University of Technology
Gil-Pita, Roberto (författare)
 (creator_code:org_t)
2010-05-19
2009
Engelska.
Ingår i: Journal of Physics: Conference Series. - : Institute of Physics Publishing Ltd.. - 1742-6596.
  • Konferensbidrag (refereegranskat)
Abstract Ämnesord
Stäng  
  • Very often in Electrical Bioimpedance (EBI) spectroscopy measurements the presence of stray capacitances creates a measurement artefact commonly known as Hook Effect. Such an artefact creates a hook-alike deviation of the EBI data noticeable when representing the measurement on the impedance plane. Such Hook Effect is noticeable at high frequencies but it also causes a data deviation at lower measurement frequencies. In order to perform any accurate analysis of the EBI spectroscopy data, the influence of the Hook Effect must be removed. An established method to compensate the hook effect is the well known Td compensation, which consist on multiplying the obtained spectrum, Zmeas() by a complex exponential in the form of exp[jTd]. Such a method cannot correct entirely the Hook Effect since the hook-alike deviation occurs a broad frequency range in both magnitude and phase of the measured impedance, and by using a real value for Td. First, a real value only modifies the phase of the measured impedance and second, it can only correct the Hook Effect at a single frequency. In addition, the process to select a value for Td by an iterative process with the aim to obtain the best Cole fitting lacks solid scientific grounds. In this work the Td compensation method is revisited and a modified approach for correcting the Hook Effect that includes a novel method for selecting the correcting values is proposed. The initial validation results confirm that the proposed method entirely corrects the Hook Effect at all frequencies.

Ämnesord

TEKNIK OCH TEKNOLOGIER  -- Elektroteknik och elektronik -- Reglerteknik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Electrical Engineering, Electronic Engineering, Information Engineering -- Control Engineering (hsv//eng)
TEKNIK OCH TEKNOLOGIER  -- Elektroteknik och elektronik -- Annan elektroteknik och elektronik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Electrical Engineering, Electronic Engineering, Information Engineering -- Other Electrical Engineering, Electronic Engineering, Information Engineering (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Medicinsk bioteknologi -- Biomedicinsk laboratorievetenskap/teknologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Medical Biotechnology -- Biomedical Laboratory Science/Technology (hsv//eng)
NATURVETENSKAP  -- Data- och informationsvetenskap -- Annan data- och informationsvetenskap (hsv//swe)
NATURAL SCIENCES  -- Computer and Information Sciences -- Other Computer and Information Science (hsv//eng)
TEKNIK OCH TEKNOLOGIER  -- Medicinteknik -- Medicinsk laboratorie- och mätteknik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering -- Medical Laboratory and Measurements Technologies (hsv//eng)

Nyckelord

bioimpedance
measurement instrumentation
applied signal processing
artefacts correction
Medical Electronics

Publikations- och innehållstyp

ref (ämneskategori)
kon (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy