- We describe the creation and characterization of a calibration CT mini-lung-phantom incorporating simulated airways and ground-glass densities. Ten duplicate mini-lung-phantoms with Three-Dimensional (3-D) printed tubes simulating airways and gradated density polyurethane foam blocks were designed and built. Dimensional accuracy and CT numbers were measured using micro-CT and clinical CT scanners. Micro-CT images of airway tubes demonstrated an average dimensional variation of 0.038 mm from nominal values. The five different densities of incorporated foam blocks, simulating ground-glass, showed mean CT numbers (±standard deviation) of −897.0 ± 1.5, −844.1 ± 1.5, −774.1 ± 2.6, −695.3 ± 1.6, and −351.0 ± 3.7 HU, respectively. Three-Dimensional printing and subtractive manufacturing enabled rapid, cost-effective production of ground-truth calibration mini-lung-phantoms with low inter-sample variation that can be scanned simultaneously with the patient undergoing lung quantitative CT.