Post-processing image filtration enabling dose reduction in standard abdominal CT

Leander, Peter (författare)
Lund University,Lunds universitet,Diagnostisk radiologi, Malmö,Forskargrupper vid Lunds universitet,Radiology Diagnostics, Malmö,Lund University Research Groups
Söderberg, Marcus (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups,Malmö University Hospital
Falt, Tobias (författare)
Malmö University Hospital
visa fler...
Albertsson, Ida (författare)
Linköpings universitet,Institutionen för klinisk och experimentell medicin,Hälsouniversitetet
Gunnarsson, Mikael (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Malmö,Forskargrupper vid Lunds universitet,Medical Radiation Physics, Malmö,Lund University Research Groups
visa färre...
 (creator_code:org_t)
2010-03-05
2010
Engelska.
Ingår i: RADIATION PROTECTION DOSIMETRY. - : Oxford University Press. - 0144-8420 .- 1742-3406. ; 139:1-3, s. 180-185
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • The collective effective radiation dose to the population is increasing due to a higher use of computerised tomography. SharpView AB, Linkoping, Sweden, has developed an adaptive non-linear post-processing image filtration that may enable the use of lower radiation doses. The present study assessed if a lower dose with image filtration had the same image quality as a higher dose without the filter applied. All imaging was performed on a Siemens Somatom Sensation 16 CT. The parameters used were 120 kV and 200 mAs (40 patients) and 130 mAs without and with image filtering (40 patients), respectively. All studies were quantitatively evaluated for noise and image quality was assessed by visual grading characteristics (VGC) analysis. After image filtration, the noise in the processed images was lowered and the image quality was improved as shown by the VGC analysis. However, images using the higher dose were still ranked as the best in five out of eight criteria as shown by the VGC analysis. Image filtration enhances CT images significantly and further studies will show if 130 mAs with image filtration may be sufficient for clinically general abdominal CT.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Nyckelord

TECHNOLOGY
TEKNIKVETENSKAP

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy