IMPLICATIONS OF PATIENT CENTRING ON ORGAN DOSE IN COMPUTED TOMOGRAPHY
-
- Kataria, Bharti (författare)
- Linköpings universitet,Avdelningen för radiologiska vetenskaper,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Region Östergötland, Röntgenkliniken i Linköping
-
- Sandborg, Michael, 1961- (författare)
- Linköpings universitet,Avdelningen för radiologiska vetenskaper,Medicinska fakulteten,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Region Östergötland, Radiofysikavdelningen US
-
- Nilsson Althen, Jonas (författare)
- Linköpings universitet,Medicinska fakulteten,Avdelningen för radiologiska vetenskaper,Region Östergötland, Radiofysikavdelningen US
-
(creator_code:org_t)
- 2016-01-06
- 2016
- Engelska.
-
Ingår i: Radiation Protection Dosimetry. - : Oxford University Press. - 0144-8420 .- 1742-3406. ; 169:1-4, s. 130-135
- Relaterad länk:
-
https://liu.diva-por... (primary) (Raw object)
-
visa fler...
-
http://liu.diva-port...
-
https://urn.kb.se/re...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- Automatic exposure control (AEC) in computed tomography (CT) facilitates optimisation of dose absorbed by the patient. The use of AEC requires appropriate ‘patient centring’ within the gantry, since positioning the patient off-centre may affect both image quality and absorbed dose. The aim of this experimental study was to measure the variation in organ and abdominal surface dose during CTexaminations of the head, neck/thorax and abdomen. The dose was compared at the isocenter with two off-centre positions—ventral and dorsal to the isocenter. Measurements were made with an anthropomorphic adult phantom and thermoluminescent dosemeters. Organs and surfaces for ventral regions received lesser dose (5.6–39.0 %) than the isocenter when the phantom was positioned 13 cm off-centre. Similarly, organ and surface doses for dorsal regions were reduced by 5.0–21.0 % at 25 cm off-centre. Therefore, correct vertical positioning of the patient at the gantry isocenter is important to maintain optimal imaging conditions.
Ämnesord
- MEDICIN OCH HÄLSOVETENSKAP -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
- MEDICAL AND HEALTH SCIENCES -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)
Nyckelord
- computed tomography
- organ dose
Publikations- och innehållstyp
- ref (ämneskategori)
- art (ämneskategori)
Hitta via bibliotek
Till lärosätets databas