Standard orientation of the pelvis : Validation on a model and ten patients
-
- Olivecrona, Lotta (författare)
- Karolinska Institutet, Department of Clinical Science, Intervention and Technology, Division of Radiology
-
- Aspelin, Peter (författare)
- Karolinska Institutet
-
- Weidenhielm, Lars (författare)
- Karolinska Institutet
-
visa fler...
-
- Stark, Andreas (författare)
- Karolinska Institutet
-
- Noz, Marilyn E. (författare)
- New York University, Department of Radiology,Nuclear Medicine
-
- Maguire Jr., Gerald Q. (författare)
- KTH,Kommunikationssystem, CoS,Computer Communications
-
- Zeleznik, Michael P. (författare)
- RAHD Oncol Prod, St Louis, MO USA
-
- Svensson, Lars E. (författare)
- KTH,Matematik (Inst.)
-
- Olivecrona, Henrik (författare)
- Soder Sjukhuset, Dept Hand Surgery
-
visa färre...
-
(creator_code:org_t)
- 2016-08-30
- 2005
- Engelska.
-
Ingår i: Acta Radiologica. - London, UK : Informa Healthcare. - 0284-1851 .- 1600-0455. ; 46:1, s. 74-82
- Relaterad länk:
-
https://urn.kb.se/re...
-
visa fler...
-
https://doi.org/10.1...
-
http://kipublication...
-
visa färre...
Abstract
Ämnesord
Stäng
- Purpose: To validate an image post-processing method for re-orienting the pelvis in CT volumes to a standardized orientation in a model and in 10 patients. Material and Methods: Twenty-four CT volumes of a pelvic model and 10 pairs of postoperative total hip arthroplasty (THA) patient CT scans were rotated to a defined pelvic standard orientation and the rotation was recorded. For precision, a test-retest procedure was used. For accuracy, three exactly represented coordinate points were used. For clinical application, the standard orientation was used for calculating the direction of acetabular cup migration from a previous model study. Results: Precision of pelvic standard orientation, calculated as maximal directional error, was better than 1degrees in the model study and better than 1.5degrees in the patient study. Accuracy, expressed as angle between ideal and measured coordinate axes, was 0.1degrees for x, y, z axes. No measurable systematic errors were found. When applied to acetabular cup migration in the model, standardization of pelvic orientation had no significant effect on the measurements. Conclusion: Reorienting the pelvis during image post-processing was shown to be accurate. It enables measurements relative to the pelvis and minimizes the dependency of patient positioning.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
Nyckelord
- CT
- hip
- arthroplasty
- total hip-arthroplasty
- acetabular component migration
- semiautomated program
- computed-tomography
- ct
- fusion
- cup
- anteversion
- alignment
- SRA - ICT