Evaluating doppler devices using a moving string test target
-
- Walker, Andrew (författare)
- Linköpings universitet,Fysiologisk mätteknik,Tekniska högskolan
-
- Philips, David J. (författare)
- Department of Surgery and Center for Bioengineering, University of Washington, Seattle
-
- Powers, Jeffry E. (författare)
- Department of Surgery and Center for Bioengineering, University of Washington, Seattle
-
(creator_code:org_t)
- Wiley InterScience, 1982
- 1982
- Engelska.
-
Ingår i: Journal of Clinical Ultrasound. - : Wiley InterScience. - 0091-2751 .- 1097-0096. ; 10:1, s. 25-30
- Relaterad länk:
-
http://urn.kb.se/res...
-
visa fler...
-
http://urn.kb.se/res...
-
https://urn.kb.se/re...
-
https://doi.org/10.1...
-
visa färre...
Abstract
Ämnesord
Stäng
- Two moving string test targets have been developed and used to characterize ultrasound Doppler instrumentation. The characteristics investigated were sample volume size and location. The tests were performed on a combined echo/Doppler instrument and on an annular array system. The procedures can be carried out routinely in the clinical laboratory to ensure that the instrument is working properly or as an aid for correct interpretation of acquired data.