Influence of the FID and off-resonance effects in dense MRI

Haraldsson, Henrik, 1977- (författare)
Linköpings universitet,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Klinisk fysiologi,Hälsouniversitetet
Sigfridsson, Andreas (författare)
Östergötlands Läns Landsting,Linköpings universitet,Klinisk fysiologi,Hälsouniversitetet,Fysiologiska kliniken
Sakuma, Hajime (författare)
Department of Radiology, Mie University, Japan
visa fler...
Engvall, Jan, 1953- (författare)
Östergötlands Läns Landsting,Linköpings universitet,Centrum för medicinsk bildvetenskap och visualisering, CMIV,Klinisk fysiologi,Hälsouniversitetet,Fysiologiska kliniken US
Ebbers, Tino (författare)
Östergötlands Läns Landsting,Linköpings universitet,Hälsouniversitetet,Fysiologiska kliniken,Mekanisk värmeteori och strömningslära,Fysiologi,Centrum för medicinsk bildvetenskap och visualisering, CMIV
visa färre...
 (creator_code:org_t)
2010-12-02
2011
Engelska.
Ingår i: Magnetic Resonance in Medicine. - : Wiley. - 0740-3194 .- 1522-2594. ; 65:4, s. 1104-1112
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Accurate functional measurement in cardiovascular diseases is important as inaccuracy may compromise diagnostic decisions. Cardiac function can be assessed using displacement encoding with stimulated echoes, resulting in three signal components. The free induction decay (FID), arising from spins undergoing T1-relaxation, is not displacement encoded and impairs the displacement acquired. Techniques for suppressing the FID exist; however, a residual will remain. The effect of the residual is difficult to distinguish and investigate in vitro and in vivo. In this work, the influence of the FID as well as of off-resonance effects is evaluated by altering the phase of the FID in relation to the stimulated echo. The results show that the FID and off-resonance effects can impair the accuracy of the displacement measurement acquired. The influence of the FID can be avoided by using an encoded reference. We therefore recommend the assessment of this influence of the FID for each displacement encoding with stimulated echoes protocol.

Nyckelord

MEDICINE
MEDICIN

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy