The microINR portable coagulometer: analytical quality and user-friendliness of a PT (INR) point-of-care instrument

Bukmann Larsen, Pia (författare)
Slagelse Hospital, Denmark
Storjord, Elin (författare)
Nordland Hospital, Norway; UiT, Norway
Bakke, Asne (författare)
Stavanger University Hospital, Norway
visa fler...
Bukve, Tone (författare)
Akershus University Hospital, Norway
Christensen, Mikael (författare)
Aarhus University Hospital, Denmark
Eikeland, Joakim (författare)
Oslo University Hospital, Norway
Engeland Haugen, Vegar (författare)
Haukeland Hospital, Norway
Husby, Kristin (författare)
Akershus University Hospital, Norway
McGrail, Rie (författare)
Aarhus University Hospital, Denmark
Meier Mikaelsen, Solveig (författare)
Sykehuset Innlandet, Norway
Monsen, Grete (författare)
Noklus, Norway
Fogh Moller, Mette (författare)
Herning Hospital, Denmark
Nybo, Jan (författare)
Aalborg University Hospital, Denmark
Revsholm, Jesper (författare)
Randers Regional Hospital, Denmark
Johanne Risoy, Aslaug (författare)
Noklus, Norway; University of Bergen, Norway
Skalsvik, Unni Marie (författare)
Nordland Hospital, Norway
Strand, Heidi (författare)
Akershus University Hospital, Norway
Serrano Teruel, Reyes (författare)
Noklus, Norway
Theodorsson, Elvar (författare)
Linköpings universitet,Avdelningen för mikrobiologi och molekylär medicin,Medicinska fakulteten,Region Östergötland, Klinisk kemi
visa färre...
 (creator_code:org_t)
2017-02-02
2017
Engelska.
Ingår i: Scandinavian Journal of Clinical and Laboratory Investigation. - : TAYLOR & FRANCIS LTD. - 0036-5513 .- 1502-7686. ; 77:2, s. 115-121
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Regular measurement of prothrombin time as an international normalized ratio PT (INR) is mandatory for optimal and safe use of warfarin. Scandinavian evaluation of laboratory equipment for primary health care (SKUP) evaluated the microINR portable coagulometer (microINR((R))) (iLine Microsystems S.L., Spain) for measurement of PT (INR). Analytical quality and user-friendliness were evaluated under optimal conditions at an accredited hospital laboratory and at two primary health care centres (PHCCs). Patients were recruited at the outpatient clinic of the Laboratory of Medical Biochemistry, St Olavs University Hospital, Trondheim, Norway (n=98) and from two PHCCs (n=88). Venous blood samples were analyzed under optimal conditions on the STA-R((R))Evolution with STA-SPA+reagent (Stago, France) (Owren method), and the results were compared to capillary measurements on the microINR((R)). The imprecision of the microINR((R)) was 6% (90% CI: 5.3-7.0%) and 6.3% (90% CI: 5.1-8.3) in the outpatient clinic and PHCC2, respectively for INR 2.5. The microINR((R)) did not meet the SKUP quality requirement for imprecision 5.0%. For INR amp;lt;2.5 at PHCC2 and at both levels in PHCC1, CV% was 5.0. The accuracy fulfilled the SKUP quality goal in both outpatient clinic and PHCCs. User-friendliness of the operation manual was rated as intermediate, defined by SKUP as neutral ratings assessed as neither good nor bad. Operation facilities was rated unsatisfactory, and time factors satisfactory. In conclusion, quality requirements for imprecision were not met. The SKUP criteria for accuracy was fulfilled both at the hospital and at the PHCCs. The user-friendliness was rated intermediate.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Hälsovetenskap -- Hälso- och sjukvårdsorganisation, hälsopolitik och hälsoekonomi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Health Sciences -- Health Care Service and Management, Health Policy and Services and Health Economy (hsv//eng)

Nyckelord

International normalized ratio; point-of-care systems; point-of-care testing; quality control; warfarin

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy