Electrical characteristics of conductive yarns and textile electrodes for medical applications
-
- Rattfält, Linda, 1979- (författare)
- Linköpings universitet,Tekniska högskolan,Fysiologisk mätteknik,Linköping University, Linköping, Sweden
-
- Lindén, Maria (författare)
- Mälardalens högskola,Linköpings universitet,Tekniska högskolan,Biomedicinsk instrumentteknik,Institutionen för datavetenskap och elektronik
-
- Hult, Peter, 1964- (författare)
- Linköpings universitet,Tekniska högskolan,Fysiologisk mätteknik,Linköping University, Linköping, Sweden
-
visa fler...
-
- Berglin, Lena (författare)
- Data- och informationsteknik Göteborgs universitet,University College of Borås, Borås, Sweden
-
- Ask, Per, 1950- (författare)
- Linköpings universitet,Tekniska högskolan,Fysiologisk mätteknik,Linköping University, Linköping, Sweden
-
visa färre...
-
(creator_code:org_t)
- 2007-10-10
- 2007
- Engelska.
-
Ingår i: Medical and Biological Engineering and Computing. - : Springer Science and Business Media LLC. - 0140-0118 .- 1741-0444. ; 45:12, s. 1251-1257
- Relaterad länk:
-
https://urn.kb.se/re...
-
visa fler...
-
https://doi.org/10.1...
-
https://urn.kb.se/re...
-
visa färre...
Abstract
Ämnesord
Stäng
- Clothing with conductive textiles for health care applications has in the last decade been of an upcoming research interest. An advantage with the technique is its suitability in distributed and home health care. The present study investigates the electrical properties of conductive yarns and textile electrodes in contact with human skin, thus representing a real ECG-registration situation. The yarn measurements showed a pure resistive characteristic proportional to the length. The electrodes made of pure stainless steel (electrode A) and 20% stainless steel/80% polyester (electrode B) showed acceptable stability of electrode potentials, the stability of A was better than that of B. The electrode made of silver plated copper (electrode C) was less stable. The electrode impedance was lower for electrodes A and B than that for electrode C. From an electrical properties point of view we recommend to use electrodes of type A to be used in intelligent textile medical applications. © International Federation for Medical and Biological Engineering 2007.
Ämnesord
- TEKNIK OCH TEKNOLOGIER -- Medicinteknik -- Medicinsk laboratorie- och mätteknik (hsv//swe)
- ENGINEERING AND TECHNOLOGY -- Medical Engineering -- Medical Laboratory and Measurements Technologies (hsv//eng)