Intercomparison of lateral scales of scanning electron microscopes and atomic force microscopes in research institutes in Northern Europe

Seppä, Santeri Jeremias (författare)
MIKES Centre for Metrology and Accreditation, Finland
Korpelainen, Virpi (författare)
MIKES Centre for Metrology and Accreditation, Finland
Bergstrand, Sten (författare)
RISE,Massa, kraft, längd och tryck
visa fler...
Karlsson, Helge (författare)
Justervesenet, Norway
Lillepea, Lauri (författare)
AS Metrosert, Estonia
Lassila, Antti (författare)
MIKES Centre for Metrology and Accreditation, Finland
visa färre...
 (creator_code:org_t)
2014-03-05
2014
Engelska.
Ingår i: Measurement science and technology. - : IOP Publishing. - 0957-0233 .- 1361-6501. ; 25:4
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • An intercomparison of lateral scales of scanning electron microscopes (SEM) and atomic force microscopes (AFM) in various research laboratories in Northern Europe was organized by the local national metrology institutes. In this paper are presented the results of the comparison, with also an example uncertainty budget for AFM grating pitch measurement. Grating samples (1D) were circulated among the participating laboratories. The participating laboratories were also asked about the calibration of their instruments. The accuracy of the uncertainty estimates seemed to vary largely between the laboratories, and for some laboratories the appropriateness of the calibration procedures could be considered. Several institutes (60% of all results in terms of En value) also had good comprehension of their measurement capability. The average difference from reference value was 6.7 and 10.0 nm for calibrated instruments and 20.6 and 39.9 nm for uncalibrated instruments for 300 nm and 700 nm gratings, respectively. The correlation of the results for both nominally 300 and 700 nm gratings shows that a simple scale factor calibration would have corrected a large part of the deviations from the reference values.

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy