SwePub
Sök i LIBRIS databas

  Utökad sökning

id:"swepub:oai:DiVA.org:liu-79790"
 

Sökning: id:"swepub:oai:DiVA.org:liu-79790" > Assessment of the a...

Assessment of the accuracy of MRI wall shear stress estimation using numerical simulations

Petersson, Sven (författare)
Linköpings universitet,Avdelningen för kardiovaskulär medicin,Hälsouniversitetet
Dyverfeldt, Petter (författare)
Linköpings universitet,Klinisk fysiologi,Hälsouniversitetet,Centrum för medicinsk bildvetenskap och visualisering, CMIV
Ebbers, Tino (författare)
Östergötlands Läns Landsting,Linköpings universitet,Klinisk fysiologi,Hälsouniversitetet,Fysiologiska kliniken US,Centrum för medicinsk bildvetenskap och visualisering, CMIV
 (creator_code:org_t)
2012-02-15
2012
Engelska.
Ingår i: Journal of Magnetic Resonance Imaging. - : Wiley-Blackwell. - 1053-1807 .- 1522-2586. ; 36:1, s. 128-138
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Purpose: To investigate the accuracy of wall shear stress (WSS) estimation using MRI. Specifically, to investigate the impact of different parameters and if MRI WSS estimates are monotonically related to actual WSS. Materials and Methods: The accuracy of WSS estimation using methods based on phase-contrast (PC) MRI velocity mapping, Fourier velocity encoding (FVE) and intravoxel velocity standard deviation mapping were studied using numerical simulations. The influence of spatial resolution, velocity encoding, wall segmentation, and voxel location were investigated over a range of WSS values. Results: WSS estimates were found to be sensitive to parameter settings in general and spatial resolution in particular. All methods underestimated WSS, except for the FVE-based method, which instead was extremely sensitive to voxel position relative to the wall. Methods using PC-based WSS estimation with wall segmentation showed to be accurate for low WSS, but were sensitive to segmentation errors. Conclusion: Even in the absence of noise and for relatively simple velocity profiles, MRI WSS estimates cannot always be assumed to be linearly or even monotonically related to actual WSS. High WSS values cannot be resolved and the estimates depend on parameter setting. Nevertheless, distinguishing areas of low and moderate WSS may be feasible.

Nyckelord

wall shear stress; phase contrast magnetic resonance imaging; atherosclerosis; hemodynamics; Fourier velocity encoding; numerical simulations
MEDICINE
MEDICIN

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Petersson, Sven
Dyverfeldt, Pett ...
Ebbers, Tino
Artiklar i publikationen
Journal of Magne ...
Av lärosätet
Linköpings universitet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy