The limiting energy resolution of SiC detectors in ion spectrometry

Strokan, N.B. (författare)
Ioffe Physicotechnical Institute, Russian Academy of Sciences, St. Petersburg, 194021, Russian Federation
Ivanov, A.M. (författare)
Ioffe Physicotechnical Institute, Russian Academy of Sciences, St. Petersburg, 194021, Russian Federation
Lebedev, A.A. (författare)
Ioffe Physicotechnical Institute, Russian Academy of Sciences, St. Petersburg, 194021, Russian Federation
visa fler...
Syväjärvi, Mikael (författare)
Linköpings universitet,Tekniska högskolan,Halvledarmaterial
Yakimova, Rositsa (författare)
Linköpings universitet,Tekniska högskolan,Halvledarmaterial
visa färre...
Ioffe Physicotechnical Institute, Russian Academy of Sciences, St Petersburg, 194021, Russian Federation Tekniska högskolan (creator_code:org_t)
Pleiades Publishing Ltd, 2005
2005
Engelska.
Ingår i: Semiconductors (Woodbury, N.Y.). - : Pleiades Publishing Ltd. - 1063-7826 .- 1090-6479. ; 39:12, s. 1420-1425
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • The Monte Carlo method is used to simulate the complete stopping of a particles in SiC. A histogram of energy losses in nuclear-scattering events is obtained. The energy-loss spectrum has the characteristic asymmetric shape with the line full width at the half-maximum FWHMnucl ˜ 4. 22 keV. The final shape of the spectral line is obtained by a convolution with the Gaussian function that describes the contribution of the ionization and noise fluctuations (originated in the detector and instrumentation) to the signal. The resulting value of FWHM for the line is equal to 8.75 keV (at a noise variance of 1.7 keV). The experimental energy resolution of the detectors was found to be poorer than the calculated value by a factor of 2. It is established that the losses of charge during its transport in the detector bulk are insignificant, so that the discrepancy between the calculated and experimental values of the resolution should be attributed to the nonoptimal design of the detector window. © 2005 Pleiades Publishing, Inc.

Nyckelord

TECHNOLOGY
TEKNIKVETENSKAP

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy