The impact of a modified cutting flute implant design on osseointegration

Jimbo, Ryo (författare)
Malmö högskola,Odontologiska fakulteten (OD)
Tovar, N. (författare)
Marin, C. (författare)
visa fler...
Teixeira, H. S. (författare)
Anchieta, R. B. (författare)
Silveira, L. M. (författare)
Janal, M. N. (författare)
Shibli, J. A. (författare)
Coelho, P. G. (författare)
visa färre...
 (creator_code:org_t)
Elsevier, 2014
2014
Engelska.
Ingår i: International Journal of Oral and Maxillofacial Surgery. - : Elsevier. - 0901-5027 .- 1399-0020. ; 43:7, s. 883-888
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Information concerning the effects of the implant cutting flute design on initial stability and its influence on osseointegration in vivo is limited. This study evaluated the early effects of implants with a specific cutting flute design placed in the sheep mandible. Forty-eight dental implants with two different macrogeometries (24 with a specific cutting flute design Blossom group; 24 with a self-tapping design - DT group) were inserted into the mandibular bodies of six sheep; the maximum insertion torque was recorded. Samples were retrieved and processed for histomorphometric analysis after 3 and 6 weeks. The mean insertion torque was lower for Blossom implants (P < 0.001). No differences in histomorphometric results were observed between the groups. At 3 weeks, P = 0.58 for bone-to-implant contact (BIC) and P = 0.52 for bone area fraction occupied (BAFO); at 6 weeks, P = 0.55 for BIC and P = 0.45 for BAFO. While no histomorphometric differences were observed, ground sections showed different healing patterns between the implants, with better peri-implant bone organization around those with the specific cutting flute design (Blossom group). Implants with the modified cutting flute design had a significantly reduced insertion torque compared to the DT implants with a traditional cutting thread, and resulted in a different healing pattern.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Odontologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Dentistry (hsv//eng)

Nyckelord

implants
cutting flute design
macro-geometry
thread design
insertion torque
histomorphometry
primary stability
secondary stability
osseointegration

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy