Search: id:"swepub:oai:DiVA.org:kth-22644" >
Model studies on ac...
Model studies on acetabular component migration in total hip arthroplasty using CT and a semiautomated program for volume merging
-
- Olivecrona, Lotta (author)
- Karolinska Hosp, Dept Radiology
-
- Olivecrona, Henrik (author)
- Soder Sjukhuset, Dept Hand Surg, Stockholm, Sweden
-
- Weidenhielm, Lars (author)
- Karolinska Institutet
-
show more...
-
- Noz, Marilyn E. (author)
- New York University, Department of Radiology,Nuclear Medicine
-
- Maguire Jr., Gerald Q. (author)
- KTH,Mikroelektronik och informationsteknik, IMIT
-
- Zeleznik, Michael P. (author)
- RAHD Oncology Products, St. Louis, MO, USA
-
show less...
-
(creator_code:org_t)
- SAGE Publications, 2003
- 2003
- English.
-
In: Acta Radiologica. - : SAGE Publications. - 0284-1851 .- 1600-0455. ; 44:4, s. 419-429
- Related links:
-
https://urn.kb.se/re...
-
show more...
-
https://doi.org/10.1...
-
http://kipublication...
-
show less...
Abstract
Subject headings
Close
- Purpose: Validation of a non-invasive CT method for detection of acetabular cup migration after total hip arthroplasty in a phantom study. Material and Methods: 26 CT examinations were obtained of a pelvic model while altering the position of the acetabular cup. Using a previously described program for volume merging, the pelvi in different examinations were fused and the 3D alterations of the position of the acetabular cup were evaluated visually and numerically and correlated to direct measurements on the model. Results: Visually, two independent examiners differentiated between 0, 1 and 2 to 3 mm migration with 100% specificity and sensitivity. Numerically, the mean error over all cases between model and CT measurements was 0.04 mm (SD +/- 0.33). The mean absolute error between model and CT data was 0.26 mm (SD +/- 0.19). Intra- and interobserver 95% accuracy and repeatability limits were below 0.5/0.7 mm, respectively. No significant interobserver difference occurred. The data were normally distributed and not dependent on observer. Conclusion: The accuracy of this non-invasive method out-performs routine plain radiography. The method gives both visual and numerical correlates to migration and can be used in clinical practice.
Subject headings
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Ortopedi (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Orthopaedics (hsv//eng)
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
Keyword
- hip prosthesis
- volume merging
- migration
- CT
- experimental
- radiostereometry
- accuracy
- SRA - ICT
- SRA - Informations- och kommunikationsteknik
Publication and Content Type
- ref (subject category)
- art (subject category)
Find in a library
To the university's database