Hemodynamic Evaluation of the Jarvik 2000 Heart During Heart Failure

Myers, T. M. (författare)
Jarvik Heart Inc., USA
Casimir-Ahn, H. (författare)
Jarvik Heart Inc., USA
Lönn, U. (författare)
Jarvik Heart Inc., USA
visa fler...
Peterzén, B. (författare)
Linköpings universitet,Institutionen för medicin och vård,Hälsouniversitetet
Wårdell, Karin (författare)
Linköpings universitet,Institutionen för medicinsk teknik,Hälsouniversitetet
Carnstam, B. (författare)
Texas Heart Institute, USA
Jarvik, R. K. (författare)
Texas Heart Institute, USA
Frazier, O. H. (författare)
Jarvik Heart Inc., USA
visa färre...
Jarvik Heart Inc, USA Institutionen för medicin och vård (creator_code:org_t)
Wolters Kluwer, 2001
2001
Engelska.
Ingår i: ASAIO journal (1992). - : Wolters Kluwer. - 1058-2916 .- 1538-943X. ; 46:2, s. 167-
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • The Jarvik 2000 Heart is an axial flow left ventricular assist system designed for long-term circulatory support. Purpose: To assess the level of hemodynamic support provided by the Jarvik 2000 in an animal model, with acute, severe heart failure. Methods: Heart failure (HF) was induced in 5 calves by progressively ligating the LAD coronary artery. A continuous infusion of esmolol hydrochloride was given to maintain the stability of HF. Compared to normal baseline values, during heart failure CI decreased by 33% (p < 0.02), LAP increased by 16% (p < 0.03), and BP decreased by 28% (p < 0.13). Once stable HF was established, the pump was turned on, and hemodynamic measurements were obtained at pump speed settings that ranged from 8,000 to 12,000 rpm. Results: Selected, mean hemodynamic values are presented in the table, CI, LAP, and BP improved significantly (p < 0.05) at pump speeds >9000 rpm. Pulse pressure (PP) narrowed by 64% (p < 0.009), and the PAP remained unchanged.Image ToolsBlood flow in the femoral artery increased but was not statistically significant. Other hemodynamic parameters such as HR, SVR, PVR, and CVP did not change significantly. Conclusions: The Jarvik 2000 Heart can provide a level of cardiac output support that normalizes the CI and LAP, while maintaining some arterial pressure pulsatility during HF. Systemic perfusion is increased while the left heart is unloaded.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Kardiologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Cardiac and Cardiovascular Systems (hsv//eng)

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy