Operating a triple stack microchannel plate-phosphor assembly for single particle counting in the 12-300 K temperature range

Rosén, Stefan (författare)
Stockholms universitet,Fysikum
Schmidt, Henning (författare)
Stockholms universitet,Fysikum
Reinhed, Peter (författare)
Stockholms universitet,Fysikum
visa fler...
Fischer, Daniel (författare)
Stockholms universitet,Fysikum
Thomas, Richard (författare)
Stockholms universitet,Fysikum
Cederquist, Henrik (författare)
Stockholms universitet,Fysikum
Liljeby, Leif (författare)
Stockholms universitet,Manne Siegbahn-laboratoriet
Bagge, Lars (författare)
Stockholms universitet,Manne Siegbahn-laboratoriet
Leontein, Sven (författare)
Stockholms universitet,Manne Siegbahn-laboratoriet
Blom, Mikael (författare)
Stockholms universitet,Manne Siegbahn-laboratoriet
visa färre...
 (creator_code:org_t)
American Institute of Physics, 2007
2007
Engelska.
Ingår i: Review of Scientific Instruments. - : American Institute of Physics. - 0034-6748 .- 1089-7623. ; 78:11, s. 113301-
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • An assembly consisting of a stack of three microchannel plates (MCPs) and a phosphor screen anode has been operated over the temperature range from 300 to 12 K. We report on measurements at 6.4 kHz (using an alpha source) and with dark counts only (15 Hz). Without any particle source,  the MCP bias current decreased by a factor of 2.1×103 when the temperature was lowered from 300 to 12 K. Using the alpha source, and a photomultiplier tube (PMT) to monitor the phosphor screen anode, we first observed an increase in the decay time of the phosphor from 12 to 45 μs when the temperature was decreased from 300 to 100 K while the decay time then decreased and reached a value of 5 μs at 12 K. The pulse height distribution from the PMT was measured between300 and 12 K and shows a spectrum typical for a MCP phosphor setup at 300 K and 12 K but is strongly degraded for intermediate temperatures. We conclude that the present MCP-phosphor detector assembly is well suited for position-sensitive particle counting operation at temperatures down to at least 12 K even for count rates beyond 6 kHz. This result is crucial and an important part of ongoing developments of new instrumentation for investigations of, e.g., interactions involving complex molecular ions with internal quantum state control.

Ämnesord

NATURVETENSKAP  -- Fysik (hsv//swe)
NATURAL SCIENCES  -- Physical Sciences (hsv//eng)

Nyckelord

Physics
Fysik
fysik
Physics

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy