Measurements before endovascular repair of abdominal aortic aneurysms : MR imaging with MRA vs. angiography and CT

Engellau, Lena (författare)
Department of Radiology, Lund University Hospital, Lund, Sweden
Albrechtsson, U. (författare)
Department of Radiology, Lund University Hospital, Lund, Sweden
Dahlström, Nils (författare)
Department of Radiology, Hudiksvall Hospital, Hudiksvall Sweden
visa fler...
Norgren, L. (författare)
Department of Vascular Diseases, Lund University, Malmo University Hospital, Malmo, Sweden
Persson, Anders, 1953- (författare)
Östergötlands Läns Landsting,Linköpings universitet,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Hälsouniversitetet,Medicinsk radiologi,Röntgenkliniken i Linköping
Larsson, E.-M. (författare)
Department of Radiology, Lund University Hospital, Lund, Sweden
visa färre...
 (creator_code:org_t)
SAGE Publications, 2003
2003
Engelska.
Ingår i: Acta Radiologica. - : SAGE Publications. - 0284-1851 .- 1600-0455. ; 44:2, s. 177-184
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Purpose: 1) To compare measurements obtained with MR imaging (MRI)/contrast-enhanced MR angiography (CE MRA) with measurements obtained with angiography (DSA) and CT, for stent-graft sizing of abdominal aortic aneurysms (AAA). 2) To compare MRA measurements obtained with the two post processing techniques MIP (maximum intensity projection) and VRT (3D volume rendering technique).Material and Methods: The prospective study included 20 consecutive patients with AAA identified by DSA and CT as suitable for endovascular repair. For the study, MRI/CE MRA was performed. Five measurement variables for stent-graft sizing were chosen. Comparisons were made between MRI/CE MRA, DSA and CT, and between observers. Comparisons were also made between MIP and VRT.Results: Significantly shorter lengths were obtained with MRA-MIP than with DSA. Three out of six diameter measurements were significantly smaller on MRI/CE MRA than on DSA and CT. No significant differences were found between the observers. One diameter measurement was significantly smaller on MIP than on VRT, while the other measurements showed no significant differences.Conclusion: The length measurements obtained with MRA-MIP were probably more correct than those with DSA. For more reliable diameter measurements with CE MRA, improvements of the technique, including VRT reconstructions and a standardized determination of the vessel boundaries, are needed.

Nyckelord

Abdominal aortic aneurysm
Angiography
CT
Endograft sizing
MR angiography
Volume rendering technique
NATURAL SCIENCES
NATURVETENSKAP

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy