Photon-counting detector CT and energy-integrating detector CT for trabecular bone microstructure analysis of cubic specimens from human radius

Klintström, Benjamin (författare)
KTH,Medicinsk avbildning,KTH Royal Inst Technol, Sweden
Henriksson, Lilian (författare)
Linköpings universitet,Avdelningen för diagnostik och specialistmedicin,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Region Östergötland, Röntgenkliniken i Linköping
Moreno, Rodrigo, 1973- (författare)
KTH,Medicinsk avbildning,KTH Royal Inst Technol, Sweden
visa fler...
Malusek, Alexandr (författare)
Linköpings universitet,Avdelningen för diagnostik och specialistmedicin,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV
Smedby, Örjan, Professor, 1956- (författare)
KTH,Medicinsk avbildning,KTH Royal Inst Technol, Sweden
Woisetschläger, Mischa (författare)
Linköpings universitet,Avdelningen för diagnostik och specialistmedicin,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Region Östergötland, Röntgenkliniken i Linköping
Klintström, Eva (författare)
Linköpings universitet,Avdelningen för diagnostik och specialistmedicin,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Region Östergötland, Röntgenkliniken i Linköping
visa färre...
 (creator_code:org_t)
2022-07-27
2022
Engelska.
Ingår i: European radiology experimental. - : Springer Nature. - 2509-9280. ; 6:1
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Background As bone microstructure is known to impact bone strength, the aim of this in vitro study was to evaluate if the emerging photon-counting detector computed tomography (PCD-CT) technique may be used for measurements of trabecular bone structures like thickness, separation, nodes, spacing and bone volume fraction. Methods Fourteen cubic sections of human radius were scanned with two multislice CT devices, one PCD-CT and one energy-integrating detector CT (EID-CT), using micro-CT as a reference standard. The protocols for PCD-CT and EID-CT were those recommended for inner- and middle-ear structures, although at higher mAs values: PCD-CT at 450 mAs and EID-CT at 600 (dose equivalent to PCD-CT) and 1000 mAs. Average measurements of the five bone parameters as well as dispersion measurements of thickness, separation and spacing were calculated using a three-dimensional automated region growing (ARG) algorithm. Spearman correlations with micro-CT were computed. Results Correlations with micro-CT, for PCD-CT and EID-CT, ranged from 0.64 to 0.98 for all parameters except for dispersion of thickness, which did not show a significant correlation (p = 0.078 to 0.892). PCD-CT had seven of the eight parameters with correlations rho > 0.7 and three rho > 0.9. The dose-equivalent EID-CT instead had four parameters with correlations rho > 0.7 and only one rho > 0.9. Conclusions In this in vitro study of radius specimens, strong correlations were found between trabecular bone structure parameters computed from PCD-CT data when compared to micro-CT. This suggests that PCD-CT might be useful for analysing bone microstructure in the peripheral human skeleton.

Ämnesord

TEKNIK OCH TEKNOLOGIER  -- Medicinteknik -- Medicinsk material- och protesteknik (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering -- Medical Materials (hsv//eng)
TEKNIK OCH TEKNOLOGIER  -- Medicinteknik -- Medicinsk bildbehandling (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Medical Engineering -- Medical Image Processing (hsv//eng)

Nyckelord

Cancellous bone
Osteoporosis
Radius
Tomography (x-ray computed)
X-ray microtomography

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy