Benedek, Hunor (författare)
Lund University,Lunds universitet,Medicinsk strålningsfysik, Lund,Sektion V,Institutionen för kliniska vetenskaper, Lund,Medicinska fakulteten,Medical Radiation Physics, Lund,Section V,Department of Clinical Sciences, Lund,Faculty of Medicine
Isacsson, Ulf (författare)
Uppsala universitet,Medicinsk strålningsvetenskap
Olevik-Dunder, Maria (författare)
Karolinska University Hospital, Stockholm, Sweden
visa fler...
Westermark, Mathias (författare)
Karolinska University Hospital, Stockholm, Sweden
Hållström, Per (författare)
Gävle Hospital, Gavle, Swe
Olofsson, Jörgen (författare)
Umeå universitet,Radiofysik
Gustafsson, Magnus (författare)
Sahlgrenska University Hospital, Göteborg, Sweden
visa färre...
 (creator_code:org_t)
2015-01-12
2015
Engelska.
Ingår i: 8TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE). - : Institute of Physics (IOP). ; 573, s. 012015-012015
  • Konferensbidrag (refereegranskat)
Abstract Ämnesord
Stäng  
  • In late 2011 The Swedish Society of Radiation Physics formed a working group to concentrate on the Quality Assurance of modern radiation therapy techniques. The given task was to identify and summarise the different QA strategies in Sweden and also the international recommendations. This was used to formulate recommendations for practical guidelines within Sweden. In this paper a brief summery of the group's work is presented. All the Swedish radiation therapy centres do a pre treatment verification measurement as QA for every new IMRT and VMAT plan. Physicists do it and they believe it to be time consuming. A general standpoint from all the centres was that new guidelines and legislation is needed to allow QA that does not require a measurement. Based on various international publications and recommendations the working group has presented two strategies, one where all new plans are checked through measurement and one where no measurement is needed. The measurement-based strategy is basically the same as the one used today with an extended machine QA part. The other presented strategy is process oriented where all the different parts of the treatment chain are checked separately. The final report can be found in Swedish on http://www.radiofysik.org.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Annan medicin och hälsovetenskap -- Övrig annan medicin och hälsovetenskap (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Other Medical and Health Sciences -- Other Medical and Health Sciences not elsewhere specified (hsv//eng)
MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Radiologi och bildbehandling (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Radiology, Nuclear Medicine and Medical Imaging (hsv//eng)

Publikations- och innehållstyp

ref (ämneskategori)
kon (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy