SwePub
Sök i LIBRIS databas

  Utökad sökning

id:"swepub:oai:DiVA.org:kth-47540"
 

Sökning: id:"swepub:oai:DiVA.org:kth-47540" > A Robust Methodolog...

A Robust Methodology for In Vivo T1 Mapping

Barral, Joelle K. (författare)
Stanford University
Gudmundson, Erik (författare)
Uppsala universitet,Avdelningen för systemteknik,Reglerteknik
Stikov, Nikola (författare)
Stanford University
visa fler...
Etezadi-Amoli, Maryam (författare)
Stanford University
Stoica, Peter (författare)
Uppsala universitet,Avdelningen för systemteknik,Reglerteknik
Nishimura, Dwight G. (författare)
Stanford University
visa färre...
 (creator_code:org_t)
2010-06-17
2010
Engelska.
Ingår i: Magnetic Resonance in Medicine. - : Wiley. - 0740-3194 .- 1522-2594. ; 64:4, s. 1057-1067
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • In this article, a robust methodology for in vivo T(1) mapping is presented. The approach combines a gold standard scanning procedure with a novel fitting procedure. Fitting complex data to a five-parameter model ensures accuracy and precision of the T(1) estimation. A reduced-dimension nonlinear least squares method is proposed. This method turns the complicated multi-parameter minimization into a straightforward one-dimensional search. As the range of possible T(1) values is known, a global grid search can be used, ensuring that a global optimal solution is found. When only magnitude data are available, the algorithm is adapted to concurrently restore polarity. The performance of the new algorithm is demonstrated in simulations and phantom experiments. The new algorithm is as accurate and precise as the conventionally used Levenberg-Marquardt algorithm but much faster. This gain in speed makes the use of the five-parameter model viable. In addition, the new algorithm does not require initialization of the search parameters. Finally, the methodology is applied in vivo to conventional brain imaging and to skin imaging. T(1) values are estimated for white matter and gray matter at 1.5T and for dermis, hypodermis, and muscle at 1.5T, 3T, and 7T. Magn Reson Med 64:1057-1067, 2010.

Ämnesord

TEKNIK OCH TEKNOLOGIER  -- Elektroteknik och elektronik -- Signalbehandling (hsv//swe)
ENGINEERING AND TECHNOLOGY  -- Electrical Engineering, Electronic Engineering, Information Engineering -- Signal Processing (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

relaxation-times
magnetization-transfer
tissue
mri
temperature
dependence
skin
t1
parameters
frequency
Signal processing
Radiological research

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy