In-situ calibration of clinical built-in KAP meters with traceability to a primary standard using a reference KAP meter

Malusek, Alexandr (författare)
Linköpings universitet,Avdelningen för radiologiska vetenskaper,Hälsouniversitetet
Helmrot, Ebba (författare)
Östergötlands Läns Landsting,Linköpings universitet,Avdelningen för radiologiska vetenskaper,Hälsouniversitetet,Radiofysikavdelningen US
Sandborg, Michael, 1961- (författare)
Östergötlands Läns Landsting,Linköpings universitet,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Avdelningen för radiologiska vetenskaper,Hälsouniversitetet,Radiofysikavdelningen US
visa fler...
Grindborg, J-E (författare)
Swedish Radiat Protect Author, Sweden
Alm Carlsson, Gudrun (författare)
Östergötlands Läns Landsting,Linköpings universitet,Avdelningen för radiologiska vetenskaper,Hälsouniversitetet,Radiofysikavdelningen US,Centrum för medicinsk bildvetenskap och visualisering, CMIV
visa färre...
 (creator_code:org_t)
2014-11-04
2014
Engelska.
Ingår i: Physics in Medicine and Biology. - : IOP Publishing: Hybrid Open Access. - 0031-9155 .- 1361-6560. ; 59:23, s. 7195-7210
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • The air kerma-area product (KAP) is used for settings of diagnostic reference levels. The International Atomic Energy Agency (IAEA) recommends that doses in diagnostic radiology (including the KAP values) be estimated with an accuracy of at least +/- 7% (k = 2). Industry standards defined by the International Electrotechnical Commission (IEC) specify that the uncertainty of KAP meter measurements should be less than +/- 25% (k = 2). Medical physicists willing to comply with the IAEAs recommendation need to apply correction factors to KAP values reported by x-ray units. The aim of this work is to present and evaluate a calibration method for built-in KAP meters on clinical x-ray units. The method is based on (i) a tandem calibration method, which uses a reference KAP meter calibrated to measure the incident radiation, (ii) measurements using an energy-independent ionization chamber to correct for the energy dependence of the reference KAP meter, and (iii) Monte Carlo simulations of the beam quality correction factors that correct for differences between beam qualities at a standard laboratory and the clinic. The method was applied to the KAP meter in a Siemens Aristos FX plus unit. It was found that values reported by the built-in KAP meter differed from the more accurate values measured by the reference KAP meter by more than 25% for high tube voltages (more than 140 kV) and heavily filtered beams (0.3 mm Cu). Associated uncertainties were too high to claim that the IECs limit of 25% was exceeded. Nevertheless the differences were high enough to justify the need for a more accurate calibration of built-in KAP meters.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

KAP-meter; kerma-area product; calibration

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy