A computerized Infusion Pump for control of tissue tracer concentration during Positron Emission Tomography in vivo Pharmacokinetic/Pharmacodynamic measurements

Eriksson, Olof (författare)
Uppsala universitet,Enheten för radiologi
Wallberg, Andreas (författare)
Syvänen, Stina (författare)
Uppsala universitet,Institutionen för farmaceutisk biovetenskap,Hammarlund-Udenaes
visa fler...
Josephsson, Raymond (författare)
Uppsala universitet,Klinisk virologi
Långström, Bengt (författare)
Uppsala universitet,Institutionen för biokemi och organisk kemi
Bergström, Mats (författare)
Uppsala universitet,Institutionen för farmaceutisk biovetenskap
visa färre...
 (creator_code:org_t)
2008-05-30
2008
Engelska.
Ingår i: BMC Medical Physics. - : Springer Science and Business Media LLC. - 1756-6649. ; 8:2
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • BACKGROUND:A computer controlled infusion pump (UIPump) for regulation of target tissue concentration of radioactive compounds was developed for use in biological research and tracer development for PET.METHODS:Based on observed tissue or plasma kinetics after a bolus injection of the tracer an algorithm calculates the infusion needed to obtain a specified target kinetic curve. A computer feeds this infusion scheme into an infusion pump connected to an animal via a venous catheter. The concept was validated using [11C]Flumazenil administrated to Sprague-Dawley rats where the whole brain distribution and kinetic of the tracer was measured over time using a microPET-scanner. The accuracy and precision of the system was assessed by producing steady-state levels of the tracer and by mimicking kinetics after oral administration.RESULTS:Various kinetic profiles could be generated, including rapid achievement of constant levels, or step-wise increased levels. The resulting tissue curves had low deviation from the target curves according to the specified criteria: AUC (%): 4.2 +/- 2.8, Maximal deviation (%): 13.6 +/- 5.0 and R2: 0.95 +/- 0.02.CONCLUSION:The UIPump-system is suitable for use in PET-research for assessment of PK/PD properties by simulation of different tracer tissue kinetics in vivo.

Nyckelord

MEDICINE
MEDICIN

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy