Combined diffuse light reflectance and electric impedance measurements for navigation aid in deep brain surgery

Johansson, Johannes D., 1977- (författare)
Linköpings universitet,Biomedicinsk instrumentteknik,Tekniska högskolan,MINT,Department of Biomedical Engineering, Linköping University, Linköping
Blomstedt, Patric (författare)
Umeå universitet,Neurokirurgi,Department of Neurosurgery, Norrland´s University Hospital, Umeå, Sweden
Haj-Hosseini, Neda (författare)
Linköpings universitet,Biomedicinsk instrumentteknik,Tekniska högskolan,MINT,Department of Biomedical Engineering, Linköping University, Linköping
visa fler...
Bergenheim, A Tommy (författare)
Umeå universitet,Neurokirurgi,Department of Neurosurgery, Norrland´s University Hospital, Umeå, Sweden
Eriksson, Ola (författare)
Linköpings universitet,Biomedicinsk instrumentteknik,Tekniska högskolan,Department of Biomedical Engineering, Linköping University, Linköping
Wårdell, Karin (författare)
Linköpings universitet,Biomedicinsk instrumentteknik,Tekniska högskolan,MINT,Department of Biomedical Engineering, Linköping University, Linköping
visa färre...
 (creator_code:org_t)
2009-02-18
2009
Engelska.
Ingår i: Stereotactic and Functional Neurosurgery. - Basel : S. Karger AG. - 1011-6125 .- 1423-0372. ; 87:2, s. 105-113
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Aim: The aim of this study is to investigate reflected light intensity combined with impedance for navigation aid during stereotactic neurosurgery.Methods: During creation of 21 trajectories for stereotactic implantation of deep brain stimulation electrodes in the globus pallidus internus or subthalamus (zona incerta or subthalamic nucleus), impedance at 512 kHz and reflected light intensity at 780 nm were measured continuously and simultaneously with a radio frequency electrode containing optical fibres. The signals were compared with anatomy determined from pre- and postoperative MRI and CT. The measurements were performed within minutes and signal analysis was done post-operatively.Results: Reflected light intensity was low from cortex, lateral ventricle, caudate nucleus and putamen. It was intermediate from globus pallidus and thalamus while it was high from subcortical white matter, internal capsule and the subthalamus. The electric impedance was less consistent but generally low in the cortex, intermediate in subcortical white matter, the putamen, the globus pallidus and the thalamus and high in the internal capsule and the subthalamus.Conclusion: Reflected light intensity and electric impedance give complementary information about passed tissue and the combination seems promising for navigation aid during stereotactic neurosurgery.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine (hsv//eng)

Nyckelord

Stereotactic surgery
navigation
electric impedance
light reflectance
MEDICINE
MEDICIN

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy