Validation of a 3D CT method for measurement of linear wear of acetabular cups : A hip simulator study

Jedenmalm, Anneli (författare)
Lunds universitet,KTH,Fiber- och polymerteknologi,Lund University, Sweden,KTH, Fiber- och polymerteknologi,Ortopedi, Lund,Sektion III,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Orthopaedics (Lund),Section III,Department of Clinical Sciences, Lund,Faculty of Medicine
Nilsson, Fritjof (författare)
KTH,Fiber- och polymerteknologi,KTH, Fiber- och polymerteknologi
Noz, Marilyn E. (författare)
New York University, Department of Radiology
visa fler...
Green, Douglas D. (författare)
Loma Linda University, Orthopaedic Research Center
Gedde, Ulf W. (författare)
KTH,Fiber- och polymerteknologi,KTH, Fiber- och polymerteknologi
Clarke, Ian C. (författare)
Loma Linda University, Orthopaedic Research Center
Stark, Andreas (författare)
Karolinska Institutet
Maguire Jr., Gerald Q. (författare)
KTH,Radio Systems Laboratory (RS Lab),Computer Communications,KTH, Radio Systems Laboratory (RS Lab)
Zeleznik, Michael P. (författare)
Saya Systems Inc., Salt Lake City, UT, USA
Olivecrona, Henrik (författare)
Karolinska Institute, Department of Molecular Medicine and Surgery
visa färre...
 (creator_code:org_t)
2011-02
2011
Engelska.
Ingår i: Acta Orthopaedica. - : Medical Journals Sweden AB. - 1745-3674 .- 1745-3682. ; 82:1, s. 35-41
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Material and methods Ultra-high molecular weight polyethylene cups with a titanium mesh molded on the outside were subjected to wear using a hip simulator. Before and after wear, they were (1) imaged with a CT scanner using a phantom model device, (2) measured using a coordinate measurement machine (CMM), and (3) weighed. CMM was used as the reference method for measurement of femoral head penetration into the cup and for comparison with CT, and gravimetric measurements were used as a reference for both CT and CMM. Femoral head penetration and wear vector angle were studied. The head diameters were also measured with both CMM and CT. The repeatability of the method proposed was evaluated with two repeated measurements using different positions of the phantom in the CT scanner. Results The accuracy of the 3D CT method for evaluation of linear wear was 0.51 mm and the repeatability was 0.39 mm. Repeatability for wear vector angle was 17 degrees A degrees. Interpretation This study of metal-meshed hip-simulated acetabular cups shows that CT has the capacity for reliable measurement of linear wear of acetabular cups at a clinically relevant level of accuracy.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Kirurgi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Surgery (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Ortopedi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Orthopaedics (hsv//eng)

Nyckelord

Total Hip-Arthroplasty
Semiautomated Program
Component Migration
Computed-Tomography
Fusion
Repeatability
Penetration
Time
Orthopaedics
Ortopedi
SRA - ICT
SRA - Informations- och kommunikationsteknik

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy