SwePub
Sök i LIBRIS databas

  Utökad sökning

id:"swepub:oai:DiVA.org:uu-103249"
 

Sökning: id:"swepub:oai:DiVA.org:uu-103249" > Small fields :

Small fields : nonequilibrium radiation dosimetry

Das, Indra J. (författare)
Ding, George X. (författare)
Ahnesjö, Anders, 1953- (författare)
Uppsala universitet,Enheten för onkologi
 (creator_code:org_t)
2007-12-20
2008
Engelska.
Ingår i: Medical physics (Lancaster). - : Wiley. - 0094-2405. ; 35:1, s. 206-215
  • Forskningsöversikt (refereegranskat)
Abstract Ämnesord
Stäng  
  • Advances in radiation treatment with beamlet-based intensity modulation, image-guided radiation therapy, and stereotactic radiosurgery (including specialized equipments like CyberKnife, Gamma Knife, tomotherapy, and high-resolution multileaf collimating systems) have resulted in the use of reduced treatment fields to a subcentimeter scale. Compared to the traditional radiotherapy with fields > or =4 x 4 cm2, this can result in significant uncertainty in the accuracy of clinical dosimetry. The dosimetry of small fields is challenging due to nonequilibrium conditions created as a consequence of the secondary electron track lengths and the source size projected through the collimating system that are comparable to the treatment field size. It is further complicated by the prolonged electron tracks in the presence of low-density inhomogeneities. Also, radiation detectors introduced into such fields usually perturb the level of disequilibrium. Hence, the dosimetric accuracy previously achieved for standard radiotherapy applications is at risk for both absolute and relative dose determination. This article summarizes the present knowledge and gives an insight into the future procedures to handle the nonequilibrium radiation dosimetry problems. It is anticipated that new miniature detectors with controlled perturbations and corrections will be available to meet the demand for accurate measurements. It is also expected that the Monte Carlo techniques will increasingly be used in assessing the accuracy, verification, and calculation of dose, and will aid perturbation calculations of detectors used in small and highly conformal radiation beams.

Nyckelord

small field dosimetry
nonequilibrium conditions
Monte Carlo
MEDICINE
MEDICIN

Publikations- och innehållstyp

ref (ämneskategori)
for (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Hitta mer i SwePub

Av författaren/redakt...
Das, Indra J.
Ding, George X.
Ahnesjö, Anders, ...
Artiklar i publikationen
Medical physics ...
Av lärosätet
Uppsala universitet

Sök utanför SwePub

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy