SwePub
Tyck till om SwePub Sök här!
Sök i LIBRIS databas

  Utökad sökning

id:"swepub:oai:DiVA.org:umu-39964"
 

Sökning: id:"swepub:oai:DiVA.org:umu-39964" > Photon beam charact...

Photon beam characteristics on the MM50 racetrack microtron and a new approach for beam quality determination

Karlsson, Mikael (författare)
Umeå universitet,Radiofysik
Nyström, Håkan (författare)
Umeå universitet,Radiofysik
Svensson, Hans (författare)
Umeå universitet,Radiofysik
 (creator_code:org_t)
1998-06-04
1993
Engelska.
Ingår i: Medical physics (Lancaster). - : Wiley. - 0094-2405. ; 20:1, s. 143-149
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • The photon beams of the MM50 racetrack microtron have special characteristics which make them more suitable than conventional photon beams for precision radiation therapy with good dosimetric control. The beam flattening is obtained by the scanning of an elementary beam instead of using a flattening filter. This will give a number of advantages such as the possibility to optimize field flattening to individual field forms and field sizes. The radiation quality is the same across the whole beam, which gives smaller changes in dose profiles with depth and also makes it easier to perform careful dose planning. Beam collimation is mainly performed by a multileaf collimator and the special design of the treatment head gives nearly ideal characteristics for dose determination in an arbitrary point in the treatment fields. The output factor has been shown to depend almost solely on scattering within the treatment field. The conventional methods for beam quality characterization have been found less suitable at high energies and a new method based on HVL measurements in water is proposed.

Nyckelord

photon beams
racetrack microtrons
collimators
X−ray dosimetry
radiotherapy
electron beams
filters
beam shaping
radiation effects

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Karlsson, Mikael
Nyström, Håkan
Svensson, Hans
Artiklar i publikationen
Medical physics ...
Av lärosätet
Umeå universitet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy