Influence of electron contamination on in vivo surface dosimetry for high-energy photon beams

Sjogren, R (författare)
Umeå universitet,Radiofysik
Karlsson, Mikael (författare)
Umeå universitet,Radiofysik
 (creator_code:org_t)
1998
1998
Engelska.
Ingår i: Medical physics (Lancaster). - 0094-2405. ; 25:6, s. 916-921
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • The influence of the electron contamination at in vivo dosimetry with diodes on the patient surface has been investigated by introducing different accessories in the beam path and by changing the field size and SSD. The results show a clear correlation between the electron contamination at an effective measuring depth of the diode and the signal from the patient diode. When the electron contamination is taken into account the agreement between the diode values and the absorbed dose is greatly improved. More accurate in vivo dosimetry with less error margins is therefore possible if better predictions of the electron contamination in high-energy photon beams can be performed. (C) 1998 American Association of Physicists in Medicine. [S0094-2405(98)00606-3].

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Sjogren, R
Karlsson, Mikael
Artiklar i publikationen
Medical physics ...
Av lärosätet
Umeå universitet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy