Evaluation of ultrasound Tissue Velocity Imaging : a phantom study of velocity estimation in skeletal muscle low-level contractions
-
- Lindberg, Frida, 1982- (författare)
- KTH,Medicinsk teknik
-
- Mårtensson, Mattias (författare)
- KTH,Medicinsk bildteknik
-
- Grönlund, Christer (författare)
- Umeå universitet,Radiofysik,Medicinsk Teknik - FoU, Umeå Univeristet
-
visa fler...
-
- Brodin, Lars-Åke (författare)
- KTH,Medicinsk teknik
-
visa färre...
-
(creator_code:org_t)
- 2013-06-07
- 2013
- Engelska.
-
Ingår i: BMC Medical Imaging. - : BioMed Central. - 1471-2342. ; 13:1, s. 16-
- Relaterad länk:
-
https://bmcmedimagin...
-
visa fler...
-
https://umu.diva-por... (primary) (Raw object)
-
https://urn.kb.se/re...
-
https://doi.org/10.1...
-
https://urn.kb.se/re...
-
visa färre...
Abstract
Ämnesord
Stäng
- Background: Tissue Velocity Imaging (TVI) is an ultrasound based technique used for quantitative analysis of the cardiac function and has earlier been evaluated according to myocardial velocities. Recent years several studies have reported applying TVI in the analysis of skeletal muscles. Skeletal tissue velocities can be very low. In particular, when performing isometric contractions or contractions of low force level the velocities may be much lower compared to the myocardial tissue velocities. Methods: In this study TVI was evaluated for estimation of tissue velocities below the typical myocardial velocities. An in-house phantom was used to see how different PRF-settings affected the accuracy of the velocity estimations. Results: With phantom peak velocity at 0.03 cm/s the error ranged from 31% up to 313% with the different PRF-settings in this study. For the peak velocities at 0.17 cm/s and 0.26 cm/s there was no difference in error with tested PFR settings, it is kept approximately around 20%. Conclusions: The results from the present study showed that the PRF setting did not seem to affect the accuracy of the velocity estimation at tissue velocities above 0.17 cm/s. However at lower velocities (0.03 cm/s) the setting was crucial for the accuracy. The PRF should therefore preferable be reduced when the method is applied in low-level muscle contraction.
Ämnesord
- TEKNIK OCH TEKNOLOGIER -- Medicinteknik (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Medical Engineering (hsv//eng)
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
Nyckelord
- Tissue Velocity Imaging
- Ultrasound
- Skeletal muscle
- Phantom evaluation
- Pulse repetition frequency
Publikations- och innehållstyp
- ref (ämneskategori)
- art (ämneskategori)
Hitta via bibliotek
Till lärosätets databas