Change in intraocular pressure measurement after myopic LASEK : a study evaluating Goldmann, Pascal and applanation resonance tonometry

Jóhannesson, Gauti (författare)
Umeå universitet,Oftalmiatrik
Hallberg, Per (författare)
Eklund, Anders (författare)
visa fler...
Koskela, Timo (författare)
Lindén, Christina (författare)
Umeå universitet,Oftalmiatrik
visa färre...
 (creator_code:org_t)
Lippincott Williams & Wilkins, 2012
2012
Engelska.
Ingår i: Journal of glaucoma. - : Lippincott Williams & Wilkins. - 1057-0829 .- 1536-481X. ; 21:4, s. 255-259
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • Purpose: To prospectively evaluate 3 tonometry methods—the gold standard, Goldmann applanation tonometry, a new method, Pascal dynamic contour tonometry (PDCT), and a method under development, applanation resonance tonometry (ART)—with respect to intraocular pressure (IOP) measurements before, 3 and 6 months after laser-assisted subepithelial keratectomy (LASEK).Materials and Methods: One randomly assigned eye of each of 53 healthy individuals, who underwent LASEK surgery for myopia was studied. Visual acuity, central corneal thickness, corneal curvature, and IOP were measured at each visit. Six IOP measurements/methods with 5 minutes pause between methods were performed.Results: All tonometry methods measured a significantly lower IOP after LASEK correction by a mean of −3.1 diopters. The IOP reduction was largest after 6 months for Goldmann applanation tonometry (−1.7±1.8 mm Hg) followed by ARTstat (−1.2±1.5 mm Hg), PDCT (−1.1±1.6 mm Hg), and ARTdyn (−1.0±1.5 mm Hg). The reduction of IOP did not differ significantly between different methods (P=0.11). There was a significant further reduction of measured IOP for PDCT between 3 and 6 months (−0.5±1.0 mm Hg). Uncorrected visual acuity improved significantly between 3 and 6 months postoperatively from 1.32±0.28 to 1.43±0.27.Conclusions: All tonometry methods measured a significant, but low, reduction of IOP 3 and 6 months after LASEK. Further change in visual acuity and IOP measurements between 3 and 6 months suggest a still ongoing postoperative process.

Ämnesord

MEDICIN OCH HÄLSOVETENSKAP  -- Klinisk medicin -- Oftalmologi (hsv//swe)
MEDICAL AND HEALTH SCIENCES  -- Clinical Medicine -- Ophthalmology (hsv//eng)

Nyckelord

LASEK
Goldmann applanation tonometry
dynamic contour tonometry
intraocular pressure
applanation resonance tonometry

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Jóhannesson, Gau ...
Hallberg, Per
Eklund, Anders
Koskela, Timo
Lindén, Christin ...
Om ämnet
MEDICIN OCH HÄLSOVETENSKAP
MEDICIN OCH HÄLS ...
och Klinisk medicin
och Oftalmologi
Artiklar i publikationen
Journal of glauc ...
Av lärosätet
Umeå universitet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy