Comparison of transendocardial and retrograde coronary venous intramyocardial catheter delivery systems in healthy and infarcted pigs

Baklanov, D. V. (författare)
Dartmouth Medical School, Hanover, NH, USA
Moodie, K. M. (författare)
Dartmouth Medical School, Hanover, NH, USA
McCarthy, F. E. (författare)
Dartmouth Medical School, Hanover, NH, USA
visa fler...
Mandrusov, E. (författare)
Guidant Corp., Santa Clara, CA, USA
Chiu, J. (författare)
Guidant Corp., Santa Clara, CA, USA
Aswonge, G. (författare)
Guidant Corp., Santa Clara, CA, USA
Cheng, J. (författare)
Guidant Corp., Santa Clara, CA, USA
Chow, M. (författare)
Guidant Corp., Santa Clara, CA, USA
Simons, M. (författare)
Dartmouth Medical School, Hanover, NH, USA
de Muinck, Ebo D. (författare)
Dartmouth Medical School, Hanover, NH, USA
visa färre...
Dartmouth Medical School, Hanover, NH, USA Guidant Corp, Santa Clara, CA, USA (creator_code:org_t)
2006
2006
Engelska.
Ingår i: Catheterization and cardiovascular interventions. - : Wiley-Blackwell. - 1522-1946 .- 1522-726X. ; 68:3, s. 416-423
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • We compared two routes for myocardial delivery of therapeutics, transendocardial (TE) delivery with an intramyocardial injection catheter, and retrograde coronary venous (RCV) delivery with a balloon occlusion catheter in the interventricular vein. Methods: TE and RCV injection of 15 mu m, neutron-activatable microspheres was compared in healthy pigs (Group I, n = 3), pigs with a 1-week-old myocardial infarction (MI; group II, n = 5), and pigs with a 2-weeks-old MI (group III, n = 4). The MI was induced by a 1-hr balloon occlusion in the LAD. Both methods were compared in the same animal using different microspheres. The RCV catheter allowed for continuous measurement of distal pressure and 2.5 x 10(6) microspheres were injected in 10 ml at 300 mmHg above balloon occlusion pressure. The TE injections were targeted to the infarct zone and 2.5 x 10(6) microspheres were distributed over 10 injections of 200 mu l. Results: The retention of microspheres decreased with increase in MI age, but was comparable between devices within the groups. RCV delivery resulted in (14.3 +/- 0.9)% microsphere retention in Group I, (10.3 +/- 0.2)% in Group II, and (6.4 +/- 0.1)% in group III (P less than 0.05 versus group I). Microsphere retention after TE was (15.1 +/- 0.7)% in group I, (18.9 +/- 0.6)% in group II, (4.1 +/- 0.1)% in Group III (P less than 0.05 versus groups I and II). The RCV catheter delivered primarily to midventricular, antero-septal segments, whereas TE targeted apical areas predominantly. Conclusions: Delivery efficacy was comparable between devices in each group however RCV targeted midventricular areas whereas TE targeted apical areas.

Nyckelord

myocardial infarction; percutaneous cardiac intervention; microspheres; myocardial uptake

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy