Multiplexed MALDI-MS arrays for screening of MIP solid phase extraction materials

Jagadeesan, Kishore (författare)
Lund University,Lunds universitet,Avdelningen för Biomedicinsk teknik,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH
Wierzbicka, Celina (författare)
Malmö högskola,Institutionen för biomedicinsk vetenskap (BMV)
Laurell, Thomas (författare)
Lund University,Lunds universitet,Avdelningen för Biomedicinsk teknik,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH
visa fler...
Sellergren, Börje (författare)
Malmö högskola,Institutionen för biomedicinsk vetenskap (BMV)
Shinde, Sudhirkumar (författare)
Malmö högskola,Institutionen för biomedicinsk vetenskap (BMV)
Ekström, Simon (författare)
Lund University,Lunds universitet,Avdelningen för Biomedicinsk teknik,Institutionen för biomedicinsk teknik,Institutioner vid LTH,Lunds Tekniska Högskola,Department of Biomedical Engineering,Departments at LTH,Faculty of Engineering, LTH
visa färre...
 (creator_code:org_t)
Elsevier, 2016
2016
Engelska.
Ingår i: Journal of chromatography. B. - : Elsevier. - 1570-0232 .- 1873-376X. ; 1021, s. 213-220
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Technology that facilitates rapid investigation of solid phase extraction protocols using very small amounts of sorbent can save both time and money. The microfabricated ISET (Integrated Selective Enrichment Target) interfaced with MALDI mass spectrometry is able to provide an efficient, economic and generic optimization process for SPE sample preparation. The SPE is performed in a rapid and parallel fashion, with a processing time off only 2h per ISET with 96 samples. Each of the 96 wells on the ISET can hold 600nL of SPE sorbent. The ability to work with small amounts of sorbent and samples in the ISET platform provides a big advantage when developing affinity sorbents, such as molecularly imprinted polymers (MIPs). Here it is demonstrated that an amount of 25mg phosphoserine imprinted MIP (pS-MIP) sorbent can allow for analysis of more than 500 ISET nanovials using a multitude of different conditions. In the presented case, the multiplexed experiments allowed for early discovery of unspecific interactions and subsequent minimization of these, resulting in a protocol that provided improved enrichment of phosphopeptides.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Medicinsk bioteknologi -- Biomedicinsk laboratorievetenskap/teknologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Medical Biotechnology -- Biomedical Laboratory Science/Technology (hsv//eng)

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy