Evaluation of the quality of blood components obtained after automated separation of whole blood by a new multiunit processor

Lagerberg, Johan W. (författare)
Uppsala universitet,Klinisk immunologi
Salado-Jimena, Jose A. (författare)
Uppsala universitet,Klinisk immunologi
Lööf, Helena (författare)
Uppsala universitet,Klinisk immunologi
visa fler...
Bontekoe, Ido J. (författare)
Uppsala universitet,Klinisk immunologi
Nielsen, Connie (författare)
Uppsala universitet,Klinisk immunologi
Verheggen, Caroline (författare)
Uppsala universitet,Klinisk immunologi
van Waeg, Geert (författare)
Uppsala universitet,Klinisk immunologi
van der Meer, Pieter F. (författare)
Uppsala universitet,Klinisk immunologi
de Korte, Dirk (författare)
Uppsala universitet,Klinisk immunologi
Hansen, Morten B. (författare)
Uppsala universitet,Klinisk immunologi
Knutson, Folke (författare)
Uppsala universitet,Klinisk immunologi
visa färre...
 (creator_code:org_t)
2012-12-11
2013
Engelska.
Ingår i: Transfusion. - : Wiley. - 0041-1132 .- 1537-2995. ; 53:8, s. 1798-1807
  • Tidskriftsartikel (refereegranskat)
Abstract Ämnesord
Stäng  
  • BACKGROUND: The Reveos system (Terumo BCT) is a fully automated device able to process four whole blood (WB) units simultaneously into a plasma unit, a red blood cell (RBC) unit, and an interim platelet (PLT) unit (IPU). Multiple IPUs can be pooled to form a transfusable PLT product. The aim of our study was to evaluate the quality of components made with the Reveos system from either fresh (2-8hr) or overnight-held WB. STUDY DESIGN AND METHODS: A prototype of the Reveos system was used to process WB. RBCs were resuspended in SAGM, leukoreduced, and assayed for in vitro quality variables during a 42-day storage period at 2 to 6 degrees C. Twenty-four-hour in vivo recovery was determined on Day42. Plasma was assayed for cellular contamination and activation variables. IPUs were pooled with SSP+ additive solution for in vitro quality assessments during a 7-day storage period at room temperature. RESULTS: Reveos-produced RBCs and plasma units met the predefined requirements. RBC recovery was superior to control units. On Day42, hemolysis was below 0.8% and in vivo recovery was above 75% for all RBCs. Cellular contamination was lower for Reveos-produced plasma. PLT yield was higher with overnight-stored WB. PLT quality was well maintained during storage with no significant differences between the two groups. Conclusion: Blood components prepared with the Reveos from fresh or overnight-held WB meet quality criteria without any relevant difference between the two groups. The Reveos system has the potential to increase efficacy and standardization of blood component preparation.

Publikations- och innehållstyp

ref (ämneskategori)
art (ämneskategori)

Hitta via bibliotek

Till lärosätets databas

Kungliga biblioteket hanterar dina personuppgifter i enlighet med EU:s dataskyddsförordning (2018), GDPR. Läs mer om hur det funkar här.
Så här hanterar KB dina uppgifter vid användning av denna tjänst.

 
pil uppåt Stäng

Kopiera och spara länken för att återkomma till aktuell vy