Is there an underestimation of intima-media thickness based on M-mode ultrasound technique in the abdominal aorta?

Dahlén, Elsa M (författare)
Linköpings universitet,Allmänmedicin,Hälsouniversitetet
Andreasson, Thomas (författare)
Sahlgrenska University Hospital, Göteborg
Cinthio, Magnus (författare)
Lund University,Lunds universitet,Avdelningen för Biomedicinsk teknik,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH,Electrical Measurements, Faculty of Engineering, LTH, Lund University
visa fler...
Nyström, Fredrik (författare)
Östergötlands Läns Landsting,Linköpings universitet,Internmedicin,Hälsouniversitetet,Endokrin- och magtarmmedicinska kliniken US
Östgren, Carl Johan (författare)
Östergötlands Läns Landsting,Linköpings universitet,Allmänmedicin,Hälsouniversitetet,Primärvården i västra länsdelen
Länne, Toste (författare)
Östergötlands Läns Landsting,Linköpings universitet,Fysiologi,Hälsouniversitetet,Thorax-kärlkliniken i Östergötland
visa färre...
 (creator_code:org_t)
Wiley-Blackwell, 2012
2012
Engelska.
Ingår i: Clinical Physiology and Functional Imaging. - : Wiley-Blackwell. - 1475-0961 .- 1475-097X. ; 32:1, s. 1-4
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Measuring intima-media thickness (IMT) in the common carotid artery (CCA) is a valuable resource for the evaluation of subclinical atherosclerosis. The main objective of this study was to explore whether a B-mode ultrasound technique, Philips ATL, and an M-mode ultrasound technique, Wall Track System (WTS), show interchangeable results when measured in CCA and the abdominal aorta (AA). A total of 24 healthy, young subjects were examined. IMT and lumen diameter (LD) of the AA and the CCA were measured twice by two skilled ultrasonographers with two different ultrasound equipment B-mode: (Philips, ATL and M-mode: WTS).The intra-observer variability of IMT in CCA and AA using B-mode showed a coefficient of variation 8% and 9%, and with M-mode 11% and 15%, respectively. Interobserver variability of IMT in CCA and AA using B-mode was 6% and 12%, and with M-mode 11% and 18%, respectively. CCA IMT was 0·53 ± 0·07 and 0·53 ± 0·09 mm using B-mode and M-mode, respectively. However, in AA, IMT was 0·61 ± 0·5 and 0·54 ± 0·10 mm using B-mode and M-mode, respectively. Thus, AA IMT was 11·5% thicker using B-mode (P<0·01). We received adequate IMT readings from the carotid artery as well as the AA using two commonly used B-mode and M-mode techniques. B-mode technique seems to show less variability, especially in the AA. More importantly, the two techniques measured different IMT thickness in the aorta, emphasizing the importance of using similar technique when comparing the impact of absolute values of IMT on cardiovascular disease.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Medicinska och farmaceutiska grundvetenskaper -- Fysiologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Basic Medicine -- Physiology (hsv//eng)

Nyckelord

MEDICINE
MEDICIN

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy